The validity of Acuros BV and TG-43 for high-dose-rate brachytherapy superficial mold treatments

被引:20
|
作者
Boman, Eeva L. [1 ,2 ,3 ]
Satherley, Thomas W. S. [1 ]
Schleich, Nanette [1 ]
Paterson, Dean B. [1 ]
Greig, Lynne [1 ]
Louwe, Rob J. W. [1 ]
机构
[1] Wellington Hosp, Blood & Canc Ctr, Wellington, New Zealand
[2] Tampere Univ Hosp, Dept Oncol, POB 2000, FI-33521 Tampere, Finland
[3] Tampere Univ Hosp, Dept Med Phys, Tampere, Finland
关键词
HDR brachytherapy; Surface mold; Skin treatment; Acuros BV; TG-43; Film dosimetry; EQUIVALENT PHANTOM MATERIALS; IR-192; BRACHYTHERAPY; WATER-EQUIVALENT; SKIN BRACHYTHERAPY; DOSIMETRY; AAPM; APPLICATORS; SCATTER; PHOTON; CS-137;
D O I
10.1016/j.brachy.2017.08.010
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work is to validate the Acuros BV dose calculation algorithm for high-dose-rate (HDR) brachytherapy superficial mold treatments in the absence of full scatter conditions and compare this with TG-43 dose calculations. We also investigate the impact of additional back scatter material (bolus) applied above surface molds to the dose distributions under the mold. METHODS AND MATERIALS: The absorbed dose at various depths was compared for simulations performed using either TG-43 or Acuros BV dose calculations. Parameter variations included treatment area, thickness of the bolus, and surface shape (flat or spherical). Film measurements were carried out in a flat phantom. RESULTS: Acuros BV calculations and film measurements agreed within 1.5% but were up to 15% lower than TG-43 dose calculations when no bolus was applied above the treatment catheters. The difference in dose at the prescription depth (1 cm below the central catheter) increased with increasing treatment area: 3.3% difference for a 3 x 3.5 cm(2) source loading area, 7.4% for 8 x 9 cm(2), and 13.4% for 18 x 19 cm(2). The dose overestimation of the TG-43 model decreased when bolus was added above the treatment catheters. CONCLUSIONS: The TG-43 dosimetry formalism cannot model surface mold treatments in the absence of full scatter conditions within 5% for loading areas larger than approximately 5 x 5 cm(2). The TG-43 model results in an overestimation of the delivered dose, which increases with treatment area. This confirms the need for model-based dose calculation algorithms as discussed in TG-186. (C) 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1280 / 1288
页数:9
相关论文
共 50 条
  • [1] Dosimetric comparison of Acuros™ BV and AAPM TG-43 formalism for interstitial iridium-192 high-dose-rate brachytherapy
    Roussakis, Yiannis
    Antorkas, Georgios
    Georgiou, Leonidas
    Strouthos, Iosif
    Karagiannis, Efstratios
    Zamboglou, Constantinos
    Ferentinos, Konstantinos
    Zamboglou, Nikolaos
    Anagnostopoulos, Georgios
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2024, 16 (03) : 211 - 218
  • [2] Limitations of the TG-43 formalism for skin high-dose-rate brachytherapy dose calculations
    Granero, Domingo
    Perez-Calatayud, Jose
    Vijande, Javier
    Ballester, Facundo
    Rivard, Mark J.
    MEDICAL PHYSICS, 2014, 41 (02)
  • [3] Limitations and Clinical Implications of the TG-43 Formalism for High-Dose-Rate Skin Brachytherapy
    Granero, D.
    Perez-Calatayud, J.
    Vijande, J.
    Ballester, F.
    Candela-Juan, C.
    Rivard, M.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [4] Comparison of Acuros BV and TG-43 Dose Calculation for HDR Skin Brachytherapy with Flap Applicator in Heterogeneous Media
    Kang, H.
    Patel, R.
    Gros, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E113 - E114
  • [5] Evaluation of BrachyDose Monte Carlo code for HDR brachytherapy: dose comparison against Acuros®BV and TG-43 algorithms
    Dagli, Ayse
    Yurt, Fatma
    Yegin, Gultekin
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2020, 19 (01) : 76 - 83
  • [6] Dosimetric comparison of Acuros™ BV with AAPM TG43 dose calculation formalism in breast interstitial high-dose-rate brachytherapy with the use of metal catheters
    Sinnatamby, Mourougan
    Nagarajan, Vivekanandan
    Reddy, Sathyanarayana
    Karunanidhi, Gunaseelan
    Singhavajala, Vivekanandam
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2015, 7 (04) : 273 - 279
  • [7] The TG-43 brachytherapy dose calculation formalism
    Rivard, Mark J.
    Brachytherapy Physics, Second Edition, 2005, (31): : 295 - 324
  • [8] A Semi-2D TG-43 Formula for Accurate In-House Verification of High-Dose-Rate Brachytherapy Dose Calculation
    Fan, Q.
    Zheng, D.
    Wang, S.
    Lin, C.
    Enke, C.
    Zhou, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E573 - E573
  • [9] Dosimetry comparison between TG-43 and Monte Carlo calculations using the Freiburg flap for skin high-dose-rate brachytherapy
    Vijande, Javier
    Ballester, Facundo
    Ouhib, Zoubir
    Granero, Domingo
    Carmen Pujades-Claumarchirant, M.
    Perez-Calatayud, Jose
    BRACHYTHERAPY, 2012, 11 (06) : 528 - 535
  • [10] Dose Comparison of TG-43 and Acuros Algorithm to Account Lateral Scatter in Dose Distribution
    Deshpande, S.
    Patwe, P.
    Avdhoot, S.
    Anand, V.
    Bajpai, R.
    Kannan, V.
    MEDICAL PHYSICS, 2013, 40 (06)