Glucose recovery in a microfluidic microdialysis biochip

被引:19
|
作者
Hsieh, YC
Zahn, JD
机构
[1] Penn State Univ, Dept Bioengn, University Pk, PA 16802 USA
[2] Penn State Univ, Inst Mat Res, University Pk, PA 16802 USA
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2005年 / 107卷 / 02期
关键词
microdialysis; diabetes; glucose sensing; microfluidics;
D O I
10.1016/j.snb.2004.11.039
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microdialysis is a technique for separating biomolecules by controlling the mass transfer rate of biomolecules diffusing across a semipermeable membrane. It has been used in the medical community to continuously monitor various metabolites (glucose, lactate, sodium, etc.) in patients. Glucose microdialysis is of particular interest in diabetes treatment. Commercially available microdialysis probes have been used to create microdialysis sampling systems with a glucose biosensor. However, most probes are operated in a diffusion limited regime and struggle to obtain high recoveries of analytes. In this paper, an on-chip microdialysis system is presented. Dialysis of sulforhodamine fluorescent dye (MW 607) and glucose (MW 180) are demonstrated. The average permeability of the membrane to the fluorescent dye is 0.3471 mu m/s (standard deviation (S.D.) 0.0235 mu m/s) and is 2.5280 mu m/s (S.D. 0.3723 mu m/s) for glucose. The results of fluorescence and glucose dialysis show consistency and follow analytically predicted results. The system is sensitive to rapid changes in glucose reservoir concentration. Within 10 min of a reservoir concentration change, the outlet concentration of the dialysate has reached steady state with an 80% glucose recovery at a flowrate of 1.5 mu l/min. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:649 / 656
页数:8
相关论文
共 50 条
  • [21] Microfluidic differential immunocapture biochip for specific leukocyte counting
    Umer Hassan
    Nicholas N Watkins
    Bobby Reddy
    Gregory Damhorst
    Rashid Bashir
    Nature Protocols, 2016, 11 : 714 - 726
  • [22] Key technologies for LSPR-sensing microfluidic biochip
    Wang Hao
    Luo Xiangang
    Yao Harimin
    Du Chunlei
    Zhao Ze Yu
    Zhu Shaoli
    DESIGN, MANUFACTURING, AND TESTING OF MICRO- AND NANO-OPTICAL DEVICES AND SYSTEMS, 2007, 6724
  • [23] Direct electrical detection of target cells on a microfluidic biochip
    Javanmard, Mehdi
    Talasaz, AmirAli H.
    Nemat-Gorgani, Mohsen
    Pease, Fabian
    Ronaghi, Mostafa
    Davis, Ronald W.
    MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS VI, 2008, 6886
  • [24] Parallel testing optimization method of digital microfluidic biochip
    Huang, Xijun
    Xu, Chuanpei
    Zhang, Long
    Hu, Cong
    Mo, Wei
    MEASUREMENT, 2022, 194
  • [25] A New Combined Routing Technique in Digital Microfluidic Biochip
    Bhattacharya, Rupam
    Roy, Pranab
    Rahaman, Hafizur
    EMERGING TECHNOLOGIES IN DATA MINING AND INFORMATION SECURITY, IEMIS 2018, VOL 1, 2019, 755 : 441 - 450
  • [26] Microfluidic differential immunocapture biochip for specific leukocyte counting
    Hassan, Umer
    Watkins, Nicholas N.
    Reddy, Bobby, Jr.
    Damhorst, Gregory
    Bashir, Rashid
    NATURE PROTOCOLS, 2016, 11 (04) : 714 - 726
  • [27] Design-for-Trust Technique for Microfluidic Biochip Layout
    Gountia, Debasis
    Roy, Sudip
    PROCEEDINGS OF 2019 IEEE REGION 10 SYMPOSIUM (TENSYMP), 2019, : 704 - 709
  • [28] Harmonic Dilution Gradient Preparation for Digital Microfluidic Biochip
    Podder, Dipannita
    Das, Kunal
    2015 ANNUAL IEEE INDIA CONFERENCE (INDICON), 2015,
  • [29] BIOSENSOR AND MICRODIALYSIS FOR MONITORING GLUCOSE
    BOUTELLE, MG
    FELLOWS, LK
    FRAY, A
    FILLENZ, M
    JOURNAL OF NEUROCHEMISTRY, 1994, 63 : S44 - S44
  • [30] Co-integrated microfluidic and THz functions for biochip devices
    Laurette, S.
    Treizebre, A.
    Bocquet, B.
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2011, 21 (06)