An empirical model of diagnostic x-ray attenuation under narrow-beam geometry

被引:11
|
作者
Mathieu, Kelsey B. [1 ]
Kappadath, S. Cheenu [2 ]
White, R. Allen [3 ]
Atkinson, E. Neely [4 ]
Cody, Dianna D. [2 ]
机构
[1] Rice Univ, Dept Bioengn, Houston, TX 77005 USA
[2] Univ Texas MD Anderson Canc Ctr, Div Diagnost Imaging, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Div Bioinformat & Computat Biol, Houston, TX 77030 USA
[4] Rice Univ, Dept Stat, Houston, TX 77005 USA
关键词
CT; mammography; radiography; polyenergetic; half-value layer; HALF-VALUE LAYER; SPECTRA; CT;
D O I
10.1118/1.3592933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling. Methods: An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-01.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation]. Results: The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R-2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation. Conclusions: The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry). (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592933]
引用
收藏
页码:4546 / 4555
页数:10
相关论文
共 50 条
  • [41] ANALYTICAL MODEL FOR A 6 MEV X-RAY BEAM
    BORGER, F
    SIMPSON, L
    OVADIA, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (03): : 444 - &
  • [42] On X-ray channeling in narrow guides
    Dabagov, S. B.
    Uberall, H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (17): : 3881 - 3887
  • [43] Geometry Repeatability for Diagnostic Cone-Beam Imaging in the Emergency Department Using a Twin Robotic X-Ray System
    Herbst, M.
    Noo, F.
    Fieselmann, A.
    Vogt, S.
    Ritschl, L.
    Mertelmeier, T.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [44] CALCULATION OF INTEGRAL QUANTITIES FROM THE NARROW BEAM ATTENUATION CURVE OF A LOW-ENERGY X-RAY-BEAM
    DELGADO, V
    ORTIZ, P
    RADIATION PROTECTION DOSIMETRY, 1993, 48 (03) : 273 - 277
  • [45] X-ray optics - a tool for X-ray diagnostic and therapy
    Bradaczek, M.
    Popescu, M.
    JOURNAL OF OPTOELECTRONICS AND ADVANCED MATERIALS, 2007, 9 (07): : 1945 - 1948
  • [46] Monte Carlo simulator of realistic x-ray beam for diagnostic applications
    Bontempi, Marco
    Andreani, Lucia
    Rossi, Pier Luca
    Visani, Andrea
    MEDICAL PHYSICS, 2010, 37 (08) : 4201 - 4209
  • [47] Constancy check of beam quality in conventional diagnostic X-ray equipment
    Costa, Alessandro M.
    Badin, Romulo S.
    Leite, Marina S.
    Caldas, Linda V. E.
    APPLIED RADIATION AND ISOTOPES, 2008, 66 (10) : 1403 - 1405
  • [48] X-RAY ATTENUATION BY PRESCRIPTION LENSES
    YOUNG, RG
    CARLTON, WH
    RADIOLOGY, 1978, 129 (03) : 811 - 811
  • [49] ATTENUATION MEASUREMENTS IN THE X-RAY REGION
    VARIER, KM
    UNNIKRISHNAN, MP
    PHYSICAL REVIEW A, 1986, 33 (04): : 2382 - 2386
  • [50] ATTENUATION LENGTHS OF X-RAY PHOTOELECTRONS
    SOKCEVIC, D
    SUNJIC, M
    SOLID STATE COMMUNICATIONS, 1974, 15 (11) : 1703 - 1706