Clinical and Biomechanical Effects of Femoral Neck Buttress Plate used for Vertical Femoral Neck Fractures

被引:6
|
作者
Steffensmeier, Andrew [1 ]
Shah, Nihar [1 ]
Archdeacon, Michael [1 ]
Watson, David [2 ,3 ]
Sanders, Roy W. [2 ,3 ]
Sagi, Henry C. [1 ]
机构
[1] Univ Cincinnati, Orthopaed Surg Dept, 231 Albert Sabin Way, Cincinnati, OH 45367 USA
[2] Univ S Florida, Dept Orthopaed Surg, Tampa, FL 33620 USA
[3] Florida Orthopaed Inst, Orthopaed Trauma Serv, Tampa, FL USA
关键词
HIP-FRACTURES; FIXATION;
D O I
10.1016/j.injury.2021.12.004
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose: To investigate both the biomechanical and clinical effect of an inferomedial femoral neck buttress plate (FNBP) used to augment a sliding hip screw (SHS) and anti-rotational screw (ARS) in the treatment of traumatic vertical femoral neck fractures. Methods: Part 1: Clinical - Retrospective review of patients under age 65 treated with open reduction of a vertical femoral neck fracture. Patients were divided into two groups: Group 1 patients (18 patients) had SHS/ARS fixation augmented with a FNBP, while Group 2 patients (18 patients) had SHS/ARS fixation alone and were matched for age and sex. Demographic data, OTA fracture classification, immediate post-operative and follow-up radiographs were analyzed for quality of reduction, femoral neck shortening (FNS), neck-shaft angle (NSA), avascular necrosis (AVN) and union. Part 2: Biomechanical - Pauwels III femoral neck osteotomy was created in five pairs of cadaveric specimens, then each fracture was reduced and stabilized with a SHS/ARS construct. Specimens were matched and split into Groups 1 and 2, similar to Part 1. Cadaveric specimens were axially loaded in cyclical fashion to analyze for construct stiffness, fracture displacement femoral neck shortening and changes in the neck shaft angle. Results: Part 1: There were 18 matched patients (14 males and 4 females) in both Group 1 and Group 2. There were no statistically significant differences between the two groups with respect to Pauwels angle, femoral neck shortening, changes in neck-shaft angle, AVN or nonunion. One reoperation in Group 1 and four in Group 2. Part 2: All five cadaveric specimens in both groups survived the 10,00 0-cycle loading regimen. We were unable to detect any significant differences between the two groups with respect to construct stiffness, change in neck-shaft angle or amount of femoral neck shortening. Conclusion: Based on the results of both clinical case series and biomechanical testing, an inferomedial neck buttress plate does not appear to offer long-term benefits with respect to maintenance of alignment or achieving union but may potentially help in obtaining the reduction. (C) 2021 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1137 / 1143
页数:7
相关论文
共 50 条
  • [41] Computational evaluation of the biomechanical effects of position changes in the femoral neck system on Pauwels type III femoral neck fractures: an in silico study
    Zhang, Xiang
    Zhang, Shenghang
    Zhong, Zhou
    Zhang, Wen
    Xiong, Zhongwei
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2025, 13
  • [42] Biomechanical comparison of fixation methods on third-generation femoral composite bone models in Pauwels type 3 femoral neck fractures: Contribution of the medial buttress plate to fixation
    Kilic, Feyzi
    Polat, Abdulkadir
    Yamak, Fatih
    Bozdag, Ergun
    Fidan, Firat
    Basar, Hakan
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2023, 57 (05) : 243 - 249
  • [43] A biomechanical reappraisal of the femoral neck
    Claxton, Alexander G.
    AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 2018, 165 : 50 - 50
  • [44] Clinical efficacy of Femoral Neck System for treatment of unstable femoral neck fractures in young adults
    Guo, ChangYong
    Huang, JiaPing
    Chen, ZhiBiao
    Cai, ZeMing
    Cai, TaoYi
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2024, 52 (05)
  • [45] Proximal Femoral Locking Plate With Cannulated Screws for the Treatment of Femoral Neck Fractures
    Lin, Dasheng
    Lian, Kejian
    Ding, Zhenqi
    Zhai, Wenliang
    Hong, Jiayuan
    ORTHOPEDICS, 2012, 35 (01) : E1 - E5
  • [46] Clinical Profile of Fallers with Femoral Neck Fractures
    Tsur, Atzmon
    Eluz, Dana
    Ltah, Dorit
    Segal, Zvi
    Shakeer, Nael
    Galin, Arkady
    PM&R, 2014, 6 (05) : 390 - 394
  • [47] Biomechanical Analysis of Fixation Devices for Basicervical Femoral Neck Fractures
    Johnson, Joseph
    Deren, Matthew
    Chambers, Alison
    Cassidy, Dale
    Koruprolu, Sarath
    Born, Christopher
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2019, 27 (01) : E41 - E48
  • [48] Femoral neck fracture in a young person - ORIF but do we use a medial buttress plate as well on the femoral neck? OTG 23
    Giordano, Vincenzo
    Rayes, Johnny
    Buckley, R. E.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (12): : 3867 - 3868
  • [49] Survivability of the Femoral Neck System for the treatment of femoral neck fractures in adults
    Stegelmann, Samuel D.
    Butler, Justin T.
    Matthews, Dylan J.
    Ostlie, Hunter C.
    Boothby, Benjamin C.
    Phillips, Seth A.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (06): : 2555 - 2563
  • [50] The Effects and Risk Factors of Femoral Neck Shortening after Internal Fixation of Femoral Neck Fractures
    Lee, Dae Hee
    Kwon, Joo Han
    Kim, Ki-Choul
    CLINICS IN ORTHOPEDIC SURGERY, 2024, 16 (05) : 718 - 723