Comparison of the IAEA TRS-398 and AAPM TG-51 absorbed dose to water protocols in the dosimetry of high-energy photon and electron beams

被引:43
|
作者
Huq, MS
Andreo, P
Song, HJ
机构
[1] Thomas Jefferson Univ, Jefferson Med Coll, Kimmel Canc Ctr, Dept Radiat Oncol, Philadelphia, PA 19107 USA
[2] Univ Stockholm, Karolinska Inst, Dept Med Radiat Phys, SE-17176 Stockholm, Sweden
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2001年 / 46卷 / 11期
关键词
D O I
10.1088/0031-9155/46/11/315
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The International Atomic Energy Agency (IAEA TRS-398) and the American Association of Physicists in Medicine (AAPM TG-51) have published new protocols for the calibration of radiotherapy beams. These protocols are based on the use of an ionization chamber calibrated in terms of absorbed dose to water in a standards laboratory's reference quality beam. This paper compares the recommendations of the two protocols in two ways: (i) by analysing in detail the differences in the basic data included in the two protocols for photon and electron beam dosimetry and (ii) by performing measurements in clinical photon and electron beams and determining the absorbed dose to water following the recommendations of the two protocols. Measurements were made with two Farmer-type ionization chambers and three plane-parallel ionization chamber types in 6, 18 and 25 MV photon beams and 6, 8, 10, 12 15 and 18 MeV electron beams. The Farmer-type chambers used were NE 2571 and PTW 30001, and the plane-parallel chambers were a Scanditronix-Wellhofer NACP and Roos, and a PTW Markus chamber. For photon beams, the measured ratios TG-51/TRS-398 of absorbed dose to water D-w ranged between 0.997 and 1.001, with a mean value of 0.999. The ratios for the beam quality correction factors k(Q) were found to agree to within about +/-0.2% despite significant differences in the method of beam quality specification for photon beams and in the basic data entering into kQ. For electron beams, dose measurements were made using direct N-D,N-w calibrations of cylindrical and plane-parallel chambers in a Co-60 gamma-ray beam, as well as cross-calibrations of plane-parallel chambers in a high-energy electron beam. For the direct N-D,N-w calibrations the ratios TG-51/TRS-398 of absorbed dose to water D-w were found to lie between 0.994 and 1.018 depending upon the chamber and electron beam energy used, with mean values of 0.996, 1.006, and 1.017, respectively, for the cylindrical, well-guarded and not well-guarded plane-parallel chambers. The D-w ratios measured for the cross-calibration procedures varied between 0.993 and 0.997. The largest discrepancies for electron beams between the two protocols arise from the use of different data for the perturbation correction factors p(wall) and P-dis of cylindrical and plane-parallel chambers, all in Co-60. A detailed analysis of the reasons for the discrepancies is made which includes comparing the formalisms, correction factors and the quantities in the two protocols.
引用
收藏
页码:2985 / 3006
页数:22
相关论文
共 50 条
  • [41] NEW PROTOCOLS FOR THE DOSIMETRY OF HIGH-ENERGY PHOTON AND ELECTRON-BEAMS
    DUTREIX, A
    MIJNHEER, B
    SVENSSON, H
    RADIOTHERAPY AND ONCOLOGY, 1985, 4 (04) : 289 - 290
  • [42] COMPARISON OF THE NMIJ AND THE ARPANSA STANDARDS FOR ABSORBED DOSE TO WATER IN HIGH-ENERGY PHOTON BEAMS
    Shimizu, M.
    Morishita, Y.
    Kato, M.
    Tanaka, T.
    Kurosawa, T.
    Takata, N.
    Saito, N.
    Ramanathan, G.
    Harty, P. D.
    Oliver, C.
    Wright, T.
    Butler, D. J.
    RADIATION PROTECTION DOSIMETRY, 2015, 164 (03) : 181 - 186
  • [43] CONSISTENCY IN THE DETERMINATION OF ABSORBED DOSE TO WATER IN HIGH-ENERGY PHOTON BEAMS
    MIJNHEER, B
    MEDICAL PHYSICS, 1986, 13 (05) : 780 - 780
  • [44] A Module for Virtual Training: Calibration of High Energy Photon and Electron Beams According the IAEA Code of Practice TRS 398
    Hartmann, G. H.
    Becker, S.
    Brandes, A. R.
    Held, C.
    Kropp, D.
    Lehmann, M.
    Mayer, A.
    Weber, Y.
    Mahr, A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 12, 2009, 25 (12): : 159 - +
  • [45] Calibration of plane-parallel chambers following the IAEA protocol (TRS-398): comparison between calibrations in 60Co and in high energy electron beams.
    Lizuain, MC
    Linero, D
    Picon, C
    Munoz, C
    De Blas, R
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S89 - S89
  • [46] Absorbed dose to water based dosimetry versus air kerma based dosimetry for high-energy photon beams: an experimental study
    Palmans, H
    Nafaa, L
    De Jans, J
    Gillis, S
    Hoornaert, MT
    Martens, C
    Piessens, M
    Thierens, H
    Van der Plaetsen, A
    Vynckier, S
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (03): : 421 - 440
  • [47] IMPLICATIONS OF THE AAPM-TG-21 PROTOCOL FOR ROUTINE DOSIMETRY OF HIGH-ENERGY PHOTON AND ELECTRON-BEAM
    PAUL, JM
    KOCH, R
    PHILIP, PC
    MEDICAL PHYSICS, 1984, 11 (03) : 383 - 383
  • [50] THE NEW AAPM PROTOCOL FOR THE DETERMINATION OF DOSE FROM HIGH-ENERGY PHOTON AND ELECTRON-BEAMS
    STERNICK, ES
    SCHULZ, RJ
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (03): : 645 - 646