ELECTRODE-SKIN IMPEDANCE COMPONENT ESTIMATION IN THE TIME-DOMAIN

被引:0
|
作者
Fortune, Benjamin C. [1 ]
McKenzie, Lachlan R. [1 ]
Chatfield, Logan T. [1 ]
Pretty, Christopher G. [1 ]
机构
[1] Univ Canterbury, Dept Mech Engn, Ctr Bioengn, Christchurch, New Zealand
关键词
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents a method to estimate the individual component values of a bipolar electrode-skin interface, with future intent of applying compensatory electrode-skin impedance balancing prior recording bio-signals with electromyography. The electrode-skin interface was stimulated by a step input and the output behaviour was characterised using a single exponential model per electrode. The method was applied to simulated circuitry, passive component circuitry and a human subject. The accuracy of the method was determined using the known values that comprised the simulated and passive component circuitry. Nine of ten simulated data sets resulted in accurate estimations, with a maximum error of 0.763% and a mean error of 0.076% per component. The method also produced successful estimates for nine of the ten physical circuitry data sets, with a maximum error of 10.2% and a mean error of 3.49% per component. The method was unsuccessful in estimating the individual electrode skin impedance components for the human subject: this was due to the system failing to reach steady state during the stimulation period. The authors suspect a DC offset caused by the half-cell potentials associated with the electrode-skin interface were the cause of the unexpected behaviour.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Electrode-Skin Impedance Model Parameter Estimation in the Frequency-Domain
    Cameron, Chris J.
    Fortune, Benjamin C.
    Pretty, Christopher G.
    Hayes, Michael P.
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [2] Electrode-skin impedance imbalance measured in the frequency domain
    Fortune, Benjamin C.
    Pretty, Christopher G.
    Cameron, Chris J.
    McKenzie, Lachlan R.
    Chatfield, Logan T.
    Hayes, Michael P.
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2021, 63
  • [3] TIME AND FREQUENCY-DEPENDENCE OF DISPOSABLE ECG ELECTRODE-SKIN IMPEDANCE
    OLSON, WH
    SCHMINCKE, DR
    HENLEY, BL
    MEDICAL INSTRUMENTATION, 1979, 13 (05): : 269 - 272
  • [4] Electrode-skin impedance from a dielectric viewpoint
    Calderwood, JH
    PHYSIOLOGICAL MEASUREMENT, 1996, 17 : A131 - A139
  • [5] A practical approach to electrode-skin impedance unbalance measurement
    Spinelli, Enrique M.
    Mayosky, Miguel A.
    Pallas-Areny, Ramon
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (07) : 1451 - 1453
  • [6] TIME-DOMAIN ESTIMATION OF MT IMPEDANCE TENSOR
    SPAGNOLINI, U
    GEOPHYSICS, 1994, 59 (05) : 712 - 721
  • [7] The effect of electrode-skin interface model in electrical impedance imaging
    Huang, JJ
    Cheng, KS
    1998 INTERNATIONAL CONFERENCE ON IMAGE PROCESSING - PROCEEDINGS, VOL 3, 1998, : 837 - 840
  • [8] Electrode-Skin contact impedance: In vivo measurements on an ovine model
    Nguyen, D. T.
    Kosobrodov, R.
    Barry, M. A.
    Chik, W.
    Jin, C.
    Oh, T. I.
    Thiagalingam, A.
    McEwan, A.
    XV INTERNATIONAL CONFERENCE ON ELECTRICAL BIO-IMPEDANCE (ICEBI) & XIV CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY (EIT), 2013, 434
  • [9] Novel electrode-skin interface for breast electrical impedance scanning
    Ji, Zhenyu
    Dong, Xiuzhen
    Shi, Xuetao
    You, Fusheng
    Fu, Feng
    Liu, Ruigang
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2009, 47 (10) : 1045 - 1052
  • [10] Comparison of Three Kinds of Electrode-Skin Interfaces for Electrical Impedance Scanning
    Yin, Yinsuo
    Ji, Zhenyu
    Zhang, Wen
    Wang, Nan
    Fu, Feng
    Liu, Ruigang
    You, Fusheng
    Shi, Xuetao
    Dong, Xiuzhen
    ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (06) : 2032 - 2039