Feasibility of Deep Learning-Based PET/MR Attenuation Correction in the Pelvis Using Only Diagnostic MR Images

被引:43
|
作者
Bradshaw, Tyler J. [1 ]
Zhao, Gengyan [2 ]
Jang, Hyungseok [3 ]
Liu, Fang [1 ]
McMillan, Alan B. [1 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Dept Radiol, Madison, WI 53706 USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI 53706 USA
[3] Univ Calif San Diego, Dept Radiol, San Diego, CA 92103 USA
基金
美国国家卫生研究院;
关键词
PET/MR; attenuation correction; deep learning; QUANTITATIVE-EVALUATION; CT IMAGES; SEGMENTATION; GENERATION; TIME;
D O I
10.18383/j.tom.2018.00016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study evaluated the feasibility of using only diagnostically relevant magnetic resonance (MR) images together with deep learning for positron emission tomography (PET)/MR attenuation correction (deepMRAC) in the pelvis. Such an approach could eliminate dedicated MRAC sequences that have limited diagnostic utility but can substantially lengthen acquisition times for multibed position scans. We used axial T2 and T1 LAVA Flex magnetic resonance imaging images that were acquired for diagnostic purposes as inputs to a 3D deep convolutional neural network. The network was trained to produce a discretized (air, water, fat, and bone) substitute computed tomography (CT) (CTsub). Discretized (CTref-discrete) and continuously valued (CTref) reference CT images were created to serve as ground truth for network training and attenuation correction, respectively. Training was performed with data from 12 subjects. CTsub, CTref, and the system MRAC were used for PET/MR attenuation correction, and quantitative PET values of the resulting images were compared in 6 test subjects. Overall, the network produced CTsub with Dice coefficients of 0.79 +/- 0.03 for cortical bone, 0.98 +/- 0.01 for soft tissue (fat: 0.94 +/- 0.0; water: 0.88 +/- 0.02), and 0.49 +/- 0.17 for bowel gas when compared with CTref-discrete. The root mean square error of the whole PET image was 4.9% by using deepMRAC and 11.6% by using the system MRAC. In evaluating 16 soft tissue lesions, the distribution of errors for maximum standardized uptake value was significantly narrower using deepMRAC (-1.0% +/- 1.3%) than using system MRAC method (0.0% +/- 6.4%) according to the Brown-Forsy the test (P < .05). These results indicate that improved PET/MR attenuation correction can be achieved in the pelvis using only diagnostically relevant MR images.
引用
收藏
页码:138 / 147
页数:10
相关论文
共 50 条
  • [1] Attenuation Correction of PET/MR Using Deep Neural Network Based on Dixon and ZTE MR Images
    Gong, Kuang
    Yang, Jaewon
    Kim, Kyungsang
    El Fakhri, Georges
    Seo, Youngho
    Li, Quanzheng
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [2] Deep Learning MR Imaging-based Attenuation Correction for PET/MR Imaging
    Liu, Fang
    Jang, Hyungseok
    Kijowski, Richard
    Bradshaw, Tyler
    McMillan, Alan B.
    RADIOLOGY, 2018, 286 (02) : 676 - 684
  • [3] Validation of model-based pelvis bone segmentation from MR images for PET/MR attenuation correction
    Renisch, S.
    Blaffert, T.
    Tang, J.
    Hu, Z.
    MEDICAL IMAGING 2012: IMAGE PROCESSING, 2012, 8314
  • [4] Feasibility of using respiration-averaged MR images for attenuation correction of cardiac PET/MR imaging
    Ai, Hua
    Pan, Tinsu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (04): : 311 - 321
  • [5] Brain PET-MR attenuation correction with deep learning
    Yaakub, S. N.
    McGinnity, C. J.
    Beck, K.
    Merida, I.
    Dunston, E.
    Muffoletto, M.
    Qureshi, A.
    Bhattacharya, S.
    MacKewn, J.
    Hammers, A.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2019, 39 : 600 - 601
  • [6] Pelvic PET/MR attenuation correction in the image space using deep learning
    Abrahamsen, Bendik Skarre
    Knudtsen, Ingerid Skjei
    Eikenes, Live
    Bathen, Tone Frost
    Elschot, Mattijs
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [7] Evaluating a radiotherapy deep learning synthetic CT algorithm for PET-MR attenuation correction in the pelvis
    Wyatt, Jonathan J.
    Kaushik, Sandeep
    Cozzini, Cristina
    Pearson, Rachel A.
    Petrides, George
    Wiesinger, Florian
    Mccallum, Hazel M.
    Maxwell, Ross J.
    EJNMMI PHYSICS, 2024, 11 (01)
  • [8] Deep Learning based Segmentation for Multi MR Imaging Protocols using Transfer Learning for PET Attenuation Correction
    Mecheter, Imene
    Amira, Abbes
    Abbod, Maysam
    Zaidi, Habib
    2020 IEEE SYMPOSIUM SERIES ON COMPUTATIONAL INTELLIGENCE (SSCI), 2020, : 2516 - 2520
  • [9] Attenuation correction and truncation completion for breast PET/MR imaging using deep learning
    Li, Xue
    Johnson, Jacob M.
    Strigel, Roberta M.
    Bancroft, Leah C. Henze
    Hurley, Samuel A.
    Estakhraji, S. Iman Zare
    Kumar, Manoj
    Fowler, Amy M.
    McMillan, Alan B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (04):
  • [10] MR-Based PET Attenuation Correction for PET/MR Imaging
    Bezrukov, Ilja
    Mantlik, Frederic
    Schmidt, Holger
    Schoelkopf, Bernhard
    Pichler, Bernd J.
    SEMINARS IN NUCLEAR MEDICINE, 2013, 43 (01) : 45 - 59