共 48 条
Difficult Removal of Retrievable IVC Filters: A Description of the "Double-Wire Restraining" Technique
被引:2
|作者:
Owens, Charles A.
[1
]
Bui, James T.
[1
]
Knuttinen, M-Grace
[1
]
Emmanuel, Neelmini
[1
]
Carrillo, Tami C.
[1
]
Gaba, Ron C.
[1
]
机构:
[1] Univ Illinois, Med Ctr, Div Intervent Radiol MC 931, Dept Radiol, Chicago, IL 60612 USA
关键词:
IVC filter;
Foreign body retrieval;
Venous intervention;
Thromboembolic disease;
Migration;
Complication;
VENA-CAVA FILTER;
CLINICAL-EXPERIENCE;
MULTICENTER;
CATHETER;
D O I:
10.1007/s00270-010-9911-1
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
We describe our experience with the use of the "double-wire restraining" technique to assist in the removal of two retrievable inferior vena cava filters: one had been misplaced in the right brachiocephalic vein with apex perforation of the vessel wall, and the second filter had migrated cephalad to straddle across both renal veins. The "double-wire restraining" technique consists of two stiff-shaft Glidewires (Terumo, Somerset, NJ) placed through the same introducer sheath and positioned on opposite sides of the filter. Both wires restrain the filter at the tip of the sheath as the sheath is advanced, thus allowing the operator to reposition the filter. This report details how this technique was used to realign two malpositioned filters and reposition the filter apices from their extravascular location, thus exposing them for ensnarement.
引用
收藏
页码:S218 / S223
页数:6
相关论文