Calibration of AC Shunts at Frequencies up to 10 kHz Using Digital Sampling

被引:0
|
作者
Djokic, Branislav [1 ]
Parks, Harold [1 ]
机构
[1] Natl Res Council Canada, Measurement Sci & Stand, Ottawa, ON, Canada
关键词
AC shunt; current measurement; measurement standards; calibration; measurement uncertainty; currentcomparator; calculable resistor; SYSTEM;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes a calibration system for AC shunts using digital sampling. It provides both ratio errors and phase displacements of a shunt under test with respect to a reference shunt. In its basic configuration, the calibration system operates at currents of up to 10 A and frequencies of up to 10 kHz. The expanded uncertainties (k = 2) of the described calibration system are estimated to be better than 30 mu A/A for magnitude and 30 mu rad for phase at 10 kHz, and lower than these at lower frequencies.
引用
收藏
页数:2
相关论文
共 50 条
  • [31] AC calibration methods of magnetic flux density coil standard up to 100 kHz
    Ulvr, Michal
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [32] AC Susceptibility Measurement in High Frequency Region up to 10 kHz using a SQUID Magnetometer MPMS
    Tsuruta, K.
    Mito, M.
    Iriguchi, K.
    Komorida, Y.
    Deguchi, H.
    Takagi, S.
    Ohta, T.
    INTERNATIONAL CONFERENCE ON MAGNETISM (ICM 2009), 2010, 200
  • [33] Monel Contribution to AC Losses in MgB2 Wires in Frequencies Up To 18 kHz
    Nikulshin, Yasha
    Wolfus, Shuki
    Friedman, Alex
    Ginodman, Vladmir
    Grasso, Gianni
    Tropeano, Matteo
    Bovone, Gianmarco
    Vignolo, Maurizio
    Ferdeghini, Carlo
    Yeshurun, Yosef
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2018, 28 (08)
  • [34] Calibration of the pressure sensitivity of microphones by a free-field method at frequencies up to 80 kHz
    Zuckerwar, Allan J.
    Herring, G.C.
    Elbing, Brian R.
    Journal of the Acoustical Society of America, 2006, 119 (01): : 320 - 329
  • [35] Calibration of the pressure sensitivity of microphones by a free-field method at frequencies up to 80 kHz
    Zuckerwar, AJ
    Herring, GC
    Elbing, BR
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2006, 119 (01): : 320 - 329
  • [36] Characterization of AC-DC Transfer Shunts up to 100 A at 100 kHz Using Two-Stage, Amplifier-Aided Current Transformers
    Kinard, Joseph
    Laug, Owen
    Waltrip, Bryan
    Lipe, Thomas
    2012 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2012, : 107 - 110
  • [37] INVESTIGATION OF AC DIFFERENTIAL CURRENT TRANSFORMERS USING A CURRENT STEP-UP CIRCUIT FOR 1-10 KHZ
    FORGER, K
    SCHLINKE, H
    ISA TRANSACTIONS, 1970, 9 (03) : 176 - &
  • [38] PRECISION ILLUMINATED OUTPUT METER CLASS-0.2 FOR FREQUENCIES UP TO 10-KHZ
    PHILIPP, W
    ARCHIV FUR TECHNISCHES MESSEN UND MESSTECHNISCHE PRAXIS, 1969, (397): : 31 - &
  • [39] APPARATUS FOR DETERMINATION OF MAGNETIC-PROPERTIES OF ELECTRICAL STEEL SHEETS AT FREQUENCIES UP TO 10 KHZ
    SCHMIDT, KH
    BIRKELBACH, G
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1978, 9 (1-3) : 123 - 129
  • [40] Calibration of Inductive Voltage Dividers at Power Frequencies Using an AC-PJVS
    Hsu, Jimmy C.
    Chen, Shih-Fang
    Kuo, Chin-Jung
    Hsiao, Mark
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,