Active-passive gradient shielding for MRI acoustic noise reduction

被引:52
|
作者
Edelstein, WA
Kidane, TK
Taracila, V
Baig, TN
Eagan, TP
Cheng, YCN
Brown, RW
Mallick, JA
机构
[1] MRSci LLC, Schenectady, NY 12309 USA
[2] Univ Texas, Inst Adv Technol, Austin, TX 78712 USA
关键词
acoustics; acoustic noise; acoustic noise reduction; noise reduction; MRI acoustic noise; MRI acoustic noise reduction; gradients; active shielding; actively shielded gradients; active-passive shielding;
D O I
10.1002/mrm.20472
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An important source of MRI acoustic noise-magnet cryostat warm-bore vibrations caused by eddy-current-induced forces-can be mitigated by a passive metal shield mounted on the outside of a vibration-isolated, vacuum-enclosed shielded gradient set. Finite-element (FE) calculations for a z-gradient indicate that a 2-mm-thick Cu layer wrapped on the gradient assembly can decrease mechanical power deposition in the warm bore and reduce warm-bore acoustic noise production by about 25 dB. Eliminating the conducting warm bore and other magnet parts as significant acoustic noise sources could lead to the development of truly quiet, fully functioning MRI systems with noise levels below 70 dB. (c) 2005 Wiley-Liss, Inc.
引用
收藏
页码:1013 / 1017
页数:5
相关论文
共 50 条
  • [1] Active-passive shielding for MRI acoustic noise reduction: Network analysis
    Kidane, Tesfaye K.
    Edelstein, William A.
    Eagan, Timothy P.
    Taracila, Victor
    Baig, Tanvir N.
    Cheng, Yu-Chung N.
    Brown, Robert W.
    IEEE TRANSACTIONS ON MAGNETICS, 2006, 42 (12) : 3854 - 3860
  • [2] Acoustic noise reduction for spiral MRI by gradient derating
    Zhou, Zeyu
    Alfayad, Abdulrahman
    Chao, Tzu Cheng
    Pipe, James G.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (04) : 1547 - 1554
  • [3] Smartphone Localization using Active-Passive Acoustic Sensing
    Murakami, Hiroaki
    Nakamura, Masanari
    Yamasaki, Shoma
    Hashizume, Hiromichi
    Sugimoto, Masanori
    2018 NINTH INTERNATIONAL CONFERENCE ON INDOOR POSITIONING AND INDOOR NAVIGATION (IPIN 2018), 2018,
  • [4] Active-passive thermotomographic systems with focusing of acoustic fields
    Burov V.A.
    Dmitriev K.V.
    Evtukhov S.N.
    Bulletin of the Russian Academy of Sciences: Physics, 2009, 73 (4) : 523 - 528
  • [5] Reduction of gradient acoustic noise in MRI using SENSE-EPI
    de Zwart, JA
    van Gelderen, P
    Kellman, P
    Duyn, JH
    NEUROIMAGE, 2002, 16 (04) : 1151 - 1155
  • [6] Active vibration control of gradient coils to reduce acoustic noise of MRI systems
    Roozen, N. B.
    Koevoets, A. H.
    den Hamer, A. J.
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2008, 13 (03) : 325 - 334
  • [7] ACTIVE-PASSIVE IMMUNIZATION
    不详
    NEW ENGLAND JOURNAL OF MEDICINE, 1950, 242 (03): : 108 - 109
  • [9] PHOTOCURRENTS ON PASSIVE AND ACTIVE-PASSIVE METALS
    BURLEIGH, TD
    CORROSION SCIENCE, 1990, 31 : 745 - 750
  • [10] ACTIVE ACOUSTIC SCREENING - REDUCTION OF NOISE IN GRADIENT COILS BY LORENTZ FORCE BALANCING
    MANSFIELD, P
    CHAPMAN, BLW
    BOWTELL, R
    GLOVER, P
    COXON, R
    HARVEY, PR
    MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (02) : 276 - 281