Elemental ratio determinations and compound-independent calibration using microsecond pulsed glow discharge time-of-flight mass spectrometry

被引:7
|
作者
Sola-Vazquez, Auristela [1 ]
Costa-Fernandez, Jose M. [1 ]
Pereiro, Rosario [1 ]
Sanz-Medel, Alfredo [1 ]
机构
[1] Univ Oviedo, Dept Phys & Analyt Chem, Fac Chem, E-33006 Oviedo, Spain
关键词
Pulsed glow discharge; Time-of-flight mass spectrometry; Speciation analysis; Elemental ratio determinations; Compound-independent calibration; MICROWAVE-INDUCED PLASMA; EMISSION SPECTROMETRY; IONIZATION SOURCE; ATOMIC-EMISSION;
D O I
10.1007/s00216-011-5145-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new prototype based on a microsecond pulsed glow discharge ion source coupled to a time-of-flight mass spectrometer was recently designed, constructed and analytically evaluated in our laboratory for simultaneous collection of elemental and molecular information, and as a gas-chromatographic detector of compounds of environmental concern. To investigate further the analytical capabilities of such a new setup, its capability for the determination of element ratios in volatile organic halogenated compounds has been explored. Moreover, compound-independent calibration has been carried out with the prototype as well. The results demonstrated that the intensity ratios (analyte to internal standard) were linear with the corresponding ratio of concentrations. Both for chlorine and bromine (measured in the prepeak) and for BrC (measured in the plateau) the slope was 1 and the intercept was 0. Moreover, detection limits were improved by more than 1 order of magnitude as compared with using external calibration. The applicability of the proposed approach has been demonstrated for the straightforward determination of chloroform in drinking and river waters.
引用
收藏
页码:2771 / 2777
页数:7
相关论文
共 50 条
  • [1] Elemental ratio determinations and compound-independent calibration using microsecond pulsed glow discharge time-of-flight mass spectrometry
    Auristela Solà-Vázquez
    José M. Costa-Fernández
    Rosario Pereiro
    Alfredo Sanz-Medel
    Analytical and Bioanalytical Chemistry, 2011, 401 : 2771 - 2777
  • [2] Microsecond-pulsed glow discharge time-of-flight mass spectrometry: Analytical advantages
    Hang, W
    Baker, C
    Smith, BW
    Winefordner, JD
    Harrison, WW
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (02) : 143 - 149
  • [3] Microsecond-pulsed Grimm glow discharge as a source for time-of-flight mass spectrometry
    Yang, CL
    Mohill, M
    Harrison, WW
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) : 1255 - 1260
  • [4] Pulsed glow discharge time-of-flight mass spectrometry
    Harrison, WW
    Hang, W
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) : 835 - 840
  • [5] MICROSECOND PULSED GLOW-DISCHARGE TIME-OF-FLIGHT MASS-SPECTROMETER
    HANG, W
    YANG, PY
    WANG, XR
    YANG, CL
    SU, YX
    HUANG, BL
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1994, 8 (08) : 590 - 594
  • [6] Explicit chemical speciation by microsecond pulsed glow discharge time-of-flight mass spectrometry: concurrent acquisition of structural, molecular and elemental information
    Majidi, V
    Moser, M
    Lewis, C
    Hang, W
    King, FL
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (01) : 19 - 25
  • [7] Explicit chemical speciation by microsecond pulsed glow discharge time-of-flight mass spectrometry: Concurrent acquisition of structural, molecular and elemental information
    Majidi, Vahid
    Moser, Matt
    Lewis, Cris
    Hang, Wei
    King, Fred L.
    1600, Royal Soc of Chemistry, Cambridge, United Kingdom (15):
  • [8] Time-of-flight mass spectrometry with a pulsed glow discharge ionization source
    Steiner, RE
    Lewis, CL
    King, FL
    ANALYTICAL CHEMISTRY, 1997, 69 (09) : 1715 - 1721
  • [9] Improving pulsed radiofrequency glow discharge for time-of-flight mass spectrometry simultaneous elemental and molecular analysis
    Auristela Solà-Vázquez
    Beatriz Fernández
    José M. Costa-Fernández
    Rosario Pereiro
    Alfredo Sanz-Medel
    Analytical and Bioanalytical Chemistry, 2014, 406 : 7431 - 7443
  • [10] Improving pulsed radiofrequency glow discharge for time-of-flight mass spectrometry simultaneous elemental and molecular analysis
    Sola-Vazquez, Auristela
    Fernandez, Beatriz
    Costa-Fernandez, Jose M.
    Pereiro, Rosario
    Sanz-Medel, Alfredo
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (29) : 7431 - 7443