Dose and image quality at CT pulmonary angiography-comparison of first and second generation dual-energy CT and 64-slice CT

被引:48
|
作者
Bauer, Ralf W. [1 ]
Kramer, Sebastian [1 ]
Renker, Matthias [1 ]
Schell, Boris [1 ]
Larson, Maya Christina [1 ]
Beeres, Martin [1 ]
Lehnert, Thomas [1 ]
Jacobi, Volkmar [1 ]
Vogl, Thomas J. [1 ]
Kerl, Josef Matthias [1 ]
机构
[1] Clin Goethe Univ, Dept Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
关键词
Dual-energy CT; 64-slice CT; Pulmonary embolism; Dose; Image quality; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; ACUTE CORONARY SYNDROME; RADIATION-EXPOSURE; LUNG PERFUSION; EMBOLISM; DIAGNOSIS; FEASIBILITY; MANAGEMENT; SETTINGS; SOCIETY;
D O I
10.1007/s00330-011-2162-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To compare dose and image quality of 64-slice, first and second generation dual-energy CT (DECT) for CT pulmonary angiography (CTPA). Methods Totally 120 patients, 30 in each group, underwent CTPA on a first generation (group 1: single-energy, 120 kV/145 mAs; group 2: DE, 140/80 kV, 70/350 mAs) or second generation dual-source DECT device (group 3: DE, 100/Sn140 kV, 120/102 mAs; group 4: DE, 80/Sn140 kV, 202/86 mAs). CTDIvol, DLP, background noise (BN), thorax diameter and attenuation in the pulmonary trunk were compared. Results Thorax diameter and attenuation in the pulmonary trunk did not differ significantly (p > 0.4 and > 0.19) between the groups. Mean CTDIvol and DLP were significantly lower (p < 0.003) in group 4 (6.2 +/- 1.6 mGy/170 +/- 41 mGycm) compared to group 1 (8.5 +/- 2.6 mGy/235 +/- 117 mGycm), group 2 (9.2 +/- 3.3 mGy/224 +/- 122 mGycm) and group 3 (8.7 +/- 2.8 mGy/246 +/- 86 mGycm). BN was significantly lower (p < 0.0001) in group 4 (12 +/- 3 HU) and group 1 (13 +/- 6 HU) compared to group 3 and 2 (16 +/- 6 HU and 23 +/- 9). Conclusion The use of second generation DECT in 80/Sn140 kV configuration allows for significant dose reduction with image quality similar to 120 kV CTPA.
引用
收藏
页码:2139 / 2147
页数:9
相关论文
共 50 条
  • [1] Dose and image quality at CT pulmonary angiography—comparison of first and second generation dual-energy CT and 64-slice CT
    Ralf W. Bauer
    Sebastian Kramer
    Matthias Renker
    Boris Schell
    Maya Christina Larson
    Martin Beeres
    Thomas Lehnert
    Volkmar Jacobi
    Thomas J. Vogl
    Josef Matthias Kerl
    European Radiology, 2011, 21 : 2139 - 2147
  • [2] Radiation dose for cardiac CT angiography - A comparison between dual source and 64-slice CT
    Hermann, Franziska
    Meyer, Tanja
    Hadamitzky, Martin
    Hendrich, Eva
    Lang, Ute
    Schwarzer, Christine
    Martinoff, Stefan
    Hausleiter, Joerg
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (10) : A142 - A142
  • [3] Image Quality and Radiation Dose for Prospectively Triggered Coronary CT Angiography: 128-Slice Single-Source CT versus First-Generation 64-Slice Dual-Source CT
    Jin Gu
    He-shui Shi
    Ping Han
    Jie Yu
    Gui-na Ma
    Sheng Wu
    Scientific Reports, 6
  • [4] Image Quality and Radiation Dose for Prospectively Triggered Coronary CT Angiography: 128-Slice Single-Source CT versus First-Generation 64-Slice Dual-Source CT
    Gu, Jin
    Shi, He-shui
    Han, Ping
    Yu, Jie
    Ma, Gui-na
    Wu, Sheng
    SCIENTIFIC REPORTS, 2016, 6
  • [5] Infant Cardiac CT Angiography with 64-Slice and 256-Slice CT: Comparison of Radiation Dose and Image Quality Using a Pediatric Phantom
    Lee, Yi-Wei
    Yang, Ching-Ching
    Mok, Greta S. P.
    Wu, Tung-Hsin
    PLOS ONE, 2012, 7 (11):
  • [6] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lukas Lenga
    Franziska Trapp
    Moritz H. Albrecht
    Julian L. Wichmann
    Addison A. Johnson
    Ibrahim Yel
    Tommaso D’Angelo
    Christian Booz
    Thomas J. Vogl
    Simon S. Martin
    European Radiology, 2019, 29 : 4603 - 4612
  • [7] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lenga, Lukas
    Trapp, Franziska
    Albrecht, Moritz H.
    Wichmann, Julian L.
    Johnson, Addison A.
    Yel, Ibrahim
    D'Angelo, Tommaso
    Booz, Christian
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN RADIOLOGY, 2019, 29 (09) : 4603 - 4612
  • [8] 64-slice CT: a comparison with invasive coronary angiography
    Ruth Kirby
    Nature Clinical Practice Cardiovascular Medicine, 2005, 2 (11): : 552 - 552
  • [9] Author Correction: Image Quality and Radiation Dose for Prospectively Triggered Coronary CT Angiography: 128-Slice Single-Source CT versus First-Generation 64-Slice Dual-Source CT
    Jin Gu
    He-shui Shi
    Ping Han
    Jie Yu
    Gui-na Ma
    Sheng Wu
    Scientific Reports, 10
  • [10] Dosimetric and image quality comparison of different 64-slice CT scanners
    Colli, V.
    Strocchi, S.
    Vite, C.
    Cacciatori, M.
    Rizzi, E.
    Marzoli, L.
    Conte, L.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 197 - 200