Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs

被引:2
|
作者
Hedayatipanah, Morad [1 ]
Salemi, Fatemeh [2 ]
Kamyari, Naser [3 ]
Yalpanian, Ashkan [4 ]
机构
[1] Hamadan Univ Med Sci, Fac Dent, Dept Periodont, Hamadan, Hamadan, Iran
[2] Hamadan Univ Med Sci, Fac Dent, Dept Oral & Maxillofacial Radiol, Hamadan, Hamadan, Iran
[3] Abadan Fac Med Sci, Dept Biostat, Abadan, Iran
[4] Hamadan Univ Med Sci, Fac Dent, Hamadan, Hamadan, Iran
关键词
Cone-Beam Computed Tomography; Dental Implants; Artifacts; CT; ACCURACY;
D O I
10.1590/pboci.2021.118
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To assess the effect of metal artifact reduction (MAR) algorithm of two cone-beam computed tomography (CBCT) software programs on the accuracy of peri-implant bone width measurements. Material and Methods: This in vitro study evaluated 35 bovine rib bone blocks measuring 8 x 8 x 11 mm. Titanium implants were inserted in bone blocks and placed in a wax model of mandible. CBCT scans were obtained with ProMax 3D and Cranex 3D CBCT systems with and without the MAR algorithm. The width of buccal and lingual bone plates surrounding the implant was measured on CBCT scans by two observers. The Cronbach's alpha was calculated to assess inter-observer agreement. The area under the curve (AUC), sensitivity and specificity values were calculated and compared. Results: The two observers had an excellent agreement. The accuracy of Cranex 3D was higher than that of ProMax 3D (p<0.05). Both CBCT systems showed higher accuracy when the MAR algorithm was not used (p<0.05). Both CBCT systems showed higher accuracy for measuring the width of the lingual plate than buccal plate (p<0.05). Conclusion: The application of MAR algorithm did not notably increase the measurement accuracy in any CBCT system. Cranex 3D showed generally higher measurement accuracy than ProMax 3D.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Efficacy of a cone beam computed tomography metal artifact reduction algorithm for the detection of peri-implant fenestrations and dehiscences
    Lins de-Azevedo-Vaz, Sergio
    Peyneau, Priscila Dias
    Ramirez-Sotelo, Laura Ricardina
    Vasconcelos, Karla de Faria
    Flores Campos, Paulo Sergio
    Haiter-Neto, Francisco
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2016, 121 (05): : 550 - 556
  • [2] Validating cone-beam computed tomography for peri-implant bone morphometric analysis
    Yan Huang
    Jeroen Van Dessel
    Maarten Depypere
    Mostafa EzEldeen
    Alexandru Andrei Iliescu
    Emanuela Dos Santos
    Ivo Lambrichts
    Xin Liang
    Reinhilde Jacobs
    Bone Research, 2014, (02) : 103 - 109
  • [3] Validating cone-beam computed tomography for peri-implant bone morphometric analysis
    Yan Huang
    Jeroen Van Dessel
    Maarten Depypere
    Mostafa EzEldeen
    Alexandru Andrei Iliescu
    Emanuela Dos Santos
    Ivo Lambrichts
    Xin Liang
    Reinhilde Jacobs
    Bone Research, 2014, 2 (02) : 103 - 109
  • [4] Validating cone-beam computed tomography for peri-implant bone morphometric analysis
    Huang, Yan
    Van Dessel, Jeroen
    Depypere, Maarten
    EzEldeen, Mostafa
    Iliescu, Alexandru Andrei
    Dos Santos, Emanuela
    Lambrichts, Ivo
    Liang, Xin
    Jacobs, Reinhilde
    BONE RESEARCH, 2014, 2
  • [5] Validating cone-beam computed tomography for peri-implant bone morphometric analysis
    Yan Huang
    Jeroen Van Dessel
    Maarten Depypere
    Mostafa EzEldeen
    Alexandru Andrei Iliescu
    Emanuela Dos Santos
    Ivo Lambrichts
    Xin Liang
    Reinhilde Jacobs
    Bone Research, 2
  • [6] Measuring peri-implant bone lesions using low-dose cone-beam computed tomography
    Leisner, Laura Christine
    Tasaka, Akinori
    Trebing, Charlotte Theresa
    Hilgenfeld, Tim
    Kosinski, Matthias Adalbert
    Kronsteiner, Dorothea
    Rammelsberg, Peter
    Schwindling, Franz Sebastian
    JOURNAL OF PROSTHODONTIC RESEARCH, 2022, 66 (02) : 326 - 332
  • [7] Efficacy of two radiographic algorithms for detection of peri-implant bone defects on cone-beam computed tomography scans
    Yousefi, Faezeh
    Heidari, Ali
    Ehsani, Azita
    Farhadian, Maryam
    Ehsani, Marzieh
    BMC ORAL HEALTH, 2025, 25 (01):
  • [8] Metal Artifact Reduction in Helical Cone-beam Computed Tomography
    Tang, Jie
    Zhang, Li
    Chen, Zhiqiang
    Xing, Yuxiang
    Cheng, Jianping
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2878 - 2881
  • [9] Higher Resolution in Cone Beam Computed Tomography Is Accompanied by Improved Bone Detection in Peri-implant Bone Despite Metal Artifact Presence
    Kerkfeld, Valentin
    Meyer, Ulrich
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (06) : 1331 - 1338
  • [10] Evaluation of a metal artifact reduction algorithm and an optimization filter in the estimation of peri-implant dehiscence defects by using cone beam computed tomography: an in-vitro study
    Bayrak, Seval
    Orhan, Kaan
    Cakmak, Emine Sebnem Kursun
    Gorurgoz, Cansu
    Odabasi, Onur
    Yilmaz, Dervis
    Atakan, Cemal
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2020, 130 (02): : 209 - 216