Biomechanical evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures

被引:185
|
作者
Marti, A
Fankhauser, C
Frenk, A
Cordey, J
Gasser, B
机构
[1] Dr HC Robert MarthyS Stiftung, Belltach, Switzerland
[2] Jacques Cordey Biomech, Filisur, Switzerland
关键词
angular stability; biomechanical study; internal fixation; monocortical and bicortical fixation; supracondylar femoral fractures;
D O I
10.1097/00005131-200109000-00004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Comparison between a Less Invasive Stabilization System (LISS) using monocortical screws with angular stability and two conventional plate systems Condylar Buttress Plate (CBP) and Dynamic Condylar Screw (DCS) for the treatment of distal femoral fractures with respect to biomechanical properties. Design: Biomechanical study using paired cadaver femurs. In Test Configuration I (distal test), a ten-millimeter gap at the diaphysis-metaphysis junction simulates a supracondylar femoral fracture. Test Configuration 2 (proximal test) has the same configuration, but the gap was cut in the isthmic region. Proximal and distal plate ends were fixed to corresponding cortical bone fragments in both tests. Optical displacement transducers served to quantify the system's ability to withstand a stepwise increased load. Reversible (deflection) and irreversible deformation (subsidence) of the bone-plate construct was investigated. Results: In Test Configuration 1, LISS showed less irreversible deformation in 72 percent of the left-right comparisons. No correlation between bone mineral density, cross-section area of bones and the measured response of the construct under load was found between pairs. In Test Configuration 2, 83 percent of the left-right comparisons showed less permanent deformation but a higher elastic deformation for LISS. Conclusions: These results suggest an enhanced ability to withstand high loads when using the monocortical screw fixation technique with angular stability. A higher elastic deformation of LISS compared with conventional plating systems in distal femoral fractures can be explained by the lower bending stiffness caused by different design and material properties.
引用
收藏
页码:482 / 487
页数:6
相关论文
共 50 条
  • [21] FRACTURES OF DISTAL THIRD OF FEMUR TREATED BY INTERNAL-FIXATION
    CHIRON, HS
    TREMOULET, J
    CASEY, P
    MULLER, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1974, (100) : 160 - 170
  • [22] Failure of less invasive stabilization system plates in the distal femur - A report of four cases
    Button, G
    Wolinsky, P
    Hak, D
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) : 565 - 570
  • [23] Treatment of distal femoral fractures with the Less Invasive Stabilization System (LISS) in paraplegics
    Badke A.
    Höntzsch D.
    Kaps H.-P.
    Weise K.
    European Journal of Trauma, 2005, 31 (5): : 499 - 502
  • [24] Coronal fractures of the distal femoral condyle: A biomechanical evaluation of four internal fixation constructs
    Hak, DJ
    Nguyen, J
    Curtiss, S
    Hazelwood, S
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2005, 36 (09): : 1103 - 1106
  • [25] LISS-Tractor. Combination of the less invasive stabilization system (LISS) with the AO distractor for fractures of the distal femur and proximal tibia
    Babst, R
    Hehli, M
    Regazzoni, P
    UNFALLCHIRURG, 2001, 104 (06): : 530 - 535
  • [26] Fixation of periprosthetic femur fractures with the less invasive stabilization system (LISS) - a new minimally invasive treatment with locked fixed-angle screws
    Kolb, W
    Guhlmann, H
    Friedel, R
    Nestmann, H
    ZENTRALBLATT FUR CHIRURGIE, 2003, 128 (01): : 53 - 59
  • [27] Comparison of Less Invasive Stabilization System Plate and Retrograde Intramedullary Nail in the Fixation of Femoral Supracondylar Fractures in the Elderly: A Biomechanical Study
    Du, Yu-ren
    Ma, Jian-xiong
    Wang, Shuo
    Sun, Lei
    Wang, Ying
    Lu, Bin
    Bai, Hao-hao
    Hu, Yong-cheng
    Ma, Xin-long
    ORTHOPAEDIC SURGERY, 2019, 11 (02) : 311 - 317
  • [28] Biomechanical comparison of CRIF and reconstruction plate for stabilization of distal femur fractures in dogs
    Paes, F.
    Ferrigno, C. R. A.
    Marinho, P. V. T.
    Pereira, C. A. M.
    Bregadioli, T.
    Dal-bo, I. S.
    Ferraro, M. A.
    ARQUIVO BRASILEIRO DE MEDICINA VETERINARIA E ZOOTECNIA, 2019, 71 (04) : 1339 - 1347
  • [29] Functional outcome following internal fixation of intraarticular fractures of the distal femur
    Sahu, Ramji Lal
    ACTA ORTHOPAEDICA BELGICA, 2017, 83 (02): : 215 - 222
  • [30] Fixation of Periprosthetic Femur Fractures Above Total Knee Arthroplasty With the Less Invasive Stabilization System: A Midterm Follow-Up Study
    Kolb, Werner
    Guhlmann, Hanno
    Windisch, Christoph
    Marx, Frank
    Koller, Heiko
    Kolb, Klaus
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 69 (03): : 670 - 676