Evaluation of the Accuracy of Different Combinations of Automatic Registration Selection Factors in Obtaining Tomotherapy Megavoltage Computed Tomography Images

被引:3
|
作者
Lee, Suk [1 ]
Cao, Yuan Jie [1 ]
Shim, Jang Bo [1 ]
Chang, Kyung Hwan [1 ]
Yang, Dae Sik [1 ]
Park, Young Je [1 ]
Yoon, Won Seob [1 ]
Kim, Chul Yong [1 ]
Cho, Kwang Hwan [2 ]
Min, C. K. [2 ]
Kim, Y. H. [2 ]
机构
[1] Korea Univ, Coll Med, Dept Radiat Oncol, Seoul 130701, South Korea
[2] Soonchunhyang Univ, Boocheon Hosp, Dept Radiat Oncol, Bouchon 420767, South Korea
关键词
Tomotherapy; MVCT; Automatic registration control; Scan image control; MODULATED RADIATION-THERAPY; GUIDED HELICAL TOMOTHERAPY; CONFORMAL RADIOTHERAPY; CT IMAGES; HEAD; IMPLEMENTATION; LOCALIZATION; VERIFICATION; DELIVERY; GUIDANCE;
D O I
10.3938/jkps.60.1961
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The purpose of our study is to find the potential best options for the registration selection parameters when performing automatic registration between kilovoltage computed tomography (KVCT) and megavoltage computed tomography (MVCT) images in helical tomotherapy. We compared nine different combinations of automatic registration parameters, including technique and resolution in automatic registration control and tomoimage filtering in scan image control. In the phantom study, we compared the translational and the rotational adjustments between the original position (lateral 0 cm; longitudinal 0 cm; vertical 0 cm) and the manually moved position (lateral 1 cm; longitudinal 1 cm; vertical 1 cm). In the patient study, we compared translational (lateral, longitudinal, vertical) and rotational (pitch, roll, yaw) adjustments between the KVCT and the MVCT images for different anatomic treatment sites. The ranges of translational and rotational displacements in the phantom study were 14.87 +/- 0.12 similar to 15.70 +/- 0.04 mm and 0 degrees similar to 1.03 +/- 0.58 degrees, respectively. In cranial cases, the ranges of translational and rotational root-mean-square values were 4.3 +/- 0.4 mm similar to 5.3 +/- 0.6 mm, and 0.3 +/- 0.3 degrees similar to 0.7 +/- 0.4 degrees, respectively. In the extra cranial cases, the ranges of translational and rotational root-mean-square values were 14.1 +/- 3.6 mm similar to 16.1 +/- 2.5 mm, and 0.5 +/- 0.3 degrees similar to 1.0 +/- 0.5 degrees, respectively. We found that the combination of "full image - super fine" could minimize the mechanical deviations in the cranial case, and that both the combinations of "bone tissue - super fine" and "full image - super fine" were the best options in extra cranial cases.
引用
收藏
页码:1961 / 1966
页数:6
相关论文
共 50 条
  • [1] Evaluation of the accuracy of different combinations of automatic registration selection factors in obtaining tomotherapy megavoltage computed tomography images
    Suk Lee
    Yuan Jie Cao
    Jang Bo Shim
    Kyung Hwan Chang
    Dae Sik Yang
    Young Je Park
    Won Seob Yoon
    Chul Yong Kim
    Kwang Hwan Cho
    C. K. Min
    Y. H. Kim
    Journal of the Korean Physical Society, 2012, 60 : 1961 - 1966
  • [2] The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes
    Forrest, LJ
    Mackie, TR
    Ruchala, K
    Turek, M
    Kapatoes, J
    Jaradat, H
    Hui, S
    Balog, J
    Vail, DM
    Mehta, MP
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (05): : 1639 - 1644
  • [3] Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy:: An evaluation of the setup verification process for the special case of a rigid head phantom
    Boswell, Sarah
    Tome, Wolfgang
    Jeraj, Robert
    Jaradat, Hazirn
    Mackie, T. Rock
    MEDICAL PHYSICS, 2006, 33 (11) : 4395 - 4404
  • [4] Performance evaluation of image reconstruction algorithms for a megavoltage computed tomography system on a helical tomotherapy unit
    Tegtmeier, Riley C.
    Ferris, William S.
    Bayouth, John E.
    Culberson, Wesley S.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (04)
  • [5] A study on the image registration accuracy of intrafraction cone beam computed tomography images
    Arumugam, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S566 - S566
  • [6] Set up accuracy of patients with head and neck tumors using the megavoltage computed tomography imaging capacity of the helical tomotherapy system
    Wandering, Aude
    Vynckier, Stefaan
    Gregoire, Vincent
    ACTA CLINICA BELGICA, 2008, 63 (02): : 131 - 131
  • [7] Megavoltage Computed Tomography Image Guidance With Helical Tomotherapy in Patients With Vertebral Tumors: Analysis of Factors Influencing Interobserver Variability
    Levegruen, Sabine
    Poettgen, Christoph
    Abu Jawad, Jehad
    Berkovic, Katharina
    Hepp, Rodrigo
    Stuschke, Martin
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 561 - 569
  • [8] Set up accuracy of patients with head and neck tumors using the megavoltage computed tomography image capacity of a helical tomotherapy system.
    Vaandering, A.
    Gregoire, V.
    Vynckier, S.
    Tomsej, M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S204 - S205
  • [9] Set-up accuracy of patients with head and neck tumors using the megavoltage computed tomography imaging capacity of the helical tomotherapy system
    Vaandering, A.
    Vynckier, S.
    Tomsej, M.
    Gregoire, V.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S70 - S70
  • [10] Prostate contouring uncertainty in megavoltage computed tomography images acquired with a helical tomotherapy unit during image-guided radiation therapy
    Song, William Y.
    Chiu, Bernard
    Bauman, Glenn S.
    Lock, Michael
    Rodrigues, George
    Ash, Robert
    Lewis, Craig
    Fenster, Aaron
    Battista, Jerry J.
    Van Dyk, Jake
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (02): : 595 - 607