Reduction of radiation dose to patients undergoing barium enema by dose audit

被引:8
|
作者
Yu, SK [1 ]
Cheung, YK
Chan, TL
Kung, CM
Yuen, MK
机构
[1] Tuen Mun Hosp, Med Phys Div, Hong Kong, Hong Kong, Peoples R China
[2] Tuen Mun Hosp, Dept Diagnost Radiol, Hong Kong, Hong Kong, Peoples R China
来源
BRITISH JOURNAL OF RADIOLOGY | 2001年 / 74卷 / 878期
关键词
D O I
10.1259/bjr.74.878.740162
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Nowadays, new fluoroscopic machines are usually equipped with a dose-area product (DAP) meter for dose measurement. In our hospital, DAP meters have been used in the Diagnostic Radiology Department for dose audit since June 1997. Demographic patient data, name of radiologist, fluoroscopic duration and DAP readings of every case were recorded by radiographers. In early 1999, questionnaires were distributed to radiologists who had performed fluoroscopic examinations during the auditing period. 23 radiologists with varying years of experience completed the questionnaire and their practice was analysed. Since familiarization with the examination technique would affect radiologists' practice, these radiologists were divided into two groups for analysis. Radiologists with less than 3 years of experience were grouped together as junior radiologists, whilst others were grouped as senior radiologists. Results of the questionnaire indicated that radiologists generally found DAP meters useful for dose evaluation in the process of technique refinement. Radiologists aware of being under continuous surveillance of their practice showed significant reduction of doses (junior radiologists 25%, p <0.005; senior radiologists 36%, p <0.05) and fluoroscopic times (junior radiologists 36%, p <0.001; senior radiologists 18%, p <0.05) compared with radiologists who were unaware that they were under surveillance but with similar radiological experience. This effect is believed to be because of increased awareness of radiation dose through audit. In addition, this "audit effect" may also affect junior radiologists in decision-making regarding the number of radiographs (p <0.05), but no effect was found for senior radiologists (p >0.5).
引用
收藏
页码:162 / 165
页数:4
相关论文
共 50 条
  • [1] Radiation Dose Reduction for Barium Enema: an Achievable Improvement in Patient Care
    Lai, B. M. H.
    Fung, V. K. P.
    Shum, J. S. F.
    Sy, A. N. L.
    Cheng, S. C. S.
    Khoo, J. L. S.
    HONG KONG JOURNAL OF RADIOLOGY, 2011, 14 (04): : 212 - 217
  • [2] Radiation doses to patients undergoing barium meal and barium enema examinations
    Delichas, MG
    Hatziioannou, K
    Papanastassiou, E
    Albanopoulou, P
    Chatzi, E
    Sioundas, A
    Psarrakos, K
    RADIATION PROTECTION DOSIMETRY, 2004, 109 (03) : 243 - 247
  • [3] Patient dose reduction by audit of grid usage in barium enemas
    Seymour, R
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (833): : 489 - 491
  • [4] Dose reduction in double contrast barium enema by use of low fluoroscopic current
    Smiddy, PF
    Quinn, AD
    Freyne, PJ
    Marsh, D
    Murphy, JM
    BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (825): : 852 - 854
  • [5] Radiation dose and diagnosticity of barium enema examinations by radiographers and radiologists: a comparative study
    Crawley, MT
    Shine, B
    Booth, A
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (844): : 399 - 405
  • [6] Radiation dose and diagnosticity of barium enema examinations by radiographers and radiologists: a comparative study (3)
    Hetherington, J
    Chapple, CL
    Rawlings, DJ
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (851): : 1219 - 1221
  • [7] ENEMA REDUCTION OF INTUSSUSCEPTION WITH A SMALL DOSE OF IOPAMIDOL MAY HAVE ADVANTAGES OVER BARIUM
    DOI, O
    KIYAMA, T
    SATO, A
    OKAFUJI, T
    KONOHANA, A
    KUROSAKA, F
    YAMADA, K
    OKAMOTO, Y
    PEDIATRIC SURGERY INTERNATIONAL, 1995, 10 (5-6) : 332 - 334
  • [8] Radiation dose and diagnosticity of barium enema examinations by radiographers and radiologists: a comparative study (2)
    McKenzie, GA
    Mathers, SA
    Graham, DT
    Chesson, RA
    Wade, JP
    Darragh, CL
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (851): : 1218 - 1219
  • [9] Real-time measurement and audit of radiation dose to patients undergoing computed radiography
    Vano, E
    Fernandez, JM
    Ten, JI
    Guibelalde, E
    Gonzalez, L
    Pedrosa, CSA
    RADIOLOGY, 2002, 225 (01) : 283 - 288
  • [10] Reducing dose at barium enema: radiographers do it digitally
    Crawley, MT
    Booth, A
    BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (896): : 652 - 656