Optimizing separation conditions for proteins and peptides using imaged capillary isoelectric focusing

被引:59
|
作者
Wu, JQ [1 ]
Li, SC [1 ]
Watson, A [1 ]
机构
[1] Convergent Biosci Ltd, Toronto, ON M8Z 2L8, Canada
关键词
capillary isoelectric focusing; detection; electrophoresis; whole-column imaging detection; proteins; antibodies; peptides; glycoproteins;
D O I
10.1016/S0021-9673(98)00326-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Separation conditions for antibodies, glycoproteins and peptides were optimized to fully realize the potential of automated imaged capillary isoelectric focusing (imaged cIEF) for protein analysis. Two commercially available capillary coatings, polyacrylamide and fluorocarbon, were found to provide reproducible results for cIEF separations. Both coatings could last more than 100 runs under normal cIEF conditions. Up to 30 mM salts (Na+) could be added to samples to prevent protein precipitation before and during isoelectric focusing performed under imaged cIEF. Short analysis time of the imaged cIEF also aided in the prevention of protein precipitation. High current at the beginning of the focusing for samples in salt could be avoided by applying a voltage gradient. Additions of up to 6 M urea and 20% glycerol could enhance solubility of proteins and peptide. Imaged cIEF was applied to the quantitation of monoclonal antibodies. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:163 / 171
页数:9
相关论文
共 50 条
  • [1] OPTIMIZING SEPARATION PARAMETERS IN CAPILLARY ISOELECTRIC-FOCUSING
    ZHU, MD
    RODRIGUEZ, R
    WEHR, T
    JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2): : 479 - 488
  • [2] A novel approach enables imaged capillary isoelectric focusing analysis of PEGylated proteins
    Zhang, Xin
    Chemmalil, Letha
    Ding, Julia
    Li, Zhengjian
    ELECTROPHORESIS, 2020, 41 (09) : 735 - 742
  • [3] Capillary isoelectric focusing of proteins
    Pritchett, TJ
    ELECTROPHORESIS, 1996, 17 (07) : 1195 - 1201
  • [4] Separation of humic acids using capillary isoelectric focusing
    Kovács, P
    Posta, J
    MICROCHEMICAL JOURNAL, 2005, 79 (1-2) : 49 - 54
  • [5] Isoelectric focusing using whole-column imaged capillary electrophoresis.
    Will, ML
    Oswald, CB
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2004, 228 : U204 - U204
  • [6] Coupling imaged capillary isoelectric focusing with mass spectrometry using a nanoliter valve
    Montealegre, Cristina
    Neusuess, Christian
    ELECTROPHORESIS, 2018, 39 (9-10) : 1151 - 1154
  • [7] Determination of the isoelectric point of proteins by capillary isoelectric focusing
    Righetti, PG
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1037 (1-2) : 491 - 499
  • [8] Capillary isoelectric focusing with whole column imaging detection for analysis of proteins and peptides
    Mao, QL
    Pawliszyn, J
    JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1999, 39 (1-2): : 93 - 110
  • [9] Capillary isoelectric focusing as a tool in the examination of antibodies, peptides and proteins of pharmaceutical interest
    Liu, X
    Sosic, Z
    Krull, IS
    JOURNAL OF CHROMATOGRAPHY A, 1996, 735 (1-2) : 165 - 190
  • [10] Quantitation of CRM197 using imaged capillary isoelectric focusing with fluorescence detection and capillary Western
    Loughney, John W.
    Ha, Sha
    Rustandi, Richard R.
    ANALYTICAL BIOCHEMISTRY, 2017, 534 : 19 - 23