Spectrophotometric determination of total proteins by derivatization with p-benzoquinone in a flow system

被引:2
|
作者
Lichtig, J
Guarilha, O
Oliveira, RMS
Masini, JC
Trujillo, LM
机构
[1] Univ Sao Paulo, Inst Quim, BR-05513970 Sao Paulo, Brazil
[2] Inst Adolfo Lutz, BR-01246902 Sao Paulo, Brazil
关键词
p-benzoquinone; total proteins; derivatization; flow system; reactor; spectrophotometric assay;
D O I
10.1081/AL-100001571
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The determination of total proteins in plasma is largely needed in clinical and biochemical areas because it is important for the diagnosis of several diseases. This paper describes a flow method for determination of total protein in human serum albumin samples using p-benzoquinone (PBQ) dissolved in dimethyl-sulphoxide (DMSO) as reagent. The reactional mixture at pH 8.0 or 9.0 was pumped through a coiled stainless steel capillary tube reactor, heated at 90 degreesC, to a spectrophotometric flow cell where the absorbance was monitored at 361 nm. A linear dynamic range was observed for protein concentrations from 50 to 500 mug mL(-1). Results were in good agreement with those obtained by the Kjeldahl's method.