We report the findings from independent prospective clinical and laboratory-based joint-simulator studies of the performance of ceramic femoral heads of 22.225 mm diameter in cross-linked polyethylene (XLP) acetabular cups. We found remarkable qualitative and quantitative agreement between the clinical and simulator results for the wear characteristics with time, and confirmed that ceramic femoral heads penetrate the XLP cups at only about half the rate of otherwise comparable metal heads. In the clinical study, 19 hips in 17 patients were followed for an average of 77 months, In the hip-joint simulator a similar prosthesis was tested for 7.3 million cycles. Both clinical and simulator results showed relatively high rates of penetration over the first 18 months or 1.5 million cycles, followed by a very much lower wear thereafter, Once an initial bedding-in of 0.2 mm to 0.4 mm had taken place the subsequent rates of penetration were very small, The initial clinical wear during bedding-in averaged 0.29 mm/year; subsequent progression was an order of magnitude lower at about 0.022 mm/year, lower than the 0.07 mm/year in metal-to-UHMWP Charnley LFAs. Our results show the excellent tribological features of alumina-ceramic-to-XLP implants, and also confirm the value of well-designed joint simulators for the evaluation of total joint replacements.
引用
收藏
页码:280 / 285
页数:6
相关论文
共 50 条
  • [41] Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years
    Dorr, LD
    Wan, ZN
    Shahrdar, C
    Sirianni, L
    Boutary, M
    Yun, A
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08): : 1816 - 1821
  • [42] POLYETHYLENE SOCKETS AND ALUMINA CERAMIC HEADS IN CEMENTED TOTAL HIP-ARTHROPLASTY - A 10-YEAR STUDY
    SUGANO, N
    NISHII, T
    NAKATA, K
    MASUHARA, K
    TAKAOKA, K
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (04): : 548 - 556
  • [43] Cementless Total Hip Arthroplasty With Alumina-on-Highly Cross-Linked Polyethylene Bearing in Young Patients With Femoral Head Osteonecrosis
    Kim, Young-Hoo
    Choi, Yoowang
    Kim, Jun-Shik
    JOURNAL OF ARTHROPLASTY, 2011, 26 (02): : 218 - 223
  • [44] Comparison of Polyethylene Wear between Highly Cross linked and Annealed UHMWPE and Conventional UHMWPE against Ceramic Heads in Total Hip Arthroplasty
    Sato, Taishi
    Nakashima, Yasuharu
    Akiyama, Mio
    Yamamoto, Takuaki
    Mawatari, Taro
    Itokawa, Takashi
    Ohishi, Masanobu
    Motomura, Goro
    Hirata, Masanobu
    Iwamoto, Yukihide
    BIOCERAMICS 24, 2013, 529-530 : 279 - 284
  • [45] A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up
    Engh, Charles A., Jr.
    Hopper, Robert H., Jr.
    Huynh, Cathy
    Ho, Henry
    Sritulanondha, Supatra
    Engh, Charles A., Sr.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (08): : 2 - 7
  • [46] Wear Rate of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Randomized Controlled Trial
    McCalden, Richard W.
    MacDonald, Steven J.
    Rorabeck, Cecil H.
    Bourne, Robert B.
    Chess, David G.
    Charron, Kory D.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (04): : 773 - 782
  • [47] Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty
    Tower, Stephen S.
    Currier, John H.
    Currier, Barbara H.
    Lyford, Kimberly A.
    Van Citters, Douglas W.
    Mayor, Michael B.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (10): : 2212 - 2217
  • [48] Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years
    Nebergall, A. K.
    Greene, M. E.
    Laursen, M. B.
    Nielsen, P. T.
    Malchau, H.
    Troelsen, A.
    BONE & JOINT JOURNAL, 2017, 99-B (05): : 577 - 584
  • [49] Cross-Linked Polyethylene for Total Hip Arthroplasty Markedly Reduces Revision Surgery at 16 Years
    de Steiger, R.
    Lorimer, M.
    Graves, S. E.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2018, 100 (15): : 1281 - 1288
  • [50] Long-Term Survivorship of Total Hip Arthroplasty with Highly Cross-Linked Polyethylene for Osteonecrosis
    Hart, Adam
    Janz, Viktor
    Trousdale, Robert T.
    Sierra, Rafael J.
    Berry, Daniel J.
    Abdel, Matthew P.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2019, 101 (17): : 1563 - 1568