Locking Plates With Computationally Enhanced Screw Trajectories Provide Superior Biomechanical Fixation Stability of Complex Proximal Humerus Fractures

被引:4
|
作者
Mischler, Dominic [1 ]
Schader, Jana Felicitas [1 ]
Dauwe, Jan [1 ,2 ]
Tenisch, Lara [1 ]
Gueorguiev, Boyko [1 ]
Windolf, Markus [1 ]
Varga, Peter [1 ]
机构
[1] AO Res Inst Davos, Davos, Switzerland
[2] UZ Leuven, Dept Trauma Surg, Leuven, Belgium
关键词
bone fracture (MeSH ID; D050723); osteosynthesis; implant optimization; 3D-printing; finite element analysis; fixation failure; ANGULAR STABLE PLATE; NONOPERATIVE TREATMENT; INTERNAL-FIXATION; ELDERLY-PATIENTS; PERFORATION; STIFFNESS; SYSTEMS; ADULTS; LOAD;
D O I
10.3389/fbioe.2022.919721
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Joint-preserving surgical treatment of complex unstable proximal humerus fractures remains challenging, with high failure rates even following state-of-the-art locked plating. Enhancement of implants could help improve outcomes. By overcoming limitations of conventional biomechanical testing, finite element (FE) analysis enables design optimization but requires stringent validation. This study aimed to computationally enhance the design of an existing locking plate to provide superior fixation stability and evaluate the benefit experimentally in a matched-pair fashion. Further aims were the evaluation of instrumentation accuracy and its potential influence on the specimen-specific predictive ability of FE. Screw trajectories of an existing commercial plate were adjusted to reduce the predicted cyclic cut-out failure risk and define the enhanced (EH) implant design based on results of a previous parametric FE study using 19 left proximal humerus models (Set A). Superiority of EH versus the original (OG) design was tested using nine pairs of human proximal humeri (N = 18, Set B). Specimen-specific CT-based virtual preoperative planning defined osteotomies replicating a complex 3-part fracture and fixation with a locking plate using six screws. Bone specimens were prepared, osteotomized and instrumented according to the preoperative plan via a standardized procedure utilizing 3D-printed guides. Cut-out failure of OG and EH implant designs was compared in paired groups with both FE analysis and cyclic biomechanical testing. The computationally enhanced implant configuration achieved significantly more cycles to cut-out failure compared to the standard OG design (p < 0.01), confirming the significantly lower peri-implant bone strain predicted by FE for the EH versus OG groups (p < 0.001). The magnitude of instrumentation inaccuracies was small but had a significant effect on the predicted failure risk (p < 0.01). The sample-specific FE predictions strongly correlated with the experimental results (R-2 = 0.70) when incorporating instrumentation inaccuracies. These findings demonstrate the power and validity of FE simulations in improving implant designs towards superior fixation stability of proximal humerus fractures. Computational optimization could be performed involving further implant features and help decrease failure rates. The results underline the importance of accurate surgical execution of implant fixations and the need for high consistency in validation studies.
引用
收藏
页数:11
相关论文
共 49 条
  • [41] Biomechanical effect of medial cortical support and medial screw support on locking plate fixation in proximal humeral fractures with a medial gap: a finite element analysis
    Yang, Pan
    Zhang, Ying
    Liu, Jian
    Xiao, Jin
    Ma, Li Min
    Zhu, Chang Rong
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2015, 49 (02) : 203 - 209
  • [42] Primary stability in locking plate fixation for proximal humeral fractures may be increased by using an additional nail osteosynthesis in combination with locking plate osteosynthesis-A biomechanical comparison
    Schoebel, Tobias
    Gemkow, Maximilian
    Wendler, Toni
    Schleifenbaum, Stefan
    Loeffler, Sabine
    Theopold, Jan
    Hepp, Pierre
    CLINICAL BIOMECHANICS, 2024, 114
  • [43] Correct positioning of the calcar screw leads to superior results in proximal humerus fractures treated with carbon-fibre-reinforced polyetheretherketone plate osteosynthesis with polyaxial locking screws
    Kimmeyer, Michael
    Schmalzl, Jonas
    Rentschler, Verena
    Schieffer, Christian
    Macken, Arno
    Gerhardt, Christian
    Lehmann, Lars-Johannes
    JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2023, 24 (01)
  • [44] Correct positioning of the calcar screw leads to superior results in proximal humerus fractures treated with carbon-fibre-reinforced polyetheretherketone plate osteosynthesis with polyaxial locking screws
    Michael Kimmeyer
    Jonas Schmalzl
    Verena Rentschler
    Christian Schieffer
    Arno Macken
    Christian Gerhardt
    Lars-Johannes Lehmann
    Journal of Orthopaedics and Traumatology, 24
  • [45] Fixation of 4-part fractures of the proximal humerus: Can we identify radiological criteria that support locking plates or IM nailing? Comparative, retrospective study of 107 cases
    Gadea, F.
    Favard, L.
    Boileau, P.
    Cuny, C.
    d'Ollone, T.
    Saragaglia, D.
    Sirveaux, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2016, 102 (08) : 963 - 970
  • [46] Internal Fixation of 4-Part Proximal Humerus Fractures and Fracture Dislocations Using Extended Deltoid-splitting Approach, Locking Plates, and Calcium Sulfate Graft: Technical Details
    Zadeh, Hamid G.
    Somasundaram, Karthikeyan
    Babu, Victor
    Movahedi, Roya
    TECHNIQUES IN SHOULDER AND ELBOW SURGERY, 2018, 19 (03): : 129 - 137
  • [47] Enhanced Biomechanical Stability in Proximal Humeral Fractures: Finite Element Analysis of a Novel Endosteal Anatomical Support Nail for Improved Fixation in Elderly Patients
    Chen, Jiawen
    Wang, Zhonghe
    Li, Changda
    He, Peiyu
    Chen, Zhongxuan
    Sun, Lijun
    Cao, Xiaoyan
    Tian, Na
    Dong, Xiang
    Tang, Peifu
    Chen, Hua
    ORTHOPAEDIC SURGERY, 2025, 17 (02) : 551 - 562
  • [48] The necessity of strength evaluation in assessment of clinical outcome after shoulder surgery : follow-up data from patients with complex proximal humerus fractures treated by locking plate fixation
    Wu, Y.
    Shang, P.
    Che, L.
    Ye, T.
    Wang, L.
    Qiu, S.
    ACTA ORTHOPAEDICA BELGICA, 2016, 82 (02): : 189 - 196
  • [49] Locking Reconstruction Double Plating of Distal Humeral Fractures: How Many Screws in the Distal Ulnar Column Segment in A3 Fracture Provide Superior Stability? A Comparative Biomechanical In Vitro Study
    Mehling, Isabella
    Schmidt-Horlohe, Kay
    Mueller, Lars Peter
    Sternstein, Werner
    Korner, Jan
    Rommens, Pol Maria
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (08) : 581 - 587