Estimates of percentage body fat in young adolescents: a comparison of dual-energy X-ray absorptiometry and air displacement plethysmography

被引:27
|
作者
Radley, D
Gately, PJ
Cooke, CB
Carroll, S
Oldroyd, B
Truscott, JG
机构
[1] Leeds Metropolitan Univ, Sch Leisure & Sport, Leeds LS6 3QS, W Yorkshire, England
[2] Univ Leeds, Ctr Bone & Body Composit Res, Leeds LS2 9JT, W Yorkshire, England
[3] Univ Leeds, Sch Healthcare Studies, Imaging Sci Grp, Leeds LS2 9JT, W Yorkshire, England
关键词
air displacement plethysmography; dual-energy X-ray absorptiometry; body composition; children;
D O I
10.1038/sj.ejcn.1601702
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: To evaluate the accuracy of percentage body fat (% fat) estimates from air displacement plethysmography ( ADP) against an increasingly recognised criterion method, dual-energy X-ray absorptiometry (DXA), in young adolescents. Design: Cross-sectional evaluation. Setting: Leeds General Infirmary, Centre for Bone and Body Composition Research, Leeds, UK. Subjects: In all, 28 adolescents (12 males and 16 females), age (mean+/-s.d.) 14.9+/-0.5 y, body mass index 21.2+/-2.9 kg/m(2) and body fat (DXA) 24.2+/-10.2% were assessed. Results: ADP estimates of % fat were highly correlated with those of DXA in both male and female subjects (r = 0.84 - 0.95, all P<0.001; s.e.e. = 3.42 - 3.89%). Mean % fat estimated by ADP using the Siri (1961) equation (ADP(Siri)) produced a nonsignificant overestimation in males (0.67%), and a nonsignificant underestimation in females (1.26%). Mean % fat estimated by ADP using the Lohman (1986) equations (ADP(Loh)) produced a nonsignificant underestimation in males (0.90%) and a significant underestimation in females (3.29%; P<0.01). Agreement between ADP and DXA methods was examined using the total error (TE) and methods of Bland and Altman (1986). Males produced a smaller TE (ADP(Siri) 3.28%; ADP(Loh) 3.49%) than females (ADP(Siri) 3.81%; ADP(Loh) 4.98%). The 95% limits of agreement were relatively similar for all % fat estimates, ranging from +/-6.57 to +/-7.58%. Residual plot analyses, of the individual differences between ADP and DXA, revealed a significant bias associated with increased % fat ( DXA), only in girls (P<0.01). Conclusions: We conclude that ADP, at present, has unacceptably high limits of agreement compared to a criterion DXA measure. The ease of use, suitability for various populations and cost of ADP warrant further investigation of this method to establish biological variables that may influence the validity of body fat estimates.
引用
收藏
页码:1402 / 1410
页数:9
相关论文
共 50 条
  • [21] Comparing Bioelectrical Impedance Analysis, Air Displacement Plethysmography, and Dual-Energy X-Ray Absorptiometry for Body Composition in Pediatric Obesity
    Thajer, Alexandra
    Vasek, Martin
    Schneider, Sophie
    Kautzky-Willer, Alexandra
    Kainberger, Franz
    Durstberger, Sebastian
    Kranzl, Andreas
    Horsak, Brian
    Greber-Platzer, Susanne
    NUTRIENTS, 2025, 17 (06)
  • [22] Estimates of body composition with dual-energy X-ray absorptiometry in adults
    Li, Chaoyang
    Ford, Earl S.
    Zhao, Guixiang
    Balluz, Lina S.
    Giles, Wayne H.
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 2009, 90 (06): : 1457 - 1465
  • [23] Prediction of body fat in adolescents: comparison of two electric bioimpedance devices with dual-energy X-ray absorptiometry
    de Faria, Eliane Rodrigues
    de Faria, Franciane Rocha
    Santos Goncalves, Vivian Siqueira
    Castro Franceschini, Sylvia do Carmo
    Gouveia Peluzio, Maria do Carmo
    da Rocha Sant'Ana, Luciana Ferreira
    Priore, Silvia Eloiza
    NUTRICION HOSPITALARIA, 2014, 30 (06) : 1270 - 1278
  • [24] Comparison of air displacement plethysmography with dual-energy X-ray absorptiometry and 3 field methods for estimating body composition in middle-aged men
    Sardinha, LB
    Lohman, TG
    Teixeira, PJ
    Guedes, DP
    Going, SB
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1998, 68 (04): : 786 - 793
  • [25] Air Displacement Plethysmography, Dual-Energy X-ray Absorptiometry, and Total Body Water to Evaluate Body Composition in Preschool-Age Children
    Crook, Tina A.
    Armbya, Narain
    Cleves, Mario A.
    Badger, Thomas M.
    Andres, Aline
    JOURNAL OF THE ACADEMY OF NUTRITION AND DIETETICS, 2012, 112 (12) : 1993 - 1998
  • [26] A comparison of dual-energy X-ray absorptiometry, air displacement plethysmography and A-mode ultrasound toassess body composition in college-age adults
    Johnson, Kelly E.
    Miller, Brian
    Gibson, Ann L.
    McLain, Trisha A.
    Juvancic-Heltzel, Judith A.
    Kappler, Rachele M.
    Otterstetter, Ronald
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2017, 37 (06) : 646 - 654
  • [27] Validation of air displacement plethysmography with dual energy X-ray absorptiometry: Effect of clothing
    Hull, H
    Fields, D
    OBESITY RESEARCH, 2004, 12 : A118 - A118
  • [28] A comparison of associations of body mass index and dual-energy x-ray absorptiometry measured percentage fat and total fat with global serum metabolites in young women
    Dorgan, Joanne F.
    Ryan, Alice S.
    LeBlanc, Erin S.
    Van Horn, Linda
    Magder, Laurence S.
    Snetselaar, Linda G.
    Zhang, Yuji
    Dallal, Cher M.
    Jung, Seungyoun
    Shepherd, John A.
    OBESITY, 2023, 31 (02) : 525 - 536
  • [29] Estimation of percentage body fat by dual-energy x-ray absorptiometry: evaluation by in vivo human elemental composition
    Wang, ZiMian
    Heymsfield, Steven B.
    Chen, Zhao
    Zhu, Shankuan
    Pierson, Richard N.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (09): : 2619 - 2635
  • [30] A comparison of dual-energy X-ray absorptiometry and somatometrics for determining body fat in rhesus macaques
    Colman, RJ
    Hudson, JC
    Barden, HS
    Kemnitz, JW
    OBESITY RESEARCH, 1999, 7 (01): : 90 - 96