Comparison of Different Fixation Types Used in Unilateral Mandibular Condylar Fractures: An In Vivo Study With New Biomechanical Model

被引:8
|
作者
Cimen, Emre
Onder, Mustafa Ercument [1 ]
Cambazoglu, Mine [2 ]
Birant, Emir [3 ]
机构
[1] Kirikkale Univ, Dept OMFS, Fac Dent, Kirikkale, Turkey
[2] Ankara Univ, Fac Dent, Dept OMFS, Ankara, Turkey
[3] Birant Makina, Ankara, Turkey
关键词
Biomechanical test; condylar fracture; double mini plate fixation; resorbable plate fixation; single-titanium mini plate fixation; CLOSED TREATMENT; NONSURGICAL TREATMENT; PLATING TECHNIQUES; INTERNAL-FIXATION; SUBCONDYLAR FRACTURES; RIGID FIXATION; CLINICAL-EVALUATION; RESORBABLE PLATES; NECK FRACTURES; FOLLOW-UP;
D O I
10.1097/SCS.0000000000002754
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: The aim of this in vivo study is to compare the single-titanium, double-titanium mini plate, and single resorbable plate systems used in internal rigid fixation of the unilateral mandibular condylar fractures on new design biomechanical model. Methods: Thirty synthetic polyurethane models were used for biomechanical testing. Fracture lines were created for each model. Fragments were fixed with single-titanium plates in Group A (n = 10), double-titanium plates in Group B (n = 10), and single biodegradable plate (PPLA) in Group C (n = 10). Masticatory forces were applied to the models and the biomechanical properties of the titanium plate and screws, resorbable plate, and screws were evaluated. Results: The average failure force for Group A, Group B, and Group C is 199, 324, 177N and the average bone displacement for Group A, Group B, Group C is 1.9, 0.3, 2.1 mm, respectively. Discussion: Double titanium plates showed the most acceptable results in the fixation of unilateral subcondylar fractures where the single titanium and biodegradable plate systems failed to provide enough stability in unilateral subcondylar fracture fixation. Biodegradable plate systems are still not an alternative in fixation of unilateral condylar fractures.
引用
收藏
页码:1277 / 1281
页数:5
相关论文
共 50 条
  • [1] Two load sharing plates fixation in mandibular condylar fractures: Biomechanical basis
    Parascandolo, Salvatore
    Spinzia, Alessia
    Parascandolo, Stefano
    Piombino, Pasquale
    Califano, Luigi
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2010, 38 (05) : 385 - 390
  • [2] Comparison of different fixation techniques for periprosthetic fractures: a biomechanical study of a new implant
    Mehmet Nuri Konya
    Ugur Yuzuguldu
    Recep Altin
    Ugur Fidan
    International Orthopaedics, 2021, 45 : 1817 - 1826
  • [3] Comparison of different fixation techniques for periprosthetic fractures: a biomechanical study of a new implant
    Konya, Mehmet Nuri
    Yuzuguldu, Ugur
    Altin, Recep
    Fidan, Ugur
    INTERNATIONAL ORTHOPAEDICS, 2021, 45 (07) : 1817 - 1826
  • [4] Biomechanical Optimization of Bone Plates Used in Rigid Fixation of Mandibular Fractures
    Lovald, Scott T.
    Wagner, Jon D.
    Baack, Bret
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 67 (05) : 973 - 985
  • [5] A Biomechanical Comparison of Different Types of Lumbopelvic Fixation
    Tis, John E.
    Helgeson, Melvin
    Lehman, Ronald A.
    Dmitriev, Anton E.
    SPINE, 2009, 34 (24) : E866 - E872
  • [6] Biomechanical evaluation of various internal fixation patterns for unilateral mandibular condylar base fractures: A three-dimensional finite element analysis
    Li, Jie
    Jiao, Jian
    Luo, Ting
    Wu, Wei
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 133
  • [7] Biomechanical evaluation of new fixation devices for mandibular angle fractures
    Wittenberg, JM
    Mukherjee, DP
    Smith, BR
    Kruse, RN
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1997, 26 (01) : 68 - 73
  • [8] Biomechanical Optimization of Bone Plates Used in Rigid Fixation of Mandibular Symphysis Fractures
    Lovald, Scott
    Baack, Bret
    Gaball, Curtis
    Olson, Garth
    Hoard, Anna
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2010, 68 (08) : 1833 - 1841
  • [9] Comparison of three different fixation constructs for radial neck fractures: a biomechanical study
    Hongwei Chen
    Dengying Wu
    Tianlong Pan
    Jun Pan
    Rui Zhang
    Xuchao Shi
    Journal of Orthopaedic Surgery and Research, 12
  • [10] Biomechanical comparison of different fixation methods in metatarsal shaft fractures: A cadaver study
    Trost, Matthias
    Knieps, Matthias
    Hackl, Michael
    Frey, Soenke
    Scaal, Martin
    Mueller, Lars Peter
    Dargel, Jens
    Gick, Sascha
    CLINICAL BIOMECHANICS, 2022, 92