Use of an intracardiac electrogram eliminates the need for a surface ECG during implantable cardioverter-defibrillator follow-up

被引:4
|
作者
Michael, Kevin A.
Peterson, Brett J.
Yue, Arthur M.
Wilson, Ryan D.
Wang, Li
Ousdigian, Kevin
Wilkoff, Bruce
Sterns, Laurence
Morgan, John M.
机构
[1] Southampton Gen Hosp, Wessex Cardiothorac Ctr, Southampton SO16 6YD, Hants, England
[2] Medtronic Inc, Minneapolis, MN USA
[3] Cleveland Clin Fdn, Cleveland, OH USA
[4] Royal Jubilee Hosp, Victoria, BC, Canada
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2007年 / 30卷 / 12期
关键词
defibrillation-ICD; electrocardiogram; new technology; pacing;
D O I
10.1111/j.1540-8159.2007.00888.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A surface electrocardiogram (SECG) for pacing threshold measurements during routine implantable cardioverter-defibrillator (ICD) follow-up can be cumbersome. This study evaluated the use of an intrathoracic far-field electrogram (EGM) derived between the Can and superior vena cava (SVC) electrode-the Leadless electrocardiogram (LLECG), in dual chamber ICDs in performing pacing threshold tests. Methods: The LLECG was evaluated prospectively during atrial and ventricular pacing threshold testing as a substudy of the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators trial (EMPIRIC) in which dual chamber ICDs were implanted in 888 patients. Threshold tests were conducted at 1 volt by decrementing the pulse width. Follow-up at three months compared pacing thresholds measured using LLECG with those using Lead I of the surface ECG (SECG). The timesaving afforded by LLECG was assessed by a questionnaire. Results: The median threshold difference between LLECG and SECG measurements for both atrial (0.00 ms, P = 0.90) and ventricular (0.00 ms, P = 0.34) threshold tests were not significant. Ninety percent of atrial and ventricular threshold differences were bounded by +/- 0.10 ms and -0.10 to +0.04 ms, respectively. We found that 99% of atrial and ventricular thresholds tests at six and 12 months attempted using LLECG were successfully completed. The questionnaire indicated that 65% of healthcare professionals found LLECG to afford at least some timesaving during device follow-ups. Conclusion: Routine follow-up can be performed reliably and expeditiously in dual chamber Medtronic (Minneapolis, MN, USA) ICDs using LLECG alone, resulting in overall timesaving.
引用
收藏
页码:1432 / 1437
页数:6
相关论文
共 50 条
  • [11] Long-term follow-up of children and young adults treated with implantable cardioverter-defibrillator: the authors' own experience with optimal implantable cardioverter-defibrillator programming
    Lewandowski, Michal
    Sterlinski, Maciej
    Maciag, Aleksander
    Syska, Pawel
    Kowalik, Ilona
    Szwed, Hanna
    Chojnowska, Lidia
    Przybylski, Andrzej
    EUROPACE, 2010, 12 (09): : 1245 - 1250
  • [12] Contributions of remote monitoring to the follow-up of implantable cardioverter-defibrillator leads under advisory
    Guedon-Moreau, Laurence
    Chevalier, Philippe
    Marquie, Christelle
    Kouakam, Claude
    Klug, Didier
    Lacroix, Dominique
    Brigadeau, Francois
    Kacet, Salem
    EUROPEAN HEART JOURNAL, 2010, 31 (18) : 2246 - 2252
  • [13] Outcome of advisory implantable cardioverter-defibrillator replacement: One-year follow-up
    Gould, Paul A.
    Gula, Lorne J.
    Champagne, Jean
    Healyey, Jeffrey S.
    Cameron, Doug
    Simpson, Christophers
    Thibault, Bernard
    Pinter, Arnold
    Tung, Stanley
    Sterns, Laurence
    Birnie, David
    Exner, Derek
    Parkash, Ratika
    Skanes, Allan C.
    Yee, Raymond
    Klein, George J.
    Krahn, Andrew D.
    HEART RHYTHM, 2008, 5 (12) : 1675 - 1681
  • [14] Predictors of implantable cardioverter-defibrillator therapies: a 6-year follow-up study
    De Almeida Fernandes, D.
    Simoes, M. R.
    Baptista, G.
    Silva, A. L.
    Santos, T.
    Guimaraes, J.
    Costa, G.
    Monteiro, E.
    Fernandes, R.
    Lopes, V.
    Antonio, N.
    Elvas, L.
    Goncalves, L.
    EUROPEAN JOURNAL OF HEART FAILURE, 2024, 26 : 22 - 22
  • [15] Long-term follow-up of paediatric patients with a subcutaneous implantable cardioverter-defibrillator
    Van der Stuijt, W.
    De Wilde, K. C.
    Baalman, S. W. E.
    Quast, A. B. E.
    Blom, N. A.
    Wilde, A. A. M.
    Knops, R. E.
    EUROPEAN HEART JOURNAL, 2020, 41 : 3209 - 3209
  • [16] A Case of Very Early Lead Fracture in an Implantable Cardioverter-Defibrillator: Management and Follow-Up
    Shawki, Marwan
    Mirzaee, Sam
    McCormack, Cameron
    Wynn, Gareth J.
    Lim, Han S.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2025, 48 (01): : 68 - 71
  • [17] COMBINED AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR AND PACEMAKER SYSTEMS - IMPLANTATION TECHNIQUES AND FOLLOW-UP
    EPSTEIN, AE
    KAY, GN
    PLUMB, VJ
    SHEPARD, RB
    KIRKLIN, JK
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (01) : 121 - 131
  • [18] Follow-up of the patient with automatic implantable cardioverter defibrillator
    Cruz Cardentey, Marleny
    Prieto Guerra, Maidelis
    Gutierrez Lopez, Alain
    Mengana Betancourt, Ana
    Ramos Cordero, Michel Y.
    Tannis Wilson, Eko
    Tamargo Barbeito, Teddy O.
    CORSALUD, 2020, 12 (03): : 283 - 291
  • [19] Longitudinal Follow-Up of Implantable Cardioverter Defibrillator Leads
    Liu, Jeffrey
    Brumberg, Genevieve
    Rattan, Rohit
    Patel, Divyang
    Adelstein, Evan
    Jain, Sandeep
    Saba, Samir
    AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (01): : 103 - 106
  • [20] Implantable cardioverter-defibrillator shocks during long-term follow-up in arrhythmogenic right ventricular cardiomyopathy
    Muller, Steven A.
    Riele, Anneline S. J. M. te
    HEART, 2023, 109 (15) : 1126 - 1127