Evaluation of intra-aneurysmal hemodynamics after flow diverter placement in a patient-specific aneurysm model

被引:9
|
作者
Balasso, Andrea [1 ]
Bauer, Jan S. [1 ]
Liebig, Thomas [2 ]
Dorn, Franziska [2 ]
Zimmer, Claus [1 ]
Liepsch, Dieter [3 ]
Prothmann, Sascha [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Dept Neuroradiol, D-80290 Munich, Germany
[2] Univ Cologne, Dept Neuroradiol, D-50931 Cologne, Germany
[3] Munich Univ Appl Sci, Munich, Germany
关键词
Stent; flow diversion; endovascular; non-Newtonian fluid; laser Doppler anemometry; PIPELINE EMBOLIZATION DEVICE; INTRACRANIAL ANEURYSMS; ENDOVASCULAR TREATMENT; SACCULAR ANEURYSMS; STENT; ARTERY; RECONSTRUCTION; THROMBOSIS; RUPTURE; STRESS;
D O I
10.3233/BIR-14019
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
BACKGROUND: The growth and rupture of cerebral aneurysms is intrinsically related to the hemodynamics prevailing in the diseased area. Therefore, a better understanding of intra-aneurysmal hemodynamics is essential for developing effective treatment methods. OBJECTIVE: The intention of this study was to evaluate the intra-aneurysmal flow and flow reduction induced by flow diverters in a true-to-scale elastic aneurysm model, obtained from real patient data. METHODS: Based on the computed tomography angiography (CTA) data of a fusiform aneurysm of a 34 year old patient, an elastic silicon rubber model of the aneurysm was produced. A physiologic pulsatile flow was created with a circulatory experimental set-up, and a non-Newtonian perfusion fluid was used as a substitute for human blood. Hemodynamics were measured by LDA before and after flow diverter implantation. RESULTS: Implantation of a flow diverter device resulted in a reduction of intra-aneurysmal maximum flow velocities of 97.8% at the inflow zone, 89.1% in the dome and 89.3% at the outflow zone, when compared to the native model. A significant reduction of 94% in the mean intra-aneurysmal velocity was found. CONCLUSIONS: This promising methodology can optimize patient treatment and will correlate with computational simulations to evaluate their reliability.
引用
收藏
页码:341 / 354
页数:14
相关论文
共 50 条
  • [1] Intra-aneurysmal blood flow based on patient-specific CT angiogram
    Morino, Taku
    Tanoue, Tetsuya
    Tateshima, Satoshi
    Vinuela, Fernando
    Tanishita, Kazuo
    EXPERIMENTS IN FLUIDS, 2010, 49 (02) : 485 - 496
  • [2] Intra-aneurysmal blood flow based on patient-specific CT angiogram
    Taku Morino
    Tetsuya Tanoue
    Satoshi Tateshima
    Fernando Vinuela
    Kazuo Tanishita
    Experiments in Fluids, 2010, 49 : 485 - 496
  • [3] Stretchable, Implantable, Nanostructured Flow-Diverter System for Quantification of Intra-aneurysmal Hemodynamics
    Howe, Connor
    Mishra, Saswat
    Kim, Yun-Soung
    Chen, Yanfei
    Ye, Sang-Ho
    Wagner, William R.
    Jeong, Jae-Woong
    Byun, Hun-Soo
    Kim, Jong-Hoon
    Chun, Youngjae
    Yeo, Woon-Hong
    ACS NANO, 2018, 12 (08) : 8706 - 8716
  • [4] Intra-aneurysmal flow reductions in a thin film nitinol flow diverter
    Chun, Youngjae
    Hur, Soojung Claire
    Kealey, Colin P.
    Levi, Daniel S.
    Mohanchandra, K. P.
    Di Carlo, Dino
    Eldredge, Jeff D.
    Vinuela, Fernando
    Carman, Gregory P.
    SMART MATERIALS & STRUCTURES, 2011, 20 (05):
  • [5] Assessment of intra-aneurysmal flow modification after flow diverter stent placement with four-dimensional flow MRI: a feasibility study
    Pereira, Vitor Mendes
    Brina, Olivier
    Delattre, Benedicte M. A.
    Ouared, Rafik
    Bouillot, Pierre
    Erceg, Gorislav
    Schaller, Karl
    Lovblad, Karl-Olof
    Vargas, Maria-Isabel
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (12) : 913 - 919
  • [6] Delayed intra-aneurysmal migration of a flow diverter construct after treatment of a giant aneurysm of the cavernous internal carotid artery
    Gawlitza, Matthias
    Soize, Sebastien
    Manceau, Pierre-Francois
    Pierot, Laurent
    JOURNAL OF NEURORADIOLOGY, 2020, 47 (03) : 233 - 236
  • [7] Intra-aneurysmal hemodynamics: evaluation of pCONus and pCANvas bifurcation aneurysm devices using DSA optical flow imaging
    Perez, Marta Aguilar
    Henkes, Hans
    Bouillot, Pierre
    Brina, Olivier
    Slater, Lee-Anne
    Pereira, Vitor Mendes
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (11) : 1197 - 1201
  • [8] The influence of aneurysm configuration on intra-aneurysmal pressure and flow
    Sorteberg, A.
    Farhoudi, D.
    INTERVENTIONAL NEURORADIOLOGY, 2006, 12 (03) : 203 - 214
  • [9] The effect of aneurysm geometry on the intra-aneurysmal flow condition
    Satoshi Tateshima
    Aichi Chien
    James Sayre
    Juan Cebral
    Fernando Viñuela
    Neuroradiology, 2010, 52 : 1135 - 1141
  • [10] The effect of aneurysm geometry on the intra-aneurysmal flow condition
    Tateshima, Satoshi
    Chien, Aichi
    Sayre, James
    Cebral, Juan
    Vinuela, Fernando
    NEURORADIOLOGY, 2010, 52 (12) : 1135 - 1141