Design and application of an assessment protocol for electromagnetic tracking systems

被引:106
|
作者
Hummel, JB [1 ]
Bax, MR
Figl, ML
Kang, Y
Maurer, C
Birkfellner, WW
Bergmann, H
Shahidi, R
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, Vienna, Austria
[2] Stanford Univ, Sch Med, Image Guidance Labs, Stanford, CA 94305 USA
[3] Ludwig Boltzmann Inst Nucl Med & Endocrinol, Vienna, Austria
基金
奥地利科学基金会;
关键词
D O I
10.1118/1.1944327
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper defines a simple protocol for competitive and quantified evaluation of electromagnetic tracking systems such as the NDI Aurora (A) and Ascension microBIRD with dipole transmitter (B). It establishes new methods and a new phantom design which assesses the reproducibility and allows comparability with different tracking systems in a consistent environment. A machined base plate was designed and manufactured in which a 50 mm grid of holes was precisely drilled for position measurements. In the center a circle of 32 equispaced holes enables the accurate measurement of rotation. The sensors can be clamped in a small mount which fits into pairs of grid holes on the base plate. Relative positional/orientational errors are found by subtracting the known distances/ rotations between the machined locations from the differences of the mean observed positions/ rotation. To measure the influence of metallic objects we inserted rods made of steel (SST 303, SST 416), aluminum, and bronze into the sensitive volume between sensor and emitter. We calculated the fiducial registration error and fiducial location error with a standard stylus calibration for both tracking systems and assessed two different methods of stylus calibration. The positional jitter amounted to 0.14 mm(A) and 0.08 mm(B). A relative positional error of 0.96 mm +/- 0.68 mm, range -0.06 mm; 2.23 mm(A) and 1.14 mm 0.78 mm, range -3.72 min; 1.57 mm(B) for a given distance of 50 mm was found. The relative rotation error was found to be 0.51 degrees (A)/0.04 degrees (B). The most relevant distortion caused by metallic objects results from SST 416. The maximum error 4.2 mm(A) / >= 100 min(B) occurs when the rod is close to the sensor(20 mm). While (B) is more sensitive with respect to metallic objects, (A) is less accurate concerning orientation measurements. (B) showed a systematic error when distances are calculated. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2371 / 2379
页数:9
相关论文
共 50 条
  • [1] Accuracy assessment protocols for electromagnetic tracking systems
    Frantz, DD
    Wiles, AD
    Leis, SE
    Kirsch, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14): : 2241 - 2251
  • [2] Accuracy assessment protocols for electromagnetic tracking systems
    Frantz, D
    Kirsch, S
    Leis, S
    Wiles, A
    CARS 2003: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2003, 1256 : 1361 - 1361
  • [3] Evaluation of a new electromagnetic tracking system using a standardized assessment protocol
    Hummel, J
    Figl, M
    Birkfellner, W
    Bax, MR
    Shahidi, R
    Maurer, CR
    Bergmann, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (10): : N205 - N210
  • [4] Design of an Automated Calibration System for Electromagnetic Tracking Systems
    Jayaram, Sankar
    Jayaram, Uma
    Palmer, Craig
    DETC 2008: PROCEEDINGS OF THE ASME INTERNATIONAL DESIGN ENGINEERING TECHNICAL CONFERENCES AND COMPUTERS AND INFORMATIONAL IN ENGINEERING CONFERENCE, VOL 3, PTS A AND B: 28TH COMPUTERS AND INFORMATION IN ENGINEERING CONFERENCE, 2009, : 1613 - 1624
  • [5] Electromagnetic Tracking for Tomotherapy Systems
    McGary, J. E.
    Chen, J.
    Xiong, Z.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S716 - S716
  • [6] Application of models in the design of underwater electromagnetic signature reduction systems
    Holmes, John J.
    NAVAL ENGINEERS JOURNAL, 2007, 119 (04) : 19 - 29
  • [7] Repeatable Assessment Protocol for Electromagnetic Trackers
    Haidegger, Tamas
    Sirokai, Beata
    Fenyvesi, Gabor
    Kovacs, Levente
    Benyo, Balazs
    Benyo, Zoltan
    MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [8] A comprehensive quality assurance protocol for electromagnetic tracking in brachytherapy
    Duerrbeck, Christopher
    Gomez-Sarmiento, Isaac Neri
    Androulakis, Ioannis
    Sauer, Birte Christina
    Kolkman-Deurloo, Inger-Karine
    Bert, Christoph
    Beaulieu, Luc
    MEDICAL PHYSICS, 2024, 51 (05) : 3184 - 3194
  • [9] Stability of miniature electromagnetic tracking systems
    Schicho, K
    Figl, M
    Donat, M
    Birkfellner, W
    Seemann, R
    Wagner, A
    Bergmann, H
    Ewers, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (09): : 2089 - 2098
  • [10] Assessment of metallic distortions of a electromagnetic tracking system
    Kirsch, Stefan R.
    Schilling, Christian
    Brunner, Georg
    MEDICAL IMAGING 2006: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2006, 6141