Functional results and complications of artificial urinary sphincter AMS 800: About 84 cases

被引:4
|
作者
Beaujon, N. [1 ]
Marcelli, F. [1 ]
Fantoni, J. -C. [1 ]
Biserte, J. [1 ]
机构
[1] Hop Claude Huriez, Serv Urol, F-59037 Lille, France
来源
PROGRES EN UROLOGIE | 2011年 / 21卷 / 03期
关键词
Urinary Artificial; Sphincter; Urinary Incontinence; Quality of Life; LONG-TERM; POSTPROSTATECTOMY INCONTINENCE; PATIENT SATISFACTION; IMPLANTATION; PROSTATECTOMY; MANAGEMENT; CONTINENCE;
D O I
10.1016/j.purol.2010.06.003
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose. - To verify the effectiveness and safetiness of artificial urinary sphincter (SUA) AMS 800 (TM) in the management of urinary incontinence by sphincter incompetence. Patients and methods. - Retrospective study of patients who received SUA by the same operator between 1992 and 2006. Fifty-seven men and 27 women whose average age was 61 years (22-82) suffered from incontinence by sphincter incompetence: after radical prostatectomy (46%), primary stress or mixed incontinence (21.4%), transurethral resection of prostate (9.5%), injury and neurological malformations (9.5%), pelvic or urethral trauma (7.1%), rectal surgery (3.6%), adenomectomy (2.4%). Functional assessment was made by telephone based on a 23-item questionnaire (Appendix 2). Results. - The mean follow-up was 52.6 months (5.3-187.1): 64% had their original SUA, 50 re-interventions were necessary for 30 patients with 13.1% mechanical complications, 6.6% urethral atrophy, and 13.9% infectious complications, 2.5% failures and 9% definitive explant. The half-time survival without revision was 8.3 years. Sixty-seven percent of patients were contacted for the functional assessment: 77% success rate (<= 1 protection), 85% of patients improved, 87% satisfied or very satisfied and 94% would go again through the same operation. Conclusion. - The artificial urinary sphincter AMS 800 (TM) remains still the reference in the management of urinary incontinence by sphincter incompetence in improving the quality of life of patients implanted at the cost of a significant revision rate and frequent residual leaks. (c) 2010 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:205 / 210
页数:6
相关论文
共 50 条
  • [31] Retrospective analysis of patients using an AMS-800 artificial urinary sphincter for stress urinary incontinence
    Marco, Saul Rico
    Ruiz, Jesus Martinez
    Martin, Miguel Segura
    Sanchez, Marta Victoria Lorenzo
    Jimenez, Sonsoles Navarro
    Honrubia, Maria Amalia Tarraga
    Martinez, Laura Domingo
    Sanchez, Antonio Santiago Salinas
    CONTINENCE, 2024, 11
  • [32] AMS-800 Artificial urinary sphincter in female patients with stress urinary incontinence: A systematic review
    Peyronnet, Benoit
    O'Connor, Eabhann
    Khavari, Rose
    Capon, Gregoire
    Manunta, Andrea
    Allue, Marta
    Hascoet, Juliette
    Nitti, Victor W.
    Game, Xavier
    Gilleran, Jason
    Castro-Sader, Luis
    Cornu, Jean-Nicolas
    Waltregny, David
    Ahyai, Sascha
    Chung, Eric
    Elliott, Daniel S.
    Fournier, Georges
    Brucker, Benjamin M.
    NEUROUROLOGY AND URODYNAMICS, 2019, 38 : S28 - S41
  • [33] ARTIFICIAL URINARY SPHINCTER-800
    BROWN, J
    MORALES, P
    UROLOGY, 1984, 23 (05) : 479 - 483
  • [34] THE AMS ARTIFICIAL URINARY SPHINCTER ON THE BULBOUS URETHRA
    NORDLING, J
    HOLMBENTZEN, M
    HALD, T
    SCANDINAVIAN JOURNAL OF UROLOGY AND NEPHROLOGY, 1986, 20 (03): : 165 - 167
  • [35] Asynchronous implantation of a penile prosthesis (AMS 700) in patients with an artificial urinary sphincter (AMS 800): What functional outcomes can we expect from the AMS 1500?
    Alechinsky, L.
    Phe, V.
    Yates, D. R.
    Bourgade, V.
    Parra, J.
    Richard, F.
    Chartier-Kastler, E.
    Bitker, M. -O.
    Roupret, M.
    PROGRES EN UROLOGIE, 2012, 22 (06): : 354 - 359
  • [36] INVALIDATION OF THE PRACTICE OF ADDING FLUID TO THE AMS 800 ARTIFICIAL URINARY SPHINCTER PRESSURE REGULATING BALLOON
    Koloff, Zachary
    Romo, Paholo Barboglio
    Yi, Yooni
    Malaeb, Bahaa
    JOURNAL OF UROLOGY, 2017, 197 (04): : E624 - E625
  • [37] Long-term experience with the double-cuff AMS 800 artificial urinary sphincter
    Kowalczyk, JJ
    Spicer, DL
    Mulcahy, JJ
    UROLOGY, 1996, 47 (06) : 895 - 897
  • [38] DEVICE SURVIVAL FOLLOWING PRIMARY IMPLANTATION OF THE AMS 800 ARTIFICIAL URINARY SPHINCTER FOR MALE STRESS URINARY INCONTINENCE
    Yafi, F. A.
    DeLay, K. J.
    Stewart, C.
    Chiang, J.
    Sangkum, P.
    Hellstrom, W. J.
    JOURNAL OF SEXUAL MEDICINE, 2017, 14 (02): : E3 - E4
  • [39] DEVICE SURVIVAL FOLLOWING PRIMARY IMPLANTATION OF THE AMS 800 ARTIFICIAL URINARY SPHINCTER FOR MALE STRESS URINARY INCONTINENCE
    Yafi, F.
    DeLay, K. J.
    Stewart, C.
    Chiang, J.
    Sangkum, P.
    Hellstrom, W. J. G.
    JOURNAL OF SEXUAL MEDICINE, 2017, 14 (01): : S52 - S52
  • [40] LONG-TERM FUNCTIONAL OUTCOMES AFTER ARTIFICIAL URINARY SPHINCTER (AMS 800®) IMPLANTATION IN FEMALE NEUROGENIC PATIENTS WITH STRESS URINARY INCONTINENCE
    Phe, V.
    Leon, P.
    Granger, B.
    Denys, P.
    Bitker, M.
    Roupret, M.
    Chartier-Kastler, E. J.
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2015, 26 : S91 - S93