Comparison of CCS Values Determined by Traveling Wave Ion Mobility Mass Spectrometry and Drift Tube Ion Mobility Mass Spectrometry

被引:92
|
作者
Hinnenkamp, Vanessa [1 ,2 ,3 ]
Klein, Julia [4 ,5 ]
Meckelmann, Sven W. [4 ,5 ]
Balsaa, Peter [1 ]
Schmidt, Torsten C. [1 ,2 ,3 ]
Schmitz, Oliver J. [4 ,5 ]
机构
[1] IWW Water Ctr, Moritzstr 26, D-45476 Mulheim, Germany
[2] Instrumental Analyt Chem, Univ Str 5, D-45141 Essen, Germany
[3] Ctr Water & Environm Res, Univ Str 5, D-45141 Essen, Germany
[4] Univ Duisburg Essen, Appl Analyt Chem, Univ Str 5, D-45141 Essen, Germany
[5] Univ Duisburg Essen, Teaching & Res Ctr Separat, Univ Str 5, D-45141 Essen, Germany
关键词
COLLISION CROSS-SECTION; GAS-PHASE; PROTEIN COMPLEXES; SMALL MOLECULES; STRUCTURAL-CHARACTERIZATION; AMINOBENZOIC ACID; CALIBRATION; DATABASE; PEPTIDE; SEPARATION;
D O I
10.1021/acs.analchem.8b02711
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Collision cross section (CCS, Omega) values determined by ion mobility mass spectrometry (IM-MS) provide the study of ion shape in the gas phase and use of these as further identification criteria in analytical approaches. Databases of CCS values for a variety of molecules determined by different instrument types are available. In this study, the comparability of CCS values determined by a drift tube ion mobility mass spectrometer (DTIM-MS) and a traveling wave ion mobility mass spectrometer (TWIM-MS) was investigated to test if a common database could be used across IM techniques. A total of 124 substances were measured with both systems and CCS values of [M + H](+) and [M + Na](+) adducts were compared. Deviations <1% were found for most substances, but some compounds show deviations up to 6.2%, which indicate that CCS databases cannot be used without care independently from the instrument type. Additionally, it was found that for several molecules [2M + Na](+) ions were formed during electrospray ionization, whereas a part of them disintegrates to [M + Na](+) ions after passing through the drift tube and before reaching the TOF region, resulting in two signals in their drift spectrum for the [M + Na](+) adduct. Finally, the impact of different LC-IM-MS settings (solvent composition, solvent flow rate, desolvation temperature, and desolvation gas flow rate) were investigated to test whether they have an influence on the CCS values or not. The results showed that these conditions have no significant impact. Only for karbutilate changes in the drift spectrum could be observed with different solvent types and flow rates using the DTIM-MS system, which could be caused by the protonation at different sites in the molecule.
引用
收藏
页码:12042 / 12050
页数:9
相关论文
共 50 条
  • [21] Ion Mobility Spectrometry: Solo or Coupled to Mass Spectrometry
    Busch, Kenneth L.
    SPECTROSCOPY, 2013, 28 (11) : 12 - +
  • [22] Developments in ion mobility spectrometry-mass spectrometry
    Collins, DC
    Lee, ML
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 372 (01) : 66 - 73
  • [23] On the separation of enantiomers by drift tube ion mobility spectrometry
    Fernandez-Maestre, Roberto
    Doerr, Markus
    ANALYTICAL METHODS, 2022, 14 (31) : 3011 - 3020
  • [24] Separation of isomeric disaccharides by traveling wave ion mobility mass spectrometry using CO2 as drift gas
    Fasciotti, Maira
    Sanvido, Gustavo B.
    Santos, Vanessa G.
    Lalli, Priscila M.
    McCullagh, Michael
    de Sa, Gilberto F.
    Daroda, Romeu J.
    Peter, Martin G.
    Eberlin, Marcos N.
    JOURNAL OF MASS SPECTROMETRY, 2012, 47 (12): : 1643 - 1647
  • [25] Structure-drift time relationships in ion mobility mass spectrometry
    Fasciotti, Maira
    Lalli, Priscila M.
    Heerdt, Gabriel
    Steffen, Rafael A.
    Corilo, Yuri E.
    de Sa, Gilberto F.
    Daroda, Romeu J.
    Reis, Francisco de A. M.
    Morgon, Nelson H.
    Pereira, Rosana C. L.
    Eberlin, Marcos N.
    Klitzke, Clecio F.
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2013, 16 (02) : 117 - 132
  • [26] Prediction of peptide drift time in ion mobility-mass spectrometry
    Wang, Bing
    Valentine, Steve
    Raghuraman, Sriram
    Plasencia, Manolo
    Zhang, Xiang
    BMC BIOINFORMATICS, 2009, 10
  • [27] Prediction of peptide drift time in ion mobility-mass spectrometry
    Bing Wang
    Steve Valentine
    Sriram Raghuraman
    Manolo Plasencia
    Xiang Zhang
    BMC Bioinformatics, 10
  • [28] Separation of phosphorothioated oligonucleotide diastereomers using multiplexed drift tube ion mobility mass spectrometry
    Demelenne, Alice
    Nys, Gwenael
    Nix, Cindy
    Fjeldsted, John C.
    Crommen, Jacques
    Fillet, Marianne
    ANALYTICA CHIMICA ACTA, 2022, 1191
  • [29] Deciphering drift time measurements from travelling wave ion mobility spectrometry-mass spectrometry studies
    Smith, David P.
    Knapman, Tom W.
    Campuzano, Iain
    Malham, Richard W.
    Berryman, Joshua T.
    Radford, Sheena E.
    Ashcroft, Atison E.
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2009, 15 (02) : 113 - 130
  • [30] Ion Trapping for Ion Mobility Spectrometry Measurements in a Cyclical Drift Tube
    Glaskin, Rebecca S.
    Ewing, Michael A.
    Clemmer, David E.
    ANALYTICAL CHEMISTRY, 2013, 85 (15) : 7003 - 7008