Deep learning-based attenuation correction in the absence of structural information for whole-body positron emission tomography imaging

被引:107
|
作者
Dong, Xue [1 ]
Lei, Yang [1 ]
Wang, Tonghe [1 ]
Higgins, Kristin [1 ,2 ]
Liu, Tian [1 ,2 ]
Curran, Walter J. [1 ,2 ]
Mao, Hui [2 ,3 ]
Nye, Jonathon A. [3 ]
Yang, Xiaofeng [1 ,2 ]
机构
[1] Emory Univ, Dept Radiat Oncol, Atlanta, GA 30322 USA
[2] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
[3] Emory Univ, Dept Radiol & Imaging Sci, Atlanta, GA 30322 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2020年 / 65卷 / 05期
基金
美国国家卫生研究院;
关键词
PET; attenuation correction; deep learning; QUANTITATIVE-EVALUATION; GUIDED ATTENUATION; PEDIATRIC CT; PET/MRI; PET; ATLAS; QUANTIFICATION; RECONSTRUCTION; SEGMENTATION; IMPACT;
D O I
10.1088/1361-6560/ab652c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Deriving accurate structural maps for attenuation correction (AC) of whole-body positron emission tomography (PET) remains challenging. Common problems include truncation, inter-scan motion, and erroneous transformation of structural voxel-intensities to PET mu-map values (e.g. modality artifacts, implanted devices, or contrast agents). This work presents a deep learning-based attenuation correction (DL-AC) method to generate attenuation corrected PET (AC PET) from non-attenuation corrected PET (NAC PET) images for whole-body PET imaging, without the use of structural information. 3D patch-based cycle-consistent generative adversarial networks (CycleGAN) is introduced to include NAC-PET-to-AC-PET mapping and inverse mapping from AC PET to NAC PET, which constrains NAC-PET-to-AC-PET mapping to be closer to one-to-one mapping. Since NAC PET images share similar anatomical structures to the AC PET image but lack contrast information, residual blocks, which aim to learn the differences between NAC PET and AC PET, are used to construct generators of CycleGAN. After training, patches from NAC PET images were fed into NAC-PET-to-AC-PET mapping to generate DL-AC PET patches. DL-AC PET image was then reconstructed through patch fusion. We conducted a retrospective study on 55 datasets of whole-body PET/CT scans to evaluate the proposed method. In comparing DL-AC PET with original AC PET, average mean error (ME) and normalized mean square error (NMSE) of the whole-body were 0.62% +/- 1.26% and 0.72% +/- 0.34%. The average intensity changes measured on sequential PET images with AC and DL-AC on both normal tissues and lesions differ less than 3%. There was no significant difference of the intensity changes between AC and DL-AC PET, which demonstrate DL-AC PET images generated by the proposed DL-AC method can reach a same level to that of original AC PET images. The method demonstrates excellent quantification accuracy and reliability and is applicable to PET data collected on a single PET scanner or hybrid platform with computed tomography (PET/CT) or magnetic resonance imaging (PET/MRI).
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Adaptive, segmented attenuation correction for whole-body PET imaging
    Xu, M
    Cutler, PD
    Luk, WK
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (01) : 331 - 336
  • [22] Comparison of Positron Emission Tomography Quantification Using Magnetic Resonance- and Computed Tomography-Based Attenuation Correction in Physiological Tissues and Lesions A Whole-Body Positron Emission Tomography/Magnetic Resonance Study in 66 Patients
    Seith, Ferdinand
    Gatidis, Sergios
    Schmidt, Holger
    Bezrukov, Ilja
    la Fougere, Christian
    Nikolaou, Konstantin
    Pfannenberg, Christina
    Schwenzer, Nina
    INVESTIGATIVE RADIOLOGY, 2016, 51 (01) : 66 - 71
  • [23] Frequency and etiology of solitary hot spots in the pelvis at whole-body positron emission tomography/computed tomography imaging
    Khademi, Sara
    Westphalen, Antonio C.
    Webb, Emily M.
    Joe, Bonnie N.
    Badiee, Shiva
    Hawkins, Randy A.
    Coakley, Fergus V.
    CLINICAL IMAGING, 2009, 33 (01) : 44 - 48
  • [24] Deep Learning-Based Positron Emission Tomography Molecular Imaging in the Assessment of Cognitive Dysfunction in Patients with Epilepsy
    Tuerxun, Mayila
    Yin, Lixin
    Chen, Huiqun
    Lin, Jingqian
    SCIENTIFIC PROGRAMMING, 2021, 2021 (2021)
  • [25] THE ROLE OF POSITRON EMISSION TOMOGRAPHY IN ONCOLOGY AND OTHER WHOLE-BODY APPLICATIONS
    HAWKINS, RA
    HOH, C
    GLASPY, J
    CHOI, Y
    DAHLBOM, M
    REGE, S
    MESSA, C
    NIETSZCHE, E
    HOFFMAN, E
    SEEGER, L
    MADDAHI, J
    PHELPS, ME
    SEMINARS IN NUCLEAR MEDICINE, 1992, 22 (04) : 268 - 284
  • [26] Whole-body tracking of single cells via positron emission tomography
    Jung, Kyung Oh
    Kim, Tae Jin
    Yu, Jung Ho
    Rhee, Siyeon
    Zhao, Wei
    Ha, Byunghang
    Red-Horse, Kristy
    Gambhir, Sanjiv Sam
    Pratx, Guillem
    NATURE BIOMEDICAL ENGINEERING, 2020, 4 (08) : 835 - 844
  • [27] Detection of lymphoma in bone marrow by whole-body positron emission tomography
    Carr, R
    Barrington, SF
    Madan, B
    O'Doherty, MJ
    Saunders, CAB
    van der Walt, J
    Timothy, AR
    BLOOD, 1998, 91 (09) : 3340 - 3346
  • [28] Whole-body tracking of single cells via positron emission tomography
    Kyung Oh Jung
    Tae Jin Kim
    Jung Ho Yu
    Siyeon Rhee
    Wei Zhao
    Byunghang Ha
    Kristy Red-Horse
    Sanjiv Sam Gambhir
    Guillem Pratx
    Nature Biomedical Engineering, 2020, 4 : 835 - 844
  • [29] Malignant melanoma staging using whole-body positron emission tomography
    Paquet, P
    Hustinx, R
    Rigo, P
    Pierard, GE
    MELANOMA RESEARCH, 1998, 8 (01) : 59 - 62
  • [30] SCATTER CORRECTION IN THE TRANSAXIAL SLICES OF A WHOLE-BODY POSITRON EMISSION TOMOGRAPH
    HOVERATH, H
    KUEBLER, WK
    OSTERTAG, HJ
    DOLL, J
    ZIEGLER, SI
    KNOPP, MV
    LORENZ, WJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06): : 717 - 728