Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy

被引:88
|
作者
Axente, Marian [1 ]
Paidi, Ajay [2 ]
Von Eyben, Rie [1 ]
Zeng, Chuan [3 ]
Bani-Hashemi, Ali [2 ]
Krauss, Andreas [4 ]
Hristov, Dimitre [1 ]
机构
[1] Stanford Hosp & Clin, Radiat Oncol, Stanford, CA 94305 USA
[2] Siemens Med Solut USA, Computed Tomog & Radiat Oncol Dept, Martinez, CA 94553 USA
[3] Univ Penn, Radiat Oncol, Philadelphia, PA 19104 USA
[4] Siemens AG, Imaging & Therapy Div, Healthcare Sect, D-91301 Forcheim, Germany
关键词
metal artifact reduction; radiotherapy; raw data correction; HELICAL COMPUTED-TOMOGRAPHY; HIP PROSTHESES; CALIBRATION; NMAR;
D O I
10.1118/1.4906245
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To clinically evaluate an iterative metal artifact reduction (IMAR) algorithm prototype in the radiation oncology clinic setting by testing for accuracy in CT number retrieval, relative dosimetric changes in regions affected by artifacts, and improvements in anatomical and shape conspicuity of corrected images. Methods: A phantom with known material inserts was scanned in the presence/absence of metal with different configurations of placement and sizes. The relative change in CT numbers from the reference data (CT with no metal) was analyzed. The CT studies were also used for dosimetric tests where dose distributions from both photon and proton beams were calculated. Dose differences and gamma analysis were calculated to quantify the relative changes between doses calculated on the different CT studies. Data from eight patients (all different treatment sites) were also used to quantify the differences between dose distributions before and after correction with IMAR, with no reference standard. A ranking experiment was also conducted to analyze the relative confidence of physicians delineating anatomy in the near vicinity of the metal implants. Results: IMAR corrected images proved to accurately retrieve CT numbers in the phantom study, independent of metal insert configuration, size of the metal, and acquisition energy. For plastic water, the mean difference between corrected images and reference images was -1.3 HU across all scenarios (N = 37) with a 90% confidence interval of [-2.4, -0.2] HU. While deviations were relatively higher in images with more metal content, IMAR was able to effectively correct the CT numbers independent of the quantity of metal. Residual errors in the CT numbers as well as some induced by the correction algorithm were found in the IMAR corrected images. However, the dose distributions calculated on IMAR corrected images were closer to the reference data in phantom studies. Relative spatial difference in the dose distributions in the regions affected by the metal artifacts was also observed in patient data. However, in absence of a reference ground truth (CT set without metal inserts), these differences should not be interpreted as improvement/deterioration of the accuracy of calculated dose. With limited data presented, it was observed that proton dosimetry was affected more than photons as expected. Physicians were significantly more confident contouring anatomy in the regions affected by artifacts. While site specific preferences were detected, all indicated that they would consistently use IMAR corrected images. Conclusions: IMAR correction algorithm could be readily implemented in an existing clinical workflow upon commercial release. While residual errors still exist in IMAR corrected images, these images present with better overall conspicuity of the patient/phantom geometry and offer more accurate CT numbers for improved local dosimetry. The variety of different scenarios included herein attest to the utility of the evaluated IMAR for a wide range of radiotherapy clinical scenarios. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:1170 / 1183
页数:14
相关论文
共 50 条
  • [31] Clinical evaluation of the iterative metal artefact reduction algorithm for post-operative CT examination after maxillofacial surgery
    Hakim, Arsany
    Slotboom, Johannes
    Lieger, Olivier
    Schlittler, Fabian
    Giger, Roland
    Michel, Chantal
    Wiest, Roland
    Wagner, Franca
    DENTOMAXILLOFACIAL RADIOLOGY, 2017, 46 (04)
  • [32] Evaluation of the Direct Density Algorithm for Metal Artifact Reduction.
    Sanchez-Esperon, Javier
    Gaztelu-Blanco, Isabela
    Estrella-Maquez, Sara
    Picon-Garcia, Fernando
    Baeza-Trujillo, Montserrat
    Luis-Simon, Florencio Javier
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3819 - S3821
  • [33] Metal Artifact Reduction in Photon-Counting Detector CT Quantitative Evaluation of Artifact Reduction Techniques
    Skornitzke, Stephan
    Mergen, Victor
    Biederer, Juergen
    Alkadhi, Hatem
    Do, Thuy D.
    Stiller, Wolfram
    Frauenfelder, Thomas
    Kauczor, Hans-Ulrich
    Euler, Andre
    INVESTIGATIVE RADIOLOGY, 2024, 59 (06) : 442 - 449
  • [34] Evaluation of Efficacy of Iterative Metal Artifact Reduction Algorithm with the Beam Hardening Correction in Image Quality and Radiation Therapy
    Kim, C.
    Vassell, N.
    Na, Y.
    MEDICAL PHYSICS, 2022, 49 (06) : E725 - E726
  • [35] A New Iterative Metal Artifact Reduction Algorithm for Both Energy-Integrating and Photon-Counting CT Systems
    Anhaus, Julian A.
    Heider, Maximilian
    Killermann, Philipp
    Hofmann, Christian
    Mahnken, Andreas H.
    INVESTIGATIVE RADIOLOGY, 2024, 59 (07) : 526 - 537
  • [36] AN ALGORITHM FOR METAL STREAKING ARTIFACT REDUCTION IN CONE BEAM CT FOR BRACHYTHERAPY
    Bazalova, M.
    Landry, G.
    Beaulieu, L.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S101 - S101
  • [37] A new metal artifact reduction algorithm based on a deteriorated CT image
    Kano, Toru
    Koseki, Michihiko
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (06) : 901 - 912
  • [38] Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy
    Li, Hua
    Noel, Camille
    Chen, Haijian
    Li, H. Harold
    Low, Daniel
    Moore, Kevin
    Klahr, Paul
    Michalski, Jeff
    Gay, Hiram A.
    Thorstad, Wade
    Mutic, Sasa
    MEDICAL PHYSICS, 2012, 39 (12) : 7507 - 7517
  • [39] A hybrid metal artifact reduction algorithm for x-ray CT
    Zhang, Yanbo
    Yan, Hao
    Jia, Xun
    Yang, Jian
    Jiang, Steve B.
    Mou, Xuanqin
    MEDICAL PHYSICS, 2013, 40 (04)
  • [40] A new iterative algorithm for ring artifact reduction in CT using ring total variation
    Salehjahromi, Morteza
    Wang, Qian
    Zhang, Yanbo
    Gjesteby, Lars A.
    Harrison, Dan
    Wang, Ge
    Edic, Peter M.
    Yu, Hengyong
    MEDICAL PHYSICS, 2019, 46 (11) : 4803 - 4815