Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study

被引:0
|
作者
van Loon, Kim [1 ]
Peelen, Linda M. [1 ,2 ]
van de Vlasakker, Emmy C. [1 ]
Kalkman, Cor J. [1 ]
van Wolfswinkel, Leo [1 ]
van Zaane, Bas [1 ]
机构
[1] Univ Utrecht, Univ Med Ctr Utrecht, Dept Anesthesiol Intens Care & Emergency Med, POB 85500, NL-3508 GA Utrecht, Netherlands
[2] Univ Utrecht, Univ Med Ctr Utrecht, Julius Ctr Hlth Sci & Primary Care, Utrecht, Netherlands
关键词
ANTECEDENTS; AGREEMENT; ARREST;
D O I
10.1007/s12630-018-1214-z
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
PurposeAltered respiratory rate (RR) has been identified as an important predictor of serious adverse events during hospitalization. Introduction of a well-tolerated continuous RR monitor could potentially reduce serious adverse events such as opioid-induced respiratory depression. The purpose of this study was to investigate the ability of different monitor devices to detect RR in low care clinical settings.MethodsThis was a prospective method-comparison study with a cross-sectional design. Thoracic impedance pneumography (IPG), frequency modulated continuous wave radar, and an acoustic breath sounds monitor were compared with the gold standard of capnography for their ability to detect RR in breaths per minute (breaths min( -1)) in awake postoperative patients in the postanesthesia care unit. The Bland and Altman method for repeated measurements and mixed effect modelling was used to obtain bias and limits of agreement (LoA). Furthermore, the ability of the three devices to assist with correct treatment decisions was evaluated in Clarke Error Grids.ResultsTwenty patients were monitored for 1,203 min, with a median [interquartile range] of 61 [60-63] min per patient. The bias (98.9% LoA) were 0.1 (-7.9 to 7.9) breaths min( -1) for the acoustic monitor, -1.6 (-10.8 to 7.6) for the radar, and -1.9 (-13.1 to 9.2) for the IPG. The extent to which the monitors guided adequate or led to inadequate treatment decisions (determined by Clarke Error Grid analysis) differed significantly between the monitors (P = 0.011). Decisions were correct 96% of the time for acoustic, 95% of the time for radar, and 94% of the time for IPG monitoring devices.ConclusionsNone of the studied devices (acoustic, IPG, and radar monitor) had LoA that were within our predefined (based on clinical judgement) limits of 2 breaths min( -1). The acoustic breath sound monitor predicted the correct treatment more often than the IPG and the radar device.
引用
收藏
页码:1324 / 1332
页数:9
相关论文
共 33 条
  • [1] Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study; [Précision des technologies de surveillance continue de la fréquence respiratoire à distance destinées aux unités cliniques de basse intensité : étude observationnelle prospective]
    van Loon K.
    Peelen L.M.
    van de Vlasakker E.C.
    Kalkman C.J.
    van Wolfswinkel L.
    van Zaane B.
    Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2018, 65 (12): : 1324 - 1332
  • [2] Implementation of a continuous respiratory rate monitor intended for low-resource settings: a novel device pilot study
    Amin, A. A.
    Mtenthaonga, P.
    Coyle, J.
    Mjumira, R.
    Mangwiro, L.
    Oden, M.
    Richards-Kortum, R.
    Dube, Q.
    Kawaza, K.
    Patil, M.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2025, 369 : S285 - S286
  • [3] Evaluating a Remote Monitoring Program for Respiratory Diseases: Prospective Observational Study
    Althobiani, Malik A.
    Ranjan, Yatharth
    Jacob, Joseph
    Orini, Michele
    Dobson, Richard James Butler
    Porter, Joanna C.
    Hurst, John R.
    Folarin, Amos A.
    JMIR FORMATIVE RESEARCH, 2023, 7
  • [4] Continuous Monitoring of Heart Rate Variability and Respiration for the Remote Diagnosis of Chronic Obstructive Pulmonary Disease: Prospective Observational Study
    Chen, Xiaolan
    Zhang, Han
    Li, Zhiwen
    Liu, Shuang
    Zhou, Yuqi
    JMIR MHEALTH AND UHEALTH, 2024, 12
  • [5] Contactless and continuous monitoring of respiratory rate in a hospital ward: a clinical validation study
    Toften, Stale
    Kjellstadli, Jonas T.
    Kvaerness, Jorn
    Pedersen, Line
    Laugsand, Lars E.
    Thu, Ole K. F.
    FRONTIERS IN PHYSIOLOGY, 2024, 15
  • [6] Evaluation of respiratory rate monitoring using a microwave Doppler sensor mounted on the ceiling of an intensive care unit: a prospective observational study
    Tanaka, Hiroyuki
    Yokose, Masashi
    Takaki, Shunsuke
    Mihara, Takahiro
    Saigusa, Yusuke
    Goto, Takahisa
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (01) : 71 - 79
  • [7] Evaluation of respiratory rate monitoring using a microwave Doppler sensor mounted on the ceiling of an intensive care unit: a prospective observational study
    Hiroyuki Tanaka
    Masashi Yokose
    Shunsuke Takaki
    Takahiro Mihara
    Yusuke Saigusa
    Takahisa Goto
    Journal of Clinical Monitoring and Computing, 2022, 36 : 71 - 79
  • [8] Interest of continuous glucose monitoring in surgical intensive care unit patients: a prospective observational study
    Alluin, A.
    Genay, S.
    Ethgen, S.
    Decaudin, B.
    Odou, P.
    Lebuffe, G.
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2013, 32 : A18 - A19
  • [9] Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study
    Masashi Takakura
    Tasuku Fujii
    Tomoya Taniguchi
    Shogo Suzuki
    Kimitoshi Nishiwaki
    Journal of Anesthesia, 2023, 37 : 394 - 400
  • [10] Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study
    Takakura, Masashi
    Fujii, Tasuku
    Taniguchi, Tomoya
    Suzuki, Shogo
    Nishiwaki, Kimitoshi
    JOURNAL OF ANESTHESIA, 2023, 37 (03) : 394 - 400