摘要
To improve the characterization of the detection efficiency of a NaI(Tl) scintillation detector in the low energy range, measurements and Monte Carlo simulations were performed with a certified reference material (CRM) and disk sources. The detection efficiency as modeled. In the MCNP simulation, the detection efficiency dropped as expected at 33.17 keV. The X-ray escape correction factors of a Tl-204 disk source and MCNP simulations were compared, a factor of the Tl-204 was similar. Moreover, Correcting the X-ray escape and using fewer sources than the CRM resulted in better detection efficiency for the NaI(Tl) detector in the low energy range.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] The detection efficiency study of NaI(Tl) scintillation detector with the different numbers of SiPMs
    Wang, Bao
    Zhang, Xiongjie
    Wang, Qingshan
    Wang, Dongyang
    Li, Dong
    Xiahou, Mingdong
    Zhou, Pengfei
    Ye, Hao
    Hu, Bin
    Zhang, Lijiao
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2022, 54 (07) : 2564 - 2571
  • [2] SCINTILLATION EFFICIENCY OF A NAI(TL) CRYSTAL FOR LOW-ENERGY ELECTRONS
    TOJO, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 241 (01): : 177 - 180
  • [3] A computational modelling of low-energy gamma ray detection efficiency of a cylindrical NaI(Tl) detector
    Ahmadi, S.
    Ashrafi, S.
    Yazdansetad, F.
    Roshanshah, N.
    Tarim, U. Akar
    Gurler, O.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 188
  • [4] THE RESOLVING CAPABILITY OF THE NAI(TL) SCINTILLATION DETECTOR
    MATHIESON, AML
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1982, 15 (FEB) : 98 - 99
  • [5] Detection Efficiency of NaI(Tl) Detector in 511-1332 keV Energy Range
    Akkurt, I.
    Gunoglu, K.
    Arda, S. S.
    SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS, 2014, 2014
  • [6] Detection efficiency of NaI(Tl) detector based on the fabricated calibration of HPGe detector
    El-Gamal, Hany
    Negm, Hani
    Hasabelnaby, Mohamed
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 360 - 366
  • [7] FABRICATION OF THIN NAI(TL) SCINTILLATION LAYERS FOR LOW ENERGY X-RAY DETECTION
    BAUER, RW
    WEINGART, RC
    NUCLEAR INSTRUMENTS & METHODS, 1967, 55 (01): : 55 - &
  • [8] APPLICATION OF NAI(TL) SCINTILLATION DETECTOR TO MEASUREMENT OF TRITIUM CONCENTRATION
    NISHIZAWA, K
    ENDO, Y
    SHINAGAWA, M
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 1977, 14 (10) : 705 - 709
  • [9] THE SCINTILLATION EFFICIENCY OF SODIUM AND IODINE RECOILS IN A NAI(TL) DETECTOR FOR DARK-MATTER SEARCHES
    SPOONER, NJC
    DAVIES, GJ
    DAVIES, JD
    PYLE, GJ
    BUCKNELL, TD
    SQUIER, GTA
    LEWIN, JD
    SMITH, PF
    PHYSICS LETTERS B, 1994, 321 (1-2) : 156 - 160
  • [10] EFFECT OF TEMPERATURE UPON THE SCINTILLATION EFFICIENCY OF NAI(TL)
    WEBB, LA
    JOHNSON, RF
    PHYSICAL REVIEW, 1955, 98 (01): : 234 - 234