Comparison of point-of-care HbA1c test versus standardized laboratory testing

被引:29
|
作者
Sicard, DA [1 ]
Taylor, JR [1 ]
机构
[1] Univ Florida, Coll Pharm, Dept Pharm Practice, Gainesville, FL 32610 USA
关键词
diabetes; hemoglobin A(1c); point-of-care;
D O I
10.1345/aph.1E504
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
BACKGROUND: Glycosylated hemoglobin (HbA(1c)) evaluation is an accepted standard of care for assessing diabetes control. Point-of-care HbA(1c) monitors are relatively new, and there are few published data on their accuracy. OBJECTIVE: To compare a point-of-care HbA(1c) monitor with standardized laboratory testing. METHODS: Twenty-three patients who had type 1 or 2 diabetes were identified from our pharmacy-run diabetes clinic at a primary care clinic as needing HbA(1c) evaluation based on the American Diabetes Association guidelines. All patients with diabetes who attended our clinic were invited to participate in the study. Patients were required to obtain both a standardized laboratory HbA(1c) evaluation and a point-of-care A1c Now value within a week of each other. An analysis evaluating both clinical and statistical significance was completed at the end of the study. RESULTS: The A1c Now monitor showed good correlation (r = 0.758) to the standardized laboratory test. We found that the most accurate A1c Now values were within a range of 6-8%. One patient's value would have resulted in a different clinical decision if the decision were based on the A1c Now value. CONCLUSIONS: It appears that A1c Now may provide adequate accuracy and be useful for some patients. The monitor would be most beneficial for patients who cannot obtain standardized laboratory values secondary to either financial or transportation constraints. However, the decision to make adjustments in therapy should not be based solely on the A1c Now point-of-care result.
引用
收藏
页码:1024 / 1028
页数:5
相关论文
共 50 条
  • [21] Point-of-care testing of HbA1c and blood glucose in a remote Aboriginal Australian community
    Martin, DD
    Shephard, MDS
    Freeman, H
    Jones, TW
    Davis, EA
    Maguire, GP
    MEDICAL JOURNAL OF AUSTRALIA, 2005, 182 (10) : 524 - 527
  • [22] Validation and effect on diabetes control of glycated haemoglobin (HbA1c) point-of-care testing
    Pillay, S.
    Aldous, C. M.
    Singh, D.
    Pillay, D.
    SAMJ SOUTH AFRICAN MEDICAL JOURNAL, 2019, 109 (02): : 112 - 115
  • [23] Point-of-care testing of HbA1C is traceable to IFCC reference method by external calibration
    Zhou, Rui
    Tong, Qing
    Zuo, Chang
    Song, Zhi-Xin
    Ma, Huai-An
    Yue, Yu-Hong
    Zhang, Rui
    Wang, Qing-Tao
    CLINICA CHIMICA ACTA, 2014, 433 : 249 - 253
  • [24] Point-of-care testing of HbA1c, renal function and lipids in remote or disadvantaged regions
    Handrinos, Lena
    Januszewski, Andrzej S.
    Trimble, Elisabeth
    Donelan, Neil
    Williams, Paul
    Jenkins, Alicia J.
    INTERNAL MEDICINE JOURNAL, 2020, 50 (12) : 1567 - 1571
  • [25] Point-of-care testing for glycated haemoglobin (HbA1c) in the Young Persons Diabetes Clinic
    Masters, J.
    Birdsall, E. B.
    Parr, J.
    Hikmat, M.
    Oyibo, S. O.
    DIABETIC MEDICINE, 2018, 35 : 188 - 188
  • [26] Precision of the Afinion HbA1c Dx Point-of-Care Test in CLIA Waived Settings
    Clendenin, Marta
    Kupfer, Kenneth
    Green, Christopher L.
    Fortner, Kyle S.
    Bailey, Timothy S.
    Brick, Stephen
    Smiley, Sarah I.
    Strauss, Guy S.
    Thorsell, Ashley A.
    Ahmed, Naseer
    DIABETES, 2022, 71
  • [27] Performance evaluation of SD A1cCare as a HbA1c analyzer for point-of-care testing
    Lee, Kyunghoon
    Jun, Sun-Hee
    Han, Minje
    Song, Sang Hoon
    Park, Kyoung Un
    Song, Woon Heung
    Song, Junghan
    CLINICAL BIOCHEMISTRY, 2015, 48 (09) : 625 - 627
  • [28] The value proposition for point-of-care testing in healthcare: HbA1c for monitoring in diabetes management as an exemplar
    Price, Christopher P.
    St John, Andrew
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2019, 79 (05): : 298 - 304
  • [29] Point-of-Care Amperometric Testing of Diabetic Marker ( HbA1c) Using Specific Electroactive Antibodies
    Chopra, Adity
    Rawat, Sanjay
    Bhalla, Vijayender
    Suri, C. Raman
    ELECTROANALYSIS, 2014, 26 (03) : 469 - 472
  • [30] Systematic Diabetes Screening Using Point-of-Care HbA1c Testing Facilitates Identification of Prediabetes
    Whitley, Heather P.
    Hanson, Courtney
    Parton, Jason M.
    ANNALS OF FAMILY MEDICINE, 2017, 15 (02) : 162 - 164