Discrimination of benign and malignant microcalcifications using dual-energy imaging

被引:2
|
作者
Kim, H. [1 ]
Lee, M. [2 ]
Kim, D. [2 ]
Kim, H. -J. [1 ,2 ]
机构
[1] Yonsei Univ, Coll Hlth Sci, Dept Radiol Sci, 1 Yonseidae Gil, Wonju 220710, Gangwon, South Korea
[2] Yonsei Univ, Coll Hlth Sci, Dept Radiat Convergence Engn, 1 Yonseidae Gil, Wonju 220710, Gangwon, South Korea
来源
JOURNAL OF INSTRUMENTATION | 2020年 / 15卷 / 02期
基金
新加坡国家研究基金会;
关键词
X-ray mammography and scinto- and MRI-mammography; X-ray radiography and digital radiography (DR); COUNTING SPECTRAL MAMMOGRAPHY; BREAST MICROCALCIFICATIONS; X-RAY; DETECTOR;
D O I
10.1088/1748-0221/15/02/C02033
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Screening mammography has a high detection efficiency for microcalcifications. In a mammogram, morphological characteristics are used to distinguish benign and potentially malignant changes. In this study, we propose a new method to distinguish malignant and benign microcalcifications using only mammographic images based on the dual-energy method. A photoncounting spectral mammography system was simulated using the Geant4 Application for Tomographic Emission (GATE) simulation tool. The dual-energy images were obtained using two energy bins. Microcalcifications of type I (calcium oxalate) and type II (calcium hydroxyapatite) were studied. For statistical analysis, the microcalcifications were classified as type I or type II based on a score calculated using the dual-energy images. The score values were calculated using the ratio of values at low energy and high energy because the attenuation difference was small in the high-energy region and large in the low-energy region. In other words, the classification of microcalcifications associated with pathogenesis was performed using the attenuation ratio as a discrimination criterion. For the appropriate dual-energy images, the effects on the energy spectra of microcalcifications were evaluated. We confirmed that the contrast and the noise were affected because the classification method used in this study is based on the pixel values of the images. In addition, we suggested the possibility of automatic classification for malignant microcalcifications using segmentation methods and the minimum and maximum thresholds of score values.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] QUANTIFICATION TECHNIQUES FOR DUAL-ENERGY CARDIAC IMAGING
    MOLLOI, SY
    MISTRETTA, CA
    MEDICAL PHYSICS, 1989, 16 (02) : 209 - 217
  • [42] Dual-Energy CT in Head and Neck Imaging
    Roele E.D.
    Timmer V.C.M.L.
    Vaassen L.A.A.
    van Kroonenburgh A.M.J.L.
    Postma A.A.
    Current Radiology Reports, 5 (5)
  • [43] Dual-Energy CT: Applications in Oncologic Imaging
    Schoepf, U.
    MEDICAL PHYSICS, 2016, 43 (06) : 3746 - 3746
  • [44] Imaging of nanoparticles with dual-energy computed tomography
    Ducote, J. L.
    Alivov, Y.
    Molloi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2031 - 2044
  • [45] AN INVERSE MATRIX APPROACH TO DUAL-ENERGY IMAGING
    SHAW, CG
    PLEWES, DP
    MEDICAL PHYSICS, 1986, 13 (04) : 577 - 577
  • [46] Dual-Energy MDCT for Imaging the Renal Mass
    Mileto, Achille
    Nelson, Rendon C.
    Paulson, Erik K.
    Marin, Daniele
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (06) : W640 - W647
  • [47] Dual-energy tissue cancellation in mammography for improved detection of microcalcifications and neoplasms: A phantom study
    Luzhbin, Dmytro
    Chang, Chieh-Chun
    Chang, Shu-Jun
    Wu, Jay
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1025
  • [48] Dual-Energy CT: Applications in Abdominal Imaging
    Bauer R.W.
    Fischer S.
    Current Radiology Reports, 3 (4)
  • [49] Numerical Simulation for Dual-energy CT Imaging
    Wang, Lixin
    Sun, Fengrong
    Babyn, Paul
    Zhong, Hai
    Zhou, Zhenzhen
    Huang, Hao
    Qin, Jiaojiao
    PROCEEDINGS OF THE 2015 2ND INTERNATIONAL CONFERENCE ON ELECTRICAL, COMPUTER ENGINEERING AND ELECTRONICS (ICECEE 2015), 2015, 24 : 316 - 320
  • [50] Dual-Energy Computed Tomography in Genitourinary Imaging
    Mileto, Achille
    Marin, Daniele
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2017, 55 (02) : 373 - +