An empirical model of diagnostic x-ray attenuation under narrow-beam geometry

被引:11
|
作者
Mathieu, Kelsey B. [1 ]
Kappadath, S. Cheenu [2 ]
White, R. Allen [3 ]
Atkinson, E. Neely [4 ]
Cody, Dianna D. [2 ]
机构
[1] Rice Univ, Dept Bioengn, Houston, TX 77005 USA
[2] Univ Texas MD Anderson Canc Ctr, Div Diagnost Imaging, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Div Bioinformat & Computat Biol, Houston, TX 77030 USA
[4] Rice Univ, Dept Stat, Houston, TX 77005 USA
关键词
CT; mammography; radiography; polyenergetic; half-value layer; HALF-VALUE LAYER; SPECTRA; CT;
D O I
10.1118/1.3592933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling. Methods: An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-01.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation]. Results: The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R-2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation. Conclusions: The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry). (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3592933]
引用
收藏
页码:4546 / 4555
页数:10
相关论文
共 50 条
  • [2] Narrow-beam X-ray tests of CCD edge response
    Stephen Kuhlmann
    Harold Spinka
    Joseph P. Bernstein
    Kevin A. Beyer
    Lisa M. Gades
    Thomas E. Kasprzyk
    Antonino Miceli
    Richard A. Spence
    Richard Talaga
    Experimental Astronomy, 2011, 29 : 135 - 144
  • [3] Narrow-beam X-ray tests of CCD edge response
    Kuhlmann, Stephen
    Spinka, Harold
    Bernstein, Joseph P.
    Beyer, Kevin A.
    Gades, Lisa M.
    Kasprzyk, Thomas E.
    Miceli, Antonino
    Spence, Richard A.
    Talaga, Richard
    EXPERIMENTAL ASTRONOMY, 2011, 29 (03) : 135 - 144
  • [4] NARROW-BEAM ATTENUATION CURVE AS A FUNCTIONAL DESCRIPTION OF A LOW-ENERGY X-RAY-BEAM
    DELGADO, V
    ORTIZ, P
    X-RAY SPECTROMETRY, 1991, 20 (06) : 343 - 344
  • [5] Radiation Protection Clothing in X-Ray Diagnostics - Comparison of Attenuation Equivalents in Narrow Beam and Inverse Broad-Beam Geometry
    Pichler, T.
    Schoepf, T.
    Ennemoser, O.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2011, 183 (05): : 470 - 476
  • [6] Angular width of a narrow beam for X-ray linear attenuation coefficient measurements
    Midgley, S.
    RADIATION PHYSICS AND CHEMISTRY, 2006, 75 (09) : 945 - 953
  • [7] NARROW BEAM X-RAY INTERFEROMETER
    INDENBOM, VL
    SLOBODETSKII, IS
    TRUNI, KG
    ZHURNAL EKSPERIMENTALNOI I TEORETICHESKOI FIZIKI, 1974, 66 (03): : 1110 - 1120
  • [8] Dual-energy digital mammography for calcification imaging - Implementation under narrow-beam geometry
    Kappadath, S
    Shaw, C
    Liu, X
    Lai, C
    Balter, P
    Whitman, G
    MEDICAL PHYSICS, 2003, 30 (06) : 1455 - 1455
  • [9] A mathematical model for detection of back-scattered radiation during diagnostic scanning with a narrow X-Ray beam
    Buckley A.A.
    Vladimirov L.V.
    Kozlov A.A.
    Biomedical Engineering, 2009, 43 (5) : 221 - 224
  • [10] ANGULAR DEPENDENCE OF X-RAY ATTENUATION FOR SAMPLES OF CYLINDRICAL AND SPHERICAL SHAPES IN A VERY NARROW BEAM
    MAZO, DM
    SOVIET PHYSICS CRYSTALLOGRAPHY, USSR, 1968, 13 (02): : 185 - &