Evaluation of Anti-dsDNA Antibody Tests: Crithidia luciliae Immunofluorescence Test, Immunoblot, Enzyme-linked Immunosorbent Assay, Chemiluminescence Immunoassay

被引:7
|
作者
Yang, Jin-young [1 ]
Oh, Eun-Jee [1 ]
Kim, Yonggoo [1 ]
Park, Yeon-Joon [1 ]
机构
[1] Catholic Univ Korea Sch Med, Dept Lab Med, Seoul, South Korea
来源
KOREAN JOURNAL OF LABORATORY MEDICINE | 2010年 / 30卷 / 06期
关键词
Anti-dsDNA; Crithidia luciliae immunofluorescence test; ELISA; Chemiluminescence immunoassay; Immunoblot assay; SYSTEMIC-LUPUS-ERYTHEMATOSUS; DOUBLE-STRANDED DNA; ANTINUCLEAR AUTOANTIBODIES; DEFINED SPECIFICITIES; RHEUMATIC DISEASES; LABORATORY ASSAYS; LINE IMMUNOASSAY; ELISA; DIAGNOSIS; CLASSIFICATION;
D O I
10.3343/kjlm.2010.30.6.675
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background : Anti-double stranded DNA antibody (anti-dsDNA) test is useful for the diagnosis and monitoring of systemic lupus erythematosus (SLE). Although several methods are available, none of them is completely satisfactory and differences among them have been reported. We evaluated the diagnostic performance of 6 commercial kits for anti-dsDNA detection. Methods : A total of 142 sera (SLE [N=74], other systemic rheumatic diseases [N=50], other diseases [N=18]) were tested by 6 different assay kits using different antigenic sources of DNA: Crithidia luciliae immunofluorescence test (CLIFF), salmon testes (immunoblot, IB), human (ELISA l), salmon testes with nucleosome linker (ELISA II), plasmid (ELISA III), and synthetic oligonucleotides (chemiluminescence immunoassay, CLIA). Results : With manufacturers' cut-off values, 6 test kits showed sensitivities of 55.4-91.9%. ELISA I had a greater sensitivity than the other five assays (P<0.001). The specificities of ELISA II, ELISA Ill, CLIA, and CLIFF were higher than those of ELISA I and IB (P<0.05). In ROC curve analysis, 3 ELISA kits and CLIA showed AUC values of 0.845-0.893, and revealed no significant differences among them (P>0.05). With cut-off values set at 95% of specificity, ELISA II had a higher sensitivity than ELISA III (63.5% vs. 41.9%, P<0.05). IB had poor concordance rates with other assays (42.0-65.0%). Pearson correlation coefficients among 4 quantitative assays were 0.667-0.798. Conclusions : Six different assays showed various performances depending on the methods and cut-off values used. Except IB, the other five assays can be used for the detection of anti-dsDNA. (Korean J Lab Med 2010;30:675-84)
引用
收藏
页码:675 / 684
页数:10
相关论文
共 50 条
  • [1] A comparison of the chemiluminescence immunoassay and Crithidia luciliae immunofluorescence test in detecting anti-dsDNA antibodies and assessing the activity of systemic lupus erythematosus
    He, Shiyi
    Wu, Xiaoning
    Li, Limin
    Jiang, Kongmei
    He, Qitian
    Xie, Li
    LUPUS, 2023, 32 (08) : 936 - 941
  • [2] Diagnostic Performance of Anti-dsDNA Tests by Indirect Immunofluorescence and Enzyme-linked Immunosorbent Assay in Patients with Systemic Lupus Erythematosus
    Nieto-Aristizabal, Ivana
    Espana, Laura Jackeline
    Ortega, Jorge
    Ruiz-Ordonez, Ingrid
    Vivas, Alvaro J.
    Castano, Gloria L.
    Castillo, Carmen
    Bautista-Vargas, Mario
    Suarez, Ana
    Canas, Carlos A.
    Tobon, Gabriel J.
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2023, 8 (04): : 713 - 725
  • [3] Comparisons of Anti-dsDNA Antibody Detection Methods by Chemiluminescent Immunoassay and Enzyme-Linked Immunosorbent Assay in Systemic Lupus Erythematosus
    Chang, Huang-Chen
    Wu, Yen-Ching
    Chen, Jun-Peng
    Wu, Yi-Da
    Huang, Wen-Nan
    Chen, Yi-Hsing
    Chen, Yi-Ming
    DIAGNOSTICS, 2021, 11 (11)
  • [4] THE CLINICAL-SIGNIFICANCE OF MEASURING DIFFERENT ANTI-DSDNA ANTIBODIES BY USING THE FARR ASSAY, AN ENZYME-IMMUNOASSAY AND A CRITHIDIA-LUCILIAE IMMUNOFLUORESCENCE TEST
    WERLE, E
    BLAZEK, M
    FIEHN, W
    LUPUS, 1992, 1 (06) : 369 - 377
  • [5] MEASUREMENT OF LOW AVIDITY ANTI-DSDNA BY THE CRITHIDIA-LUCILIAE TEST AND THE PEG ASSAY
    SMEENK, R
    VANDERLELIJ, G
    AARDEN, L
    CLINICAL AND EXPERIMENTAL IMMUNOLOGY, 1982, 49 (03): : 603 - 610
  • [6] Evaluation of a high avidity anti-dsDNA IgG enzyme-linked immunosorbent assay for the diagnosis of systemic lupus erythematosus
    Suh-Lailam, Brenda B.
    Chiaro, Tyson R.
    Davis, Wayne K.
    Wilson, Andrew R.
    Tebo, Anne E.
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL PATHOLOGY, 2011, 4 (08): : 748 - 754
  • [7] Clinical Evaluation of a New Quantitative Enzyme-Linked Immunosorbent Assay for Anti-dsDNA Antibodies Measured in Patients With Systemic Lupus Erythematosus
    Cui, Liyan
    Zhang, Jie
    Hu, Xiaozhou
    Zhang, Wenjing
    Ding, Jie
    LABMEDICINE, 2010, 41 (12): : 724 - 727
  • [8] Three different antigens for a highly specific anti-dsDNA ELISA in the diagnosis of SLE in comparison to the Crithidia luciliae immunofluorescence test.
    Matthias, T
    Wigand, R
    CLINICAL CHEMISTRY, 1996, 42 (06) : 117 - 117
  • [9] Comparison of enzyme-linked immunosorbent assay test with immunoblot assay in the diagnosis of pemphigus in Indian patients
    Khandpur, Sujay
    Sharma, Vinod K.
    Sharma, Alpana
    Pathria, Gaurav
    Satyam, Abhigyan
    INDIAN JOURNAL OF DERMATOLOGY VENEREOLOGY & LEPROLOGY, 2010, 76 (01): : 27 - 32
  • [10] EVALUATION OF AN ENZYME-LINKED IMMUNOSORBENT-ASSAY FOR TREPONEMAL ANTIBODY
    STEVENS, RW
    SCHMITT, ME
    JOURNAL OF CLINICAL MICROBIOLOGY, 1985, 21 (03) : 399 - 402