Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE

被引:31
|
作者
Schroder, David T. [2 ]
Kelly, Natalie H. [1 ]
Wright, Timothy M. [1 ]
Parks, Michael L. [2 ]
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Adult Reconstruct & Joint Replacement Div, New York, NY 10021 USA
关键词
TOTAL HIP; POLYETHYLENE;
D O I
10.1007/s11999-010-1552-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Highly crosslinked UHMWPE is associated with increased wear resistance in hip simulator and clinical studies. Laboratory and case studies, however, have described rim fracture in crosslinked acetabular liners. Controversy exists, therefore, on the relative merits of crosslinked liners over conventional liners in terms of wear performance versus resistance to fatigue cracking. Questions/purposes We asked whether crosslinked liners would show less surface damage than conventional liners but would be more susceptible to fatigue damage. Methods We examined 36 conventional UHMWPE and 39 crosslinked UHMWPE retrieved implants with similar patient demographics and identical design for evidence of wear damage, including articular surface damage, impingement, screw-hole creep, and rim cracks. Results We observed no difference in wear damage scores for the two liners. Conventional liners more frequently impinged but were more often elevated with smaller head sizes. We observed creep in approximately 70% of both types of liners. Incipient rim cracks were found in five crosslinked liners, and one liner had a rim fracture. Only one conventional liner had an incipient rim crack. Conclusions Contrary to our expectation, damage was similar between crosslinked and conventional UHMWPE liners. Moreover, the 15% occurrence (six of 39) of incipient or complete fractures in crosslinked liners as compared with a 3% occurrence (one of 36) in conventional liners may have implications for the long-term performance of crosslinked liners. Longer-term studies will be necessary to establish the fate of rim cracks and thus the overall clinical fatigue performance of crosslinked liners.
引用
收藏
页码:387 / 394
页数:8
相关论文
共 50 条
  • [31] High crystallinity makes excellent wear resistance in crosslinked UHMWPE
    Wang, Zirui
    Li, Quan
    Chen, Xianchun
    Zhang, Qin
    Wang, Ke
    POLYMER, 2023, 280
  • [32] Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)
    Muratoglu, OK
    Bragdon, CR
    O'Connor, DO
    Jasty, M
    Harris, WH
    Gul, R
    McGarry, F
    BIOMATERIALS, 1999, 20 (16) : 1463 - 1470
  • [33] Characterization of the wear performance of crosslinked UHMWPE and relationship to molding procedures
    St John, KR
    Poggie, RA
    CROSSLINKED AND THERMALLLY TREATED ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE FOR JOINT REPLACEMENTS, 2004, 1445 : 221 - 227
  • [34] Effects of simulated oxidation on the in vitro wear and mechanical properties of irradiated and melted highly crosslinked UHMWPE
    Oral, Ebru
    Neils, Andrew L.
    Doshi, Brinda N.
    Fu, Jun
    Muratoglu, Orhun K.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (02) : 316 - 322
  • [35] The effect of surface topograph of retrieved femoral heads on the wear of UHMWPE sockets
    Hall, RM
    Siney, P
    Unsworth, A
    Wroblewski, BM
    MEDICAL ENGINEERING & PHYSICS, 1997, 19 (08) : 711 - 719
  • [36] Prediction of multiaxial mechanical behavior for conventional and highly crosslinked UHMWPE using a hybrid constitutive model
    Bergström, JS
    Rimnac, CM
    Kurtz, SM
    BIOMATERIALS, 2003, 24 (08) : 1365 - 1380
  • [37] Comparison of conventional and highly crosslinked UHMWPE patellae evaluated by a new in vitro patellofemoral joint simulator
    Burroughs, Brian R.
    Rubash, Harry E.
    Estok, Daniel
    Jasty, Murali
    Krevolin, Janet
    Muratoglu, Orhun K.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2006, 79B (02) : 268 - 274
  • [38] Multiaxial fatigue behavior of conventional and highly crosslinked UHMWPE during cyclic small punch testing
    Villarraga, ML
    Kurtz, SM
    Herr, MP
    Edidin, AA
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2003, 66A (02) : 298 - 309
  • [39] Notch strengthening and hardening behavior of conventional and highly crosslinked UHMWPE under applied tensile loading
    Sobieraj, MC
    Kurtz, SM
    Rimnac, CM
    BIOMATERIALS, 2005, 26 (17) : 3411 - 3426
  • [40] The effect of lipid absorption on the mechanical properties of crosslinked and conventional UHMWPE
    Turner, JL
    Kurtz, SM
    Bracco, P
    Costa, L
    PROCEEDINGS OF THE IEEE 29TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2003, : 178 - 178