Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups

被引:188
|
作者
Wroblewski, BM [1 ]
Siney, PD [1 ]
Dowson, D [1 ]
Collins, SN [1 ]
机构
[1] UNIV LEEDS,DEPT MECH ENGN,LEEDS LS2 9JT,W YORKSHIRE,ENGLAND
来源
关键词
D O I
10.1302/0301-620X.78B2.0780280
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report the findings from independent prospective clinical and laboratory-based joint-simulator studies of the performance of ceramic femoral heads of 22.225 mm diameter in cross-linked polyethylene (XLP) acetabular cups. We found remarkable qualitative and quantitative agreement between the clinical and simulator results for the wear characteristics with time, and confirmed that ceramic femoral heads penetrate the XLP cups at only about half the rate of otherwise comparable metal heads. In the clinical study, 19 hips in 17 patients were followed for an average of 77 months, In the hip-joint simulator a similar prosthesis was tested for 7.3 million cycles. Both clinical and simulator results showed relatively high rates of penetration over the first 18 months or 1.5 million cycles, followed by a very much lower wear thereafter, Once an initial bedding-in of 0.2 mm to 0.4 mm had taken place the subsequent rates of penetration were very small, The initial clinical wear during bedding-in averaged 0.29 mm/year; subsequent progression was an order of magnitude lower at about 0.022 mm/year, lower than the 0.07 mm/year in metal-to-UHMWP Charnley LFAs. Our results show the excellent tribological features of alumina-ceramic-to-XLP implants, and also confirm the value of well-designed joint simulators for the evaluation of total joint replacements.
引用
收藏
页码:280 / 285
页数:6
相关论文
共 50 条
  • [1] Ceramic on highly cross-linked polyethylene in cementless total hip arthroplasty
    Kong, J. -S.
    Moon, K. -H.
    BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2007, : 67 - 70
  • [2] Oxidation in ultrahigh molecular weight polyethylene and cross-linked polyethylene acetabular cups tested against roughened femoral heads in a hip joint simulator
    Taddei, P
    Affatato, S
    Fagnano, C
    Toni, A
    BIOMACROMOLECULES, 2006, 7 (06) : 1912 - 1920
  • [3] Raman characterisation of conventional and cross-linked polyethylene in acetabular cups run on a hip joint simulator
    Taddei, P.
    Di Foggia, M.
    Affatato, S.
    JOURNAL OF RAMAN SPECTROSCOPY, 2011, 42 (06) : 1344 - 1352
  • [4] Cross-linked UHMWPE on alumina total hip arthroplasty
    Oonishi, H
    Clarke, IC
    Good, V
    Amino, H
    Masuda, S
    ARTHROPLASTY 2000: RECENT ADVANCES IN TOTAL JOINT REPLACEMENT, 2001, : 3 - 9
  • [5] Clinical Results of the Wear Performance of Cross-Linked Polyethylene in Total Hip Arthroplasty Prospective Randomized Trial
    Ise, Kentaro
    Kawanabe, Keiichi
    Tamura, Jiro
    Akiyama, Haruhiko
    Goto, Koji
    Nakamura, Takashi
    JOURNAL OF ARTHROPLASTY, 2009, 24 (08): : 1216 - 1220
  • [6] Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study
    Zoran Bascarevic
    Zoran Vukasinovic
    Nemanja Slavkovic
    Borislav Dulic
    Goran Trajkovic
    Violeta Bascarevic
    Sladjan Timotijevic
    International Orthopaedics, 2010, 34 : 1129 - 1135
  • [7] Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study
    Bascarevic, Zoran
    Vukasinovic, Zoran
    Slavkovic, Nemanja
    Dulic, Borislav
    Trajkovic, Goran
    Bascarevic, Violeta
    Timotijevic, Sladjan
    INTERNATIONAL ORTHOPAEDICS, 2010, 34 (08) : 1129 - 1135
  • [8] Mean 16-Year Results of Total Hip Arthroplasty With Alumina Ceramic Femoral Heads on Highly Cross-Linked Polyethylene in Patients 50 Years or Less
    Youngman, Tyler R.
    Verhotz, Daniel R.
    Layon, Daniel R.
    Parilla, Frank W.
    Pashos, Gail E.
    Thornton, Tanner
    Bendich, Ilya
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2023, 38 (07): : S346 - S350
  • [9] Conventional Versus Cross-Linked Polyethylene for Total Hip Arthroplasty
    Surace, Michele F.
    Monestier, Luca
    Vulcano, Ettore
    Harwin, Steven F.
    Cherubino, Paolo
    ORTHOPEDICS, 2015, 38 (09) : 556 - 561
  • [10] Impingement of acetabular cups in a hip simulator - Comparison of highly cross-linked and conventional polyethylene
    Holley, KG
    Furman, BD
    Babalola, OM
    Lipman, JD
    Padgett, DE
    Wright, TM
    JOURNAL OF ARTHROPLASTY, 2005, 20 (07): : 77 - 86