Measurement uncertainty evaluation by kurtosis method at calibration of electrical resistance standards using a comparator

被引:0
|
作者
Zakharov, I [1 ,2 ]
Botsiura, O. [2 ]
Semenikhin, V [1 ]
机构
[1] Natl Sci Ctr, Inst Metrol, Myronosytska Str 42, UA-61002 Kharkiv, Ukraine
[2] Kharkiv Natl Univ Radio Elect, Nauky Ave 14, UA-61166 Kharkiv, Ukraine
来源
关键词
electrical resistance standard; resistance comparator; calibration; measurement uncertainty; uncertainty budget; kurtosis method; Monte-Carlo method;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The article provides examples of the application of electrical resistance standards in metrological practice. Existing methods for their calibration are analyzed. It is established that calibration using a comparator is the most accurate and common method for calibrating electrical resistance standards. A model for transferring the size of the resistance unit in calibration of electrical resistance standards using a comparator is considered. An expression is given for the evaluation of the value of measurand. The procedure for estimating the expanded measurement uncertainty based on the kurtosis method is described, and the uncertainty budget is drawn up. An example of evaluation of the measurement uncertainty in calibration of the resistance coil P321 using a resistance comparator P3015 is given. The coincidence of the obtained results with those got using the Monte Carlo method is shown.
引用
收藏
页码:12 / 16
页数:5
相关论文
共 50 条
  • [21] The Measurement Uncertainty Evaluation of Yaw Sensor Calibration
    Tipsuwanporn, Vittaya
    Chinchusak, Wanchai
    Jaturunruangsri, Supaporn
    2016 16TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS), 2016, : 555 - 558
  • [22] Evaluation of measurement uncertainty for thermometers with calibration equations
    Chen, CC
    ACCREDITATION AND QUALITY ASSURANCE, 2006, 11 (1-2) : 75 - 82
  • [23] Defining an Uncertainty Budget in Electrical Power and Energy Reference Standards Calibration
    Demerdziev, Kiril
    Cundeva-Blajer, Marija
    Dimchev, Vladimir
    Srbinovska, Mare
    Kokolanski, Zivko
    PROCEEDINGS OF 18TH INTERNATIONAL CONFERENCE ON SMART TECHNOLOGIES (IEEE EUROCON 2019), 2019,
  • [24] Evaluation of the uncertainty component due to resistance measurement in the calibration of PRTs at the Al freezing point
    Yamazawa, K
    Harada, K
    Arai, M
    SICE 2002: PROCEEDINGS OF THE 41ST SICE ANNUAL CONFERENCE, VOLS 1-5, 2002, : 2444 - 2447
  • [25] Calibration method of microscope and its measurement uncertainty
    Bao Jian-ping
    Wang Hui-sheng
    Zhang Xiao-fei
    Third International Symposium on Precision Mechanical Measurements, Pts 1 and 2, 2006, 6280 : U706 - U712
  • [26] Improved Evaluation of Planar Calibration Standards Using the TDR Preselection Method
    Vancl, J.
    ACTA POLYTECHNICA, 2007, 47 (4-5) : 102 - 106
  • [27] NONLINEAR METHOD FOR INCLUDING MASS UNCERTAINTY OF GRAVIMETRIC STANDARDS AND SYSTEM MEASUREMENT ERRORS IN FITTING OF CALIBRATION CURVES
    PICKLES, WL
    MCCLURE, JW
    HOWELL, RH
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1978, 175 (MAR): : 13 - 13
  • [28] CALIBRATION PROCEDURE FOR STABILIZED LASERS USING THE METHOD OF OPTICAL BEATS MEASUREMENT UNCERTAINTY
    Dugheanu, Elena
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2007, 69 (01): : 63 - 70
  • [29] Calibration procedure for stabilized lasers using the method of optical beats measurement uncertainty
    National Institute of Metrology, Bucharest, Romania
    UPB Sci Bull Ser A, 2007, 1 (63-70): : 63 - 70
  • [30] A New Self-Calibration Method of an LCR Meter for RF Resistance Calibration Using Capacitance Standards
    Suzuki, Kiyoshi
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (04) : 993 - 996