The validity of Acuros BV and TG-43 for high-dose-rate brachytherapy superficial mold treatments

被引:20
|
作者
Boman, Eeva L. [1 ,2 ,3 ]
Satherley, Thomas W. S. [1 ]
Schleich, Nanette [1 ]
Paterson, Dean B. [1 ]
Greig, Lynne [1 ]
Louwe, Rob J. W. [1 ]
机构
[1] Wellington Hosp, Blood & Canc Ctr, Wellington, New Zealand
[2] Tampere Univ Hosp, Dept Oncol, POB 2000, FI-33521 Tampere, Finland
[3] Tampere Univ Hosp, Dept Med Phys, Tampere, Finland
关键词
HDR brachytherapy; Surface mold; Skin treatment; Acuros BV; TG-43; Film dosimetry; EQUIVALENT PHANTOM MATERIALS; IR-192; BRACHYTHERAPY; WATER-EQUIVALENT; SKIN BRACHYTHERAPY; DOSIMETRY; AAPM; APPLICATORS; SCATTER; PHOTON; CS-137;
D O I
10.1016/j.brachy.2017.08.010
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work is to validate the Acuros BV dose calculation algorithm for high-dose-rate (HDR) brachytherapy superficial mold treatments in the absence of full scatter conditions and compare this with TG-43 dose calculations. We also investigate the impact of additional back scatter material (bolus) applied above surface molds to the dose distributions under the mold. METHODS AND MATERIALS: The absorbed dose at various depths was compared for simulations performed using either TG-43 or Acuros BV dose calculations. Parameter variations included treatment area, thickness of the bolus, and surface shape (flat or spherical). Film measurements were carried out in a flat phantom. RESULTS: Acuros BV calculations and film measurements agreed within 1.5% but were up to 15% lower than TG-43 dose calculations when no bolus was applied above the treatment catheters. The difference in dose at the prescription depth (1 cm below the central catheter) increased with increasing treatment area: 3.3% difference for a 3 x 3.5 cm(2) source loading area, 7.4% for 8 x 9 cm(2), and 13.4% for 18 x 19 cm(2). The dose overestimation of the TG-43 model decreased when bolus was added above the treatment catheters. CONCLUSIONS: The TG-43 dosimetry formalism cannot model surface mold treatments in the absence of full scatter conditions within 5% for loading areas larger than approximately 5 x 5 cm(2). The TG-43 model results in an overestimation of the delivered dose, which increases with treatment area. This confirms the need for model-based dose calculation algorithms as discussed in TG-186. (C) 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1280 / 1288
页数:9
相关论文
共 50 条
  • [31] VERIFICATION OF HIGH-DOSE-RATE BRACHYTHERAPY PLANS
    MOWER, HW
    LO, TC
    GIRSHOVICH, L
    LAWSON, BJ
    COSTEY, GE
    RADIOLOGY, 1992, 185 : 129 - 130
  • [32] COMPARISON OF DVH-PARAMETERS FOR DOSE CALCULATION BY TG43 AND BOLTZMANN BASED ALGORITHM ACUROS BV™
    Scheuch, J.
    Nesvacil, N.
    Resch, A.
    Poetter, R.
    Kirisits, C.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S151 - S151
  • [33] Monte Carlo study of TG-43 dosimetry parameters of GammaMed Plus high dose rate 192Ir brachytherapy source using TOPAS
    Wu, Jianan
    Xie, Yaoqin
    Ding, Zhen
    Li, Feipeng
    Wang, Luhua
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 146 - 153
  • [34] High-dose-rate skin brachytherapy with interstitial, surface, or a combination of interstitial and surface mold technique
    Brovchuk, Serhii
    Park, Sang-June
    Shepil, Zoia
    Romanenko, Serhii
    Vaskevych, Oleg
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (02) : 107 - 114
  • [35] Custom High-Dose-Rate Brachytherapy Surface Mold Applicators: The Importance Source to Skin Distance
    Park, S.
    Demanes, J.
    Kamrava, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3385 - 3386
  • [36] RapidBrachyTG43: A Geant4-based TG-43 parameter and dose calculation module for brachytherapy dosimetry
    Kalinowski, Jonathan
    Enger, Shirin A.
    MEDICAL PHYSICS, 2024, 51 (05) : 3746 - 3757
  • [37] Low-dose-rate brachytherapy is superior to high-dose-rate brachytherapy for bladder cancer
    Pos, FJ
    Horenblas, S
    Lebesque, J
    Moonen, L
    Schneider, C
    Sminia, P
    Bartelink, H
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (03): : 696 - 705
  • [38] Characterization of Dose Distributions in High-Dose-Rate Surface Brachytherapy
    Buzurovic, I.
    Hansen, J.
    Bhagwat, M.
    O'Farrell, D.
    Damato, A.
    Friesen, S.
    Devlin, P.
    Cormack, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3369 - 3370
  • [39] Dosimetric comparison of AcurosBV with AAPM TG43 dose calculation formalism in cervical intraductal high-dose-rate brachytherapy using three different applicators
    Bi, Su-yan
    Chen, Zhi-jian
    Sun, Xing-ru
    Dai, Zhi-tao
    PRECISION RADIATION ONCOLOGY, 2022, 6 (03): : 234 - 242
  • [40] Postoperative high-dose-rate brachytherapy in the prevention of keloids
    Veen, Ronald E.
    Kal, Henk B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (04): : 1205 - 1208