Evaluation of primary stability of cylindrical and tapered implants in different bone types by measuring implant displacement: An In vitro study

被引:13
|
作者
Sugiura, Tsutomu [1 ]
Yamamoto, Kazuhiko [1 ]
Horita, Satoshi [1 ]
Murakami, Kazuhiro [1 ]
Kirita, Tadaaki [1 ]
机构
[1] Nara Med Univ, Dept Oral & Maxillofacial Surg, 840 Shijo Cho, Kashihara, Nara 6348522, Japan
基金
日本学术振兴会;
关键词
Bone density; dental implants; torque; RESONANCE FREQUENCY-ANALYSIS; INSERTION TORQUE; DENTAL IMPLANTS; THICKNESS; MICROMOTION; DENSITY; DESIGN; REGIONS; HYBRID; STRAIN;
D O I
10.4103/ccd.ccd_788_18
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aims: The purposes of this study were to investigate the primary stability of cylindrical and tapered implants in different bone types by measuring implant displacement and to examine the relationship between insertion torque value (ITV) and implant displacement. Materials and Methods: Four different polyurethane bone models consisted of low-density or low-to-medium-density cancellous bone with or without a cortical bone layer. A total of 120 cylindrical and tapered implants were placed into bone blocks (n = 15 in each group), and the ITV was measured. A lateral load of 15 N was applied to the top of the abutment, and implant displacement was recorded. Results: Implant displacement was significantly affected by cancellous bone density and to a lesser degree by cortical bone thickness. The displacement of tapered implants was significantly smaller than that of cylindrical implants in the presence of cortical bone. However, both implant groups showed similar ITV in the low-density cancellous bone model with the cortical bone layer. There was a correlation between ITV and displacement in the cylindrical and tapered implants. However, no correlation was observed between ITV and displacement within each bone type. Conclusions: Implant stability depended mainly on the bone type, whereas implant design had a limited influence on primary stability. The use of tapered implants may be advantageous for improving primary stability in patients with low-density cancellous bone only when crestal cortical bone exists. The same ITV of cylindrical and tapered implants did not necessarily represent similar primary stability in the bone type.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 50 条
  • [21] Relationship between Implant Geometry and Primary Stability in Different Bony Defects and Variant Bone Densities: An In Vitro Study
    Ibrahim, Ahmad
    Heitzer, Marius
    Bock, Anna
    Peters, Florian
    Moehlhenrich, Stephan Christian
    Hoelzle, Frank
    Modabber, Ali
    Kniha, Kristian
    MATERIALS, 2020, 13 (19) : 1 - 16
  • [22] Evaluation of Implant Stability and Marginal Bone Level of Maxillary Immediate Implants Augmented with Two Different Bone Grafting Materials: A Comparative Study
    Alnuaimy, Husam A.
    Hassan, Thair A.
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2020, 8 (01): : 139 - 143
  • [23] Influence of the Localization of Frontal Bone Defects on Primary Stability Values of 2 Different Implant Designs: An In Vitro Study
    Mercan, Ugur
    Bilhan, Hakan
    Meral, Deniz Gokce
    Kazokoglu, Sehnaz
    Aykol-Sahin, Gokce
    Geckili, Onur
    IMPLANT DENTISTRY, 2017, 26 (01) : 87 - 94
  • [24] Two Different Implant Designs and Impact of Related Drilling Protocols on Primary Stability in Different Bone Densities: An In Vitro Comparison Study
    Sennerby, Lars
    Pagliani, Luca
    Petersson, Anders
    Verrocchi, Damiano
    Volpe, Stefano
    Andersson, Peter
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (03) : 564 - 568
  • [25] Biomechanical and Bone Histomorphological Evaluation of Two Surfaces on Tapered and Cylindrical Root Form Implants: An Experimental Study in Dogs
    Negri, Bruno
    Luis Calvo-Guirado, Jose
    Mate Sanchez de Val, Jose Eduardo
    Delgado Ruiz, Rafael Arcesio
    Ramirez Fernandez, Maria Piedad
    Gomez Moreno, Gerardo
    Aguilar Salvatierra, Antonio
    Guardia, Javier
    Munoz Guzon, Fernando
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (06) : 799 - 808
  • [26] In Vitro Evaluation of the Primary Stability of Short and Conventional Implants
    Pimentel Lopes de Oliveira, Guilherme Jose
    Borelli Barros-Filho, Luiz Antonio
    Borelli Barros, Luiz Antonio
    Queiroz, Thalita Pereira
    Marcantonio, Elcio, Jr.
    JOURNAL OF ORAL IMPLANTOLOGY, 2016, 42 (06) : 458 - 463
  • [27] Effect of Tapered and Cylindrical Implants on Stress Distribution in Different Bone Qualities: Finite Element Analysis
    Sivrikaya, E. Can
    Omezli, M. Melih
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2019, 34 (06) : E99 - E105
  • [28] Relationship between the CT Value and Cortical Bone Thickness at Implant Recipient Sites and Primary Implant Stability with Comparison of Different Implant Types
    Howashi, Miori
    Tsukiyama, Yoshihiro
    Ayukawa, Yasunori
    Isoda-Akizuki, Kei
    Kihara, Masafumi
    Imai, Yu
    Sogo, Motofumi
    Koyano, Kiyoshi
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2016, 18 (01) : 107 - 116
  • [29] Primary Stability of Trabecular Metal Implant in Comparison to Fully Threaded Implants: In Vitro Study Simulating Immediate Implant Placement
    Magic, Marko
    Zupanek, Gorana
    Lazic, Zoran
    El Chaar, Edgard
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (06) : 584 - 589
  • [30] Evaluation of Peri-Implant Crestal Bone Loss with Different Implant Systems, Primary Stability, Bone Density and Soft Tissue Thickness: A Retrospective Study
    Shenoy A.
    Rohinikumar S.
    Maiti S.
    Sivaswamy V.
    Rajaraman V.
    Journal of Long-Term Effects of Medical Implants, 2023, 33 (04) : 53 - 58