MRI thermometry in phantoms by use of the proton resonance frequency shift method: application to interstitial laser thermotherapy

被引:44
|
作者
Olsrud, J [1 ]
Wirestam, R
Brockstedt, S
Nilsson, AMK
Tranberg, KG
Stahlberg, F
Persson, BRR
机构
[1] Univ Lund Hosp, Dept Radiat Phys, SE-22185 Lund, Sweden
[2] Univ Lund Hosp, Dept Diagnost Radiol, SE-22185 Lund, Sweden
[3] Univ Lund Hosp, Dept Surg, SE-22185 Lund, Sweden
[4] Lund Inst Technol, Dept Phys, SE-22100 Lund, Sweden
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1998年 / 43卷 / 09期
关键词
D O I
10.1088/0031-9155/43/9/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work the temperature dependence of the proton resonance frequency was assessed in agarose gel with a high melting temperature (95 degrees C) and in porcine liver in vitro at temperatures relevant to thermotherapy (25-80 degrees C). Furthermore, an optically tissue-like agarose gel phantom was developed and evaluated for use in MRI. The phantom was used to visualize temperature distributions from a diffusing laser fibre by means of the proton resonance frequency shift method. An approximately linear relationship (0.0085 ppm degrees C-1) between proton resonance frequency shift and temperature change was found for agarose gel, whereas deviations from a linear relationship were observed for porcine liver. The optically tissue-like agarose gel allowed reliable MRI temperature monitoring, and the MR relaxation times (T-1 and T-2) and the optical properties were found to be independently alterable. Temperature distributions around a diffusing laser fibre, during irradiation and subsequent cooling, were assessed with high spatial resolution (voxel size = 4.3 mm(3)) and with random uncertainties ranging from 0.3 degrees C to 1.4 degrees C (1 SD) with a 40 s scan time.
引用
收藏
页码:2597 / 2613
页数:17
相关论文
共 45 条