Strategies for Passive Sensitivity Improvement of NDIR Ethylene Gas Detectors

被引:1
|
作者
Sklorz, Adam [1 ]
Lang, Walter [1 ]
机构
[1] Inst Microsensors Actuators & Syst IMSAS, D-28359 Bremen, Germany
来源
EUROSENSORS XXV | 2011年 / 25卷
关键词
Ethylene detector; ethylene sensor; NDIR gas measurement; sensitivity improvement; IR gas measurement;
D O I
10.1016/j.proeng.2011.12.284
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Three sensitivity improvement strategies for non-dispersive-infrared (NDIR) ethylene gas detectors were defined and examined: the application of low-cost Fresnel lenses, usage of a conically shaped measurement chamber and a simple digital signal processing lead to a seven times better IR-radiation efficiency compared to a system configuration without consideration of these methods. The approaches don't affect IR-detectors, IR-sources or the optical pathway and thus, the methods can be called "passive". It was shown that usage of lock-in amplifiers is not necessary to achieve good sensitivity results and noise equivalent concentrations of about 6 ppmv. (C) 2011 Published by Elsevier Ltd.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Detection limit improvement for NDIR ethylene gas detectors using passive approaches
    Sklorz, Adam
    Janssen, Steffen
    Lang, Walter
    SENSORS AND ACTUATORS B-CHEMICAL, 2012, 175 : 246 - 254
  • [2] Merging Ethylene NDIR Gas Sensors with Preconcentrator-Devices for Sensitivity Enhancement
    Sklorz, Adam
    Schaefer, Anika
    Lang, Walter
    EUROSENSORS XXIV CONFERENCE, 2010, 5 : 1192 - 1195
  • [3] Merging ethylene NDIR gas sensors with preconcentrator-devices for sensitivity enhancement
    Sklorz, Adam
    Schafer, Anika
    Lang, Walter
    SENSORS AND ACTUATORS B-CHEMICAL, 2012, 170 : 21 - 27
  • [4] High-sensitivity ethylene gas sensor based on NDIR and dual-channel lock-in amplifier
    Zhang, Yunbo
    Jiang, Pai
    Cao, Weijie
    Li, Xiaoping
    Lai, Jianjun
    OPTIK, 2020, 223
  • [5] GAS CHROMATOGRAPHY IN TOXICOLOGICAL ANALYSIS - IMPROVEMENT OF SELECTIVITY AND SENSITIVITY BY ADDITION OF MULTIPLE DETECTORS
    BRANDENB.H
    PHARMACEUTICA ACTA HELVETIAE, 1970, 45 (07): : 394 - &
  • [6] SENSITIVITY IMPROVEMENT FOR FAST GAAS PHOTOCONDUCTIVE DETECTORS
    PASCAL, D
    BRU, C
    HOBARBOUDEBOUS, F
    LAVAL, S
    PHYSICA B & C, 1985, 129 (1-3): : 493 - 496
  • [7] Novel pyroelectric single crystals PIN-PMN-PT and their applications for NDIR gas detectors
    Chen, Jianwei
    Zhu, Lili
    Zhang, Mengyuan
    Feng, Peigui
    Sun, Ruiyu
    Di, Wenning
    Lin, Di
    Jiao, Jie
    Luo, Haosu
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2024, 63 (01)
  • [8] On sensitivity of gas-discharge detectors to light
    ChirikovZorin, IE
    Pukhov, OE
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 371 (03): : 375 - 379
  • [9] EFFECT OF THE CARRIER GAS ON THE SENSITIVITY OF THERMAL CONDUCTIVITY DETECTORS IN GAS CHROMATOGRAPHY
    SCOTT, BA
    WILLIAMSON, AG
    NATURE, 1959, 183 (4671) : 1322 - 1323
  • [10] ERROR IN MEASURING RADON IN SOIL GAS BY MEANS OF PASSIVE DETECTORS
    TANNER, AB
    NUCLEAR GEOPHYSICS, 1991, 5 (1-2): : 25 - 30