Fourier rebinning of time-of-flight PET data

被引:56
|
作者
Defrise, M [1 ]
Casey, ME
Michel, C
Conti, M
机构
[1] Vrije Univ Brussels, Dept Nucl Med, B-1090 Brussels, Belgium
[2] CPS Innovat, Knoxville, TN 37392 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 12期
关键词
D O I
10.1088/0031-9155/50/12/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper explores fast reconstruction strategies for 3D time-of-flight (TOF) positron emission tomography (PET), based on 2D data rebinning. Starting from pre-corrected 3D TOF data, a rebinning algorithm estimates for each transaxial slice the 2D TOF sinogram that would have been acquired by a single-ring scanner. The rebinned sinograms can then be reconstructed using any algorithm for 2D TOF reconstruction. We introduce TOF-FORE, an approximate rebinning algorithm obtained by extending the Fourier rebinning method for non-TOF data. In addition, we identify two partial differential equations that must be satisfied by consistent 3D TOF data, and use them to derive exact rebinning algorithms and to characterize the degree of the approximation in TOF-FORE. Numerical simulations demonstrate that TOF-FORE is more accurate than two different TOF extensions of the single-slice rebinning method, and suggest that TOF-FORE will be a valuable tool for practical TOF PET in the range of axial apertures and time resolutions typical of current scanners.
引用
收藏
页码:2749 / 2763
页数:15
相关论文
共 50 条
  • [1] Evaluation of Fourier rebinning of time-of-flight PET data to non time-of-flight
    Li, Quanzheng
    Cho, S.
    Ahn, S.
    Ren, R.
    Leahy, R.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [2] Axial Fourier Rebinning for Time-of-Flight PET
    Li, Yusheng
    Matej, Samuel
    Metzler, Scott D.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [3] Exact and approximate Fourier rebinning of PET data from time-of-flight to non time-of-flight
    Cho, Sanghee
    Ahn, Sangtae
    Li, Quanzheng
    Leahy, Richard M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03): : 467 - 484
  • [4] Optimal Rebinning of Time-of-Flight PET Data
    Ahn, Sangtae
    Cho, Sanghee
    Li, Quanzheng
    Lin, Yanguang
    Leahy, Richard M.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2011, 30 (10) : 1808 - 1818
  • [5] FORCE: Fourier Rebinning and Consistency Equations for Time-of-Flight PET
    Li, Yusheng
    Matej, Samuel
    Metzler, Scott D.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [6] Fourier rebinning and consistency equations for time-of-flight PET planograms
    Li, Yusheng
    Defrise, Michel
    Matej, Samuel
    Metzler, Scott D.
    INVERSE PROBLEMS, 2016, 32 (09)
  • [7] Continuous and discrete data rebinning in time-of-flight PET
    Defrise, Michel
    Panin, Vladimir
    Michel, Christian
    Casey, Michael E.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (09) : 1310 - 1322
  • [8] Optimized Weighting for Fourier Rebinning of Three-Dimensional Time-of-Flight PET Data to Non-Time-of-Flight
    Ahn, Sangtae
    Cho, Sanghee
    Li, Quanzheng
    Leahy, Richard M.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2989 - 2996
  • [9] Discrete axial rebinning for time-of-flight PET
    Defrise, Michel
    Panin, Vladimir
    Michel, Christian
    Casey, Michael E.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2370 - 2374
  • [10] A unified Fourier theory for time-of-flight PET data
    Li, Yusheng
    Matej, Samuel
    Metzler, Scott D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02): : 601 - 624