Catheter Ablation of Atrial Fibrillation Using Zero-Fluoroscopy Technique: A Randomized Trial

被引:96
|
作者
Bulava, Alan [1 ,2 ]
Hanis, Jiri [1 ]
Eisenberger, Martin [1 ,2 ]
机构
[1] Budweis Hosp, Dept Cardiol, Budweis 37001, Czech Republic
[2] Univ South Bohemia, Fac Hlth & Social Studies, Budweis, Czech Republic
来源
关键词
atrial fibrillation; radiation; zero fluoroscopy; pulmonary vein isolation; contact force; catheter ablation; OCCUPATIONAL-HEALTH-HAZARDS; INTERVENTIONAL LABORATORY TIME; PULMONARY VEIN ISOLATION; RADIATION-EXPOSURE; SAFER ENVIRONMENT; TACHYCARDIA; MANAGEMENT; RISK;
D O I
10.1111/pace.12634
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundRecent advances in 3D mapping systems, such as simultaneous visualization of multiple catheters and contact force measurement, have allowed a significant reduction in fluoroscopic times during radiofrequency (RF) ablation (RFA) procedures. The objective was to investigate whether RFA of paroxysmal atrial fibrillation (PAF) using the CARTO 3 system (Biosense Webster, Diamond Bar, CA, USA) and intracardiac echocardiography (ICE) can be performed safely without fluoroscopy. Methods and ResultsEighty patients with PAF were randomized in a 1:1 ratio to undergo either fluoroscopically guided pulmonary vein isolation (PVI) (X+) or PVI without fluoroscopy (X-). In the X- fluoroscopy group, catheter placement, transseptal puncture, left atrial geometry reconstruction, and PVI were accomplished solely using ICE imaging and CARTO mapping. The total procedure duration and RF application time in both the X- and X+ groups were comparable (92.5 22.9 minutes vs 99.9 +/- 15.9 minutes, P = 0.11 and 1785 +/- 548 seconds vs 1755 +/- 450 seconds, P = 0.79, respectively). Zero fluoroscopic time was achieved in all patients in the X- group with the exception of one patient, where 8 seconds of fluoroscopy was needed to assess proper position of the guide-wire in the femoral vein. No serious procedure-related complications were recorded and no differences in arrhythmia-free survival at 12 months were found between the groups. ConclusionRFA using ICE imaging and the CARTO 3 mapping system with contact force measurement is capable of eliminating fluoroscopy in patients undergoing PVI. Exclusion of fluoroscopic imaging does not seem to compromise patient safety and does not affect overall procedure duration, RF application time, or mid-term efficacy.
引用
收藏
页码:797 / 806
页数:10
相关论文
共 50 条
  • [31] Catheter ablation of atrial fibrillation using FireMagic TrueForce ablation catheter: The TRUEFORCE trial
    Zhang, Rongfeng
    Chu, Huimin
    Liu, Shaowen
    Yang, Bing
    Han, Bing
    Xiao, Xianjie
    Ma, Chengming
    Sun, Yuanjun
    Yin, Xiaomeng
    Xia, Yunlong
    Gao, Lianjun
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2023, 46 (08): : 986 - 993
  • [32] Safety of Atrial fibrillation ablation using Zero fluoroscopy in the era of uninterrupted anticoagulation
    Kanjwal, Khalil
    Ruzieh, Mohammed
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (03) : 785 - 785
  • [33] Dabigatran in Catheter Ablation of Atrial Fibrillation: A Call for a Randomized Control Trial
    Bin Abdulhak, Aref A.
    Khan, Abdur Rahman
    Wimmer, Alan P.
    AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (12): : 2087 - 2088
  • [34] Zero Fluoroscopy Ablation of Atrial Fibrillation: A Safety and Feasibility Study
    Di Biase, Luigi
    Horton, Rodney
    Trivedi, Chintan
    Mohanty, Prasant
    Mohanty, Sanghamitra
    Bai, Rong
    Sanchez, Javier
    Gallinghouse, G. Joseph
    Zagrodzky, Jason
    Al-Ahmad, Amin
    Hranitzky, Patrick
    Burkhardt, J. David
    Natale, Andrea
    CIRCULATION, 2014, 130
  • [35] Zero-fluoroscopy trans-septal puncture and catheter ablation of a left atrial tachycardia in a pregnant woman with a prosthetic mitral valve
    Ollitrault, Pierre
    Pellissier, Arnaud
    Ferchaud, Virginie
    Havard, Justine
    Champ-Rigot, Laure
    Milliez, Paul
    EUROPACE, 2022, 24 (03): : 383 - 383
  • [36] SINGLE CENTER EXPERIENCE WITH ZERO FLUOROSCOPY ATRIAL FIBRILLATION ABLATION
    Raval, Yash
    Alam, Zenith
    Tamboli, Sarosh
    Osman, Ahmed F.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 81 (08) : 220 - 220
  • [37] ZERO FLUOROSCOPY CARDIAC ABLATION: A NOVEL WORKFLOW FOR ATRIAL FIBRILLATION
    De Raffele, Martina
    Bertini, Matteo
    Malagu', Michele
    Vitali, Francesco
    Balla, Cristina
    Pompei, Graziella
    Fiorio, Alessio
    Trichilo, Michele
    Guardigli, Gabriele
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2022, 24
  • [38] Efficacy and safety of zero-fluoroscopy ablation for supraventricular tachycardias Use of optional contact force measurement for zero-fluoroscopy ablation in a clinical routine setting
    Seizer, P.
    Bucher, V.
    Frische, C.
    Heinzmann, D.
    Gramlich, M.
    Mueller, I.
    Henning, A.
    Hofbeck, M.
    Kerst, G.
    Gawaz, M.
    Schreieck, J.
    HERZ, 2016, 41 (03) : 241 - 245
  • [39] Initial experience with zero-fluoroscopy pulmonary vein isolation in patients with atrial fibrillation: single-center observational trial
    Torma, Dalma
    Janosi, Kristof
    Debreceni, Dorottya
    Bocz, Botond
    Keseru, Mark
    Simor, Tamas
    Kupo, Peter
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [40] Meta-analysis of pulmonary vein isolation ablation for atrial fibrillation conventional vs low- and zero-fluoroscopy approaches
    Huang, Henry D.
    Abid, Qurrat-ul-ain
    Ravi, Venkatesh
    Sharma, Parikshit
    Larsen, Timothy
    Krishnan, Kousik
    Trohman, Richard G.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (06) : 1403 - 1412