Commutability and interchangeability of commercial quality control materials with feline plasma for common biochemical analytes

被引:10
|
作者
Baral, Randolph M. [1 ,2 ]
Dhand, Navneet K. [2 ]
Freeman, Kathleen P. [3 ]
机构
[1] Paddington Cat Hosp, 210 Oxford St, Paddington, NSW 2021, Australia
[2] Univ Sydney, Fac Vet Sci, Sydney, NSW, Australia
[3] IDEXX Labs Ltd, Wetherby, England
关键词
Biochemistry analyzer; clinical chemistry; linear regression; Passing-Bablok; reference materials; residuals; REFERENCE INTERVALS; CLINICAL-CHEMISTRY; VALUES; SPECIFICATIONS; LABORATORIES; STRATEGIES; GUIDELINES; BREEDS;
D O I
10.1111/vcp.12357
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
BackgroundSpecies-specific plasma or serum pools are considered the ideal standard material for quality control materials (QCM) instead of commercially available human QCM. However, using plasma or serum pools is limited by volume restrictions, degradation over time, and a narrow range of analyte concentrations. Concentrations of QCM analytes should be consistent or commutable with those from species-specific plasma/serum samples, and the precision from plasma pools should be comparable or interchangeable with commercial human QCM. ObjectivesThe aims of this study were to determine the commutability and interchangeability of 2 levels of commercial QCM (MAS chemTRAK-H [CT]) with feline plasma pools (PP) from normal and renal disease cats measured using a commercial laboratory analyzer and a veterinary in-house analyzer. MethodsAgreement between the 2 analyzers was assessed for 16 analytes by correlation and Passing-Bablok regression analyses of feline plasma samples. The difference between each CT data point and the regression line (residuals) was determined and standardized, and CT were considered commutable' with PP if the standardized residual was within a range of -3 to 3. Coefficients of variation (CVs) for CT and PP for 16 analytes on 2 analyzers were compared by bootstrap analysis to determine interchangeability. ResultsMost CT analytes were within the range of patient plasma sample analytes, thus commutable. Only 2 analytes had equivalent precision for both levels of CT and both levels of PP, and 5 additional analytes had similar precision for at least one level of CT compared to at least one level of PP. ConclusionsThe QCM assessed is commutable to feline PP within the tested ranges for 2 particular analyzers. Commutability does not grant interchangeability.
引用
收藏
页码:300 / 310
页数:11
相关论文
共 37 条
  • [1] LACK OF COMMUTABILITY BETWEEN QUALITY CONTROL MATERIALS AND PLASMA SAMPLES IN A TROPONIN I ASSAY
    Candas-Estebanez, B.
    Cano-Corres, R.
    Dot-Bach, D.
    Valero-Politi, J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S713 - S713
  • [2] Standardization and harmonization in hematology: Instrument alignment, quality control materials, and commutability issue
    Vidali, Matteo
    Carobene, Anna
    Esposito, Sara Apassiti
    Napolitano, Gavino
    Caracciolo, Alessandra
    Seghezzi, Michela
    Previtali, Giulia
    Lippi, Giuseppe
    Buoro, Sabrina
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (03) : 364 - 371
  • [3] Standardization and harmonization in hematology: Instrument alignment, quality control materials, and commutability issue
    Vidali, Matteo
    Carobene, Anna
    Apassiti Esposito, Sara
    Napolitano, Gavino
    Caracciolo, Alessandra
    Seghezzi, Michela
    Previtali, Giulia
    Lippi, Giuseppe
    Buoro, Sabrina
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020,
  • [4] Fibrinogen determination according to Clauss: commutability assessment of International and commercial standards and quality control samples
    van den Besselaar, Antonius M. H. P.
    van Rijn, Claudia J. J.
    Cobbaert, Christa M.
    Reijnierse, G. Louis A.
    Hollestelle, Martine J.
    Niessen, Rene W. L. M.
    Hudig, Francisca
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (11) : 1761 - 1769
  • [5] Commutability of control materials for external quality assessment of serum apolipoprotein A-I measurement
    Zeng, Jie
    Qi, Tianqi
    Wang, Shu
    Zhang, Tianjiao
    Zhou, Weiyan
    Zhao, Haijian
    Ma, Rong
    Zhang, Jiangtao
    Yan, Ying
    Dong, Jun
    Zhang, Chuanbao
    Chen, Wenxiang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (05) : 789 - 795
  • [6] Commutability of reference and control materials:an essential factor for assuring the quality of measurements in Laboratory Medicine
    Braga, Federica
    Panteghini, Mauro
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (07) : 967 - 973
  • [7] Evaluating Commutability of Control Materials in Three Nordic External Quality Assessment Schemes for Lipoproteins
    Norling, Anna
    Bokman, C. Fredrik
    Goransson, Emma
    Wande, Birgitta
    Pedersen, Morten M.
    Berghall, Heidi
    Svala, Emilia
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2021, 6 (06): : 1580 - 1591
  • [8] Interchangeability of estimates of day-to-day imprecision between commercial control materials and serum pools
    Fuentes-Arderiu, X
    González-de-la-Presa, B
    CLINICAL CHEMISTRY, 2002, 48 (03) : 573 - 574
  • [9] Inter-method differences in the measurement of some specific plasma proteins: Commutability of control materials
    Arcangeli, L
    Cueroni, P
    Franzini, C
    Galletta, G
    Moschini, A
    Scapellato, L
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1996, 34 (01): : 37 - 42
  • [10] Lack of commutability between a quality control material and plasma samples in a troponin I measurement system
    Candas-Estebanez, Beatriz
    Cano-Corres, Ruth
    Dot-Bach, Dolors
    Valero-Politi, Jose