Novel wearable and contactless heart rate, respiratory rate, and oxygen saturation monitoring devices: a systematic review and meta-analysis

被引:16
|
作者
Chan, P. Y. [1 ]
Ryan, N. P. [1 ]
Chen, D. [1 ]
McNeil, J. [2 ]
Hopper, I [2 ]
机构
[1] Eastern Hlth, Dept Intens Care Med, Melbourne, Vic, Australia
[2] Monash Univ, Sch Publ Hlth & Prevent Med, Melbourne, Vic, Australia
基金
英国医学研究理事会;
关键词
clinical monitoring; contactless devices; infrared; wearable devices; EMERGENCY-DEPARTMENT; ATRIAL-FIBRILLATION; ACOUSTIC METHOD; ACCURACY; VALIDATION; PHOTOPLETHYSMOGRAPHY; CAPNOGRAPHY; ANESTHESIA; AGREEMENT; PRESSURE;
D O I
10.1111/anae.15834
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We performed a systematic review and meta-analysis to identify, classify and evaluate the body of evidence on novel wearable and contactless devices that measure heart rate, respiratory rate and oxygen saturations in the clinical setting. We included any studies of hospital inpatients, including sleep study clinics. Eighty-four studies were included in the final review. There were 56 studies of wearable devices and 29 of contactless devices. One study assessed both types of device. A high risk of patient selection and rater bias was present in proportionally more studies assessing contactless devices compared with studies assessing wearable devices (p = 0.023 and p < 0.0001, respectively). There was high but equivalent likelihood of blinding bias between the two types of studies (p = 0.076). Wearable device studies were commercially available devices validated in acute clinical settings by clinical staff and had more real-time data analysis (p = 0.04). Contactless devices were more experimental, and data were analysed post-hoc. Pooled estimates of mean (95%CI) heart rate and respiratory rate bias in wearable devices were 1.25 (-0.31-2.82) beats.min(-1) (pooled 95% limits of agreement -9.36-10.08) and 0.68 (0.05-1.32) breaths.min(-1) (pooled 95% limits of agreement -5.65-6.85). The pooled estimate for mean (95%CI) heart rate and respiratory rate bias in contactless devices was 2.18 (3.31-7.66) beats.min(-1) (pooled limits of agreement -6.71-10.88) and 0.30 (-0.26-0.87) breaths.min(-1) (pooled 95% limits of agreement -3.94-4.29). Only two studies of wearable devices measured SpO2; these reported mean measurement biases of 3.54% (limits of agreement -5.65-11.45%) and 2.9% (-7.4-1.7%). Heterogeneity was observed across studies, but absent when devices were grouped by measurement modality and reference standard. We conclude that, while studies of wearable devices were of slightly better quality than contactless devices, in general all studies of novel devices were of low quality, with small (< 100) patient datasets, typically not blinded and often using inappropriate statistical techniques. Both types of devices were statistically equivalent in accuracy and precision, but wearable devices demonstrated less measurement bias and more precision at extreme vital signs. The statistical variability in precision and accuracy between studies is partially explained by differences in reference standards.
引用
收藏
页码:1268 / 1280
页数:13
相关论文
共 50 条
  • [41] Evaluation of a Camera-Based Monitoring Solution Against Regulated Medical Devices to Measure Heart Rate, Respiratory Rate, Oxygen Saturation, and Blood Pressure
    Talukdar, Debjyoti
    De Deus, Luis Felipe
    Sehgal, Nikhil
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (11)
  • [42] Accuracy of Heart Rate and Respiratory Rate Measurements Using Two Types of Wearable Devices
    Takahashi, Yusuke
    Okura, Kazuki
    Minakata, Shin
    Watanabe, Motoyuki
    Hatakeyama, Kazutoshi
    Chida, Satoaki
    Saito, Kimio
    Matsunaga, Toshiki
    Shimada, Yoichi
    PROGRESS IN REHABILITATION MEDICINE, 2022, 7
  • [43] Novel wearable and contactless monitoring devices to identify deteriorating patients in the clinical setting: a systematic review protocol
    Chan, Peter Y.
    McNeil, John
    Nguyen, Tam
    Ryan, Nicholas
    Hopper, Ingrid
    SYSTEMATIC REVIEWS, 2020, 9 (01)
  • [44] Novel wearable and contactless monitoring devices to identify deteriorating patients in the clinical setting: a systematic review protocol
    Peter Y. Chan
    John McNeil
    Tam Nguyen
    Nicholas Ryan
    Ingrid Hopper
    Systematic Reviews, 9
  • [45] An Adaptive Heart Rate Monitoring Algorithm for Wearable Healthcare Devices
    Kuo, Tz-Hau
    Teng, Chi-Min
    Wu, Ming-Feng
    Wen, Chih-Yu
    ELECTRONICS, 2021, 10 (17)
  • [46] THE EFFECT OF SKIN TONE ON ACCURACY OF HEART RATE MEASUREMENT IN WEARABLE DEVICES: A SYSTEMATIC REVIEW
    Koerber, Daniel
    Khan, Shawn
    Shamsheri, Tahmina
    Kirubarajan, Abirami
    Mehta, Sangeeta
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2022, 79 (09) : 1990 - 1990
  • [47] Contactless Camera-Based Heart Rate and Respiratory Rate Monitoring Using AI on Hardware
    Kolosov, Dimitrios
    Kelefouras, Vasilios
    Kourtessis, Pandelis
    Mporas, Iosif
    SENSORS, 2023, 23 (09)
  • [48] Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate
    Monnink, Stefan H. J.
    van Vliet, Mariska
    Kuiper, Mathijs J.
    Constandse, Jan C.
    Hoftijzer, Dieke
    Muller, Marjolein
    Ronner, Eelko
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2024, : 451 - 457
  • [49] Video-Based Physiologic Monitoring During an Acute Hypoxic Challenge: Heart Rate, Respiratory Rate, and Oxygen Saturation
    Addison, Paul S.
    Jacquel, Dominique
    Foo, David M. H.
    Antunes, Andre
    Borg, Ulf R.
    ANESTHESIA AND ANALGESIA, 2017, 125 (03): : 860 - 873
  • [50] Clinical applications of contactless photoplethysmography for monitoring in adults: A systematic review and meta-analysis
    Bautista, Melissa Joanne
    Kowal, Mikolaj
    Cave, Daniel G. W.
    Downey, Candice
    Jayne, David G.
    JOURNAL OF CLINICAL AND TRANSLATIONAL SCIENCE, 2023, 7 (01)