Development and validation of analytical methodology for near-infrared conformance testing of pharmaceutical intermediates

被引:13
|
作者
Forbes, RA
Persinger, ML
Smith, DR
机构
[1] Analytical Development, Lilly Research Laboratories, A Division of Eli Lilly and Company, Clinton
关键词
pharmaceutical analysis; near-infrared spectroscopy; method validation; loracarbef; conformity index; conformance testing; wavelength distance method;
D O I
10.1016/S0731-7085(96)01875-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An approach was developed to create and validate analytical methods to perform near-infrared (NIR) conformance testing on two isolated intermediates used in the manufacture of loracarbef monohydrate bulk drug substance. Method calibration sets were developed from second-derivative NIR reflectance spectra for 30 representative batches of each intermediate. In conformance testing, second-derivative NIR spectra for samples from newly manufactured batches are compared with the calibration set. If the new spectrum is not statistically different to the average of the calibration set, the sample passes the conformance test. Using authentic batch samples of typical and low-potency lots, the methods were validated for accuracy, selectivity, ruggedness and repeatability of the methods.
引用
收藏
页码:315 / 327
页数:13
相关论文
共 50 条
  • [21] PHARMACEUTICAL FERMENTATION MONITORING BY NEAR-INFRARED SPECTROMETRY
    LANGE, AJ
    SINCLAIR, SL
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1994, 207 : 102 - BIOT
  • [22] Development and Analytical Validation of Robust Near-Infrared Multivariate Calibration Models for the Quality Inspection Control of Mozzarella Cheese
    Botelho, Bruno G.
    Mendes, Bruna A. P.
    Sena, Marcelo M.
    FOOD ANALYTICAL METHODS, 2013, 6 (03) : 881 - 891
  • [23] Development and Analytical Validation of Robust Near-Infrared Multivariate Calibration Models for the Quality Inspection Control of Mozzarella Cheese
    Bruno G. Botelho
    Bruna A. P. Mendes
    Marcelo M. Sena
    Food Analytical Methods, 2013, 6 : 881 - 891
  • [24] VALIDATION PROCEDURES IN NEAR-INFRARED SPECTROMETRY
    FORINA, M
    DRAVA, G
    BOGGIA, R
    LANTERI, S
    CONTI, P
    ANALYTICA CHIMICA ACTA, 1994, 295 (1-2) : 109 - 118
  • [25] VALIDATION OF NEAR-INFRARED SPECTROSCOPY IN MAN
    MANCINI, DM
    FARRELL, L
    WILSON, JR
    CIRCULATION, 1992, 86 (04) : 401 - 401
  • [26] VALIDATION OF NEAR-INFRARED SPECTROSCOPY IN HUMANS
    MANCINI, DM
    BOLINGER, L
    LI, H
    KENDRICK, K
    CHANCE, B
    WILSON, JR
    JOURNAL OF APPLIED PHYSIOLOGY, 1994, 77 (06) : 2740 - 2747
  • [27] Development and Validation of a Sensor Prototype for Near-Infrared Imaging of the Newborn Brain
    Ahnen, Linda
    Stachel, Helene
    Kleiser, Stefan
    Hagmann, Cornelia
    Jiang, Jingjing
    Kalyanov, Alexander
    Lindner, Scott
    Wolf, Martin
    Sanchez, Salvador
    OXYGEN TRANSPORT TO TISSUE XXXIX, 2017, 977 : 163 - 168
  • [28] Development and Validation of a Near-Infrared Optical System for Tracking Surgical Instruments
    Lin, Qinyong
    Cai, Ken
    Yang, Rongqian
    Chen, Huazhou
    Wang, Zhigang
    Zhou, Jing
    JOURNAL OF MEDICAL SYSTEMS, 2016, 40 (04) : 1 - 14
  • [29] Development and Validation of a Near-Infrared Optical System for Tracking Surgical Instruments
    Qinyong Lin
    Ken Cai
    Rongqian Yang
    Huazhou Chen
    Zhigang Wang
    Jing Zhou
    Journal of Medical Systems, 2016, 40
  • [30] VALIDATION OF THE NEAR-INFRARED SPECTROPHOTOMETRIC METHODOLOGY FOR MEASURING CEREBRAL BLOOD-VOLUME CHANGES
    STIFF, J
    WILSON, DA
    HANLEY, DF
    TRAYSTMAN, RJ
    ROGERS, MC
    PEDIATRIC RESEARCH, 1985, 19 (04) : A147 - A147