The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements

被引:42
|
作者
Brockett, Claire L. [1 ]
Harper, Phil [2 ]
Williams, Sophie [1 ]
Isaac, Graham H. [3 ]
Dwyer-Joyce, Rob S. [4 ]
Jin, Zhongmin [1 ]
Fisher, John [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Triboson Ltd, Sheffield, S Yorkshire, England
[3] Univ Leeds, Sch Mech Engn, DePuy Int & Inst Med & Biol Engn, Leeds, W Yorkshire, England
[4] Univ Sheffield, Dept Mech Engn, Tribol Grp, Sheffield, S Yorkshire, England
关键词
D O I
10.1007/s10856-007-3298-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Large diameter metal-on-metal bearings (MOM) are becoming increasingly popular, addressing the needs of young and more active patients. Clinical data has shown excellent short-to-mid-term results, though incidences of transient squeaking have been noted between implantation and up to 2 years post-operative. Geometric design features, such as clearance, have been significant in influencing the performance of the bearings. Sets of MOM bearings with different clearances were investigated in this study using a hip friction simulator to examine the influence of clearance on friction, lubrication and squeaking. The friction factor was found to be highest in the largest clearance bearings under all test conditions. The incidence of squeaking was also highest in the large clearance bearings, with all bearings in this group squeaking throughout the study. A very low incidence of squeaking was observed in the other two clearance groups. The measured lubricating film was found to be lowest in the large clearance bearings. This study suggests that increasing the bearing clearance results in reduced lubricant film thickness, increased friction and an increased incidence of squeaking.
引用
收藏
页码:1575 / 1579
页数:5
相关论文
共 50 条
  • [31] METAL-ON-METAL HIP REPLACEMENTS AND SUBCLINICAL EVIDENCE OF MYOCARDIAL DYSFUNCTION
    Lodge, Freya
    Khatun, Rabeya
    Fraser, Alan
    Yousef, Zaheer
    HEART, 2017, 103 : A11 - A12
  • [32] Large diameter metal-on-metal total hip arthroplasty for Crowe IV developmental dysplasia of the hip
    Parmaksizoglu, Atilla Sancar
    Ozkaya, Ufuk
    Bilgili, Fuat
    Basilgan, Seckin
    Kabukcuoglu, Yavuz
    HIP INTERNATIONAL, 2009, 19 (04) : 309 - 314
  • [33] Predictive wear modeling of the articulating metal-on-metal hip replacements
    Gao, Leiming
    Dowson, Duncan
    Hewson, Robert W.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (03) : 497 - 506
  • [34] Failure rates of stemmed metal-on-metal hip replacements reply
    Smith, Alison J.
    Dieppe, Paul
    Vernon, Kelly
    Porter, Martyn
    Blom, Ashley W.
    LANCET, 2012, 380 (9837): : 106 - 106
  • [35] Wear Simulation of Metal-on-Metal Hip Replacements With Frictional Contact
    Mattei, Lorenza
    Di Puccio, Francesca
    JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2013, 135 (02):
  • [36] Biotribocorrosion of metal-on-metal hip replacements: How surface degradation can influence metal ion formation
    Hesketh, James
    Meng, Qingen
    Dowson, Duncan
    Neville, Anne
    TRIBOLOGY INTERNATIONAL, 2013, 65 : 128 - 137
  • [37] A predictive analysis of long-term friction and wear characteristics of metal-on-metal total hip replacements
    Dowson, D.
    Wang, F. C.
    Wang, W. Z.
    Jin, Z. M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2007, 221 (J3) : 460 - 460
  • [38] Taper Design Affects Failure of Large-head Metal-on-metal Total Hip Replacements
    Nassif, Nader A.
    Nawabi, Danyal H.
    Stoner, Kirsten
    Elpers, Marcella
    Wright, Timothy
    Padgett, Douglas E.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (02) : 564 - 571
  • [39] Effect of 3D physiological loading and motion on elastohydrodynamic lubrication of metal-on-metal total hip replacements
    Gao, Leiming
    Wang, Fengcai
    Yang, Peiran
    Jin, Zhongmin
    MEDICAL ENGINEERING & PHYSICS, 2009, 31 (06) : 720 - 729
  • [40] Material Loss at the Taper Junction of Retrieved Large Head Metal-on-Metal Total Hip Replacements
    Matthies, Ashley K.
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Cro, Suzie
    Panagiotidou, Anna
    Blunn, Gordon
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (11) : 1677 - 1685