The value of the white precursor cell channel (WPC) on the Sysmex XN-1000 analyser in a specialist paediatric hospital

被引:20
|
作者
Jones, Amanda S. [1 ]
Tailor, Hitesh [2 ]
Liesner, Ri [1 ]
Machin, Samuel J. [2 ]
Briggs, Carol J. [2 ]
机构
[1] NHS Fdn Trust, Great Ormond St Hosp Children, Camelia Botnar Labs, Dept Haematol, London, England
[2] Univ Coll London Hosp, Dept Haematol, London, England
关键词
PAEDIATRIC HAEMATOLOGY; NEO-NATAL HAEMATOLOG; ANALYTICAL METHODS; PERFORMANCE EVALUATION; BLOOD; LYMPHOCYTES; MORPHOLOGY;
D O I
10.1136/jclinpath-2014-202640
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Background Historically, haematology analyser flags for abnormal white blood cells (WBCs) show good sensitivity but lower specificity, causing unnecessary blood film reviews. While the WBC differential channel on Sysmex XE and XN instruments reports a combined flag for blasts/abnormal lymphocytes, the new white precursor cell channel (WPC) on the XN series has been introduced to separate this into a specific flag for either cell type or, if no abnormality, remove the flag entirely. Aims To compare the efficiency of abnormal WBC flags from the XN WPC to our existing analyser and determine whether WPC can reduce false positive flags and blood films required. Methods Abnormal WBC flags from the Sysmex XE-5000 and XN-1000 were compared to manual differential and blood film morphology on 300 K(2)EDTA samples from infants and children. Results The XN WPC flag for blasts was more sensitive and specific than flags indicating blasts on the XE-5000, with a reduction in false positives from 64% (XE) to 36% (XN). Overall efficiency of the WPC flag for abnormal lymphocytes was 94% vs 79% on the XE. WPC reduced false positive flags for blasts and abnormal lymphocytes on neonatal samples by 50%. Automatic reflex analysis by WPC correctly removed a false positive flag from the white cell differential channel on 46% of samples. Total abnormal WBC flags from XN WPC were less (73) than the XE-5000 (92). Conclusions XN WPC demonstrated superior efficiency of abnormal WBC flags on paediatric samples, compared to the XE-5000, with greater sensitivity and specificity of flagging, reducing blood films for review.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 22 条
  • [21] Evaluation of automated synovial fluid total cell count and percent polymorphonuclear leukocytes on a Sysmex XN-1000 analyzer for identifying patients at risk of septic arthritis
    Saadalla, Abdulrahman
    Aguirre, Jose Jara
    Wockenfus, Amy M.
    Kelley, Brandon R.
    Swanson, Rebecca L.
    Howard, Matthew T.
    Karon, Brad S.
    CLINICA CHIMICA ACTA, 2020, 510 : 416 - 420
  • [22] WPC CHANNEL OF THE SYSMEX XN20 ANALYZER HAS LIMITED VALUE TO DISCRIMINATE BETWEEN BLASTS AND ABNORMAL LYMPHOCYTES IN ONCOHEMATOLOGIC PATIENTS WITH BLASTS IN PERIPHERAL BLOOD
    Ramon Furundarena, Jose
    Gonzalez, Carmen
    Uranga, Alasne
    Bourantas, Dimitrios
    Aranbarri, Ane
    Ondarra, Laida
    Lombardi, Clara
    Araiz, Maria
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 44 - 44