A nonparametric method for automatic correction of intensity nonuniformity in MRI data

被引:3839
|
作者
Sled, JG
Zijdenbos, AP
Evans, AC
机构
[1] Montreal Neurol Inst, McConnell Brain Imaging Ctr, Montreal, PQ H3A 2B4, Canada
[2] McGill Univ, Montreal, PQ H3A 2B4, Canada
关键词
intensity nonuniformity; magnetic resonance imaging; RF field inhomogeneity; shading artifact;
D O I
10.1109/42.668698
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A novel approach to correcting for intensity nonuniformity in magnetic resonance (MR) data is described that achieves high performance without requiring a model of the tissue classes present. The method has the advantage that it can be applied at an early stage in an automated data analysis, before a tissue model is available. Described as nonparametric nonuniform intensity normalization (N3), the method is independent of pulse sequence and insensitive to pathological data that might otherwise violate model assumptions. To eliminate the dependence of the field estimate on anatomy, an iterative approach is employed to estimate both the multiplicative bias field and the distribution of the true tissue intensities, The performance of this method is evaluated using both real and simulated MR data.
引用
收藏
页码:87 / 97
页数:11
相关论文
共 50 条
  • [21] A new method of IRFPA nonuniformity correction
    Dai, Shaosheng
    Zhang, Tianqi
    Gao, Jian
    Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics), 2007, 4492 LNCS (PART 2): : 640 - 645
  • [22] A new method of IRFPA nonuniformity correction
    Dai, Shaosheng
    Zhang, Tianqi
    Gao, Jian
    ADVANCES IN NEURAL NETWORKS - ISNN 2007, PT 2, PROCEEDINGS, 2007, 4492 : 640 - +
  • [23] Automatic correction of intensity inhomogeneities improves unsupervised assessment of abdominal fat by MRI
    Positano, Vincenzo
    Cusi, Kenneth
    Santarelli, Maria Filomena
    Sironi, Annamaria
    Petz, Roberta
    DeFronzo, Ralph
    Landini, Luigi
    Gastaldelli, Amalia
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (02) : 403 - 410
  • [24] Nonparametric Density Flows for MRI Intensity Normalisation
    Castro, Daniel C.
    Glocker, Ben
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2018, PT I, 2018, 11070 : 206 - 214
  • [25] Correction of aeroheating-induced intensity nonuniformity in infrared images
    Liu, Li
    Yan, Luxin
    Zhao, Hui
    Dai, Xiaobing
    Zhang, Tianxu
    INFRARED PHYSICS & TECHNOLOGY, 2016, 76 : 235 - 241
  • [26] Method of nonuniformity correction for IRFPA with nonlinear response
    Feng Lin
    Liu Shuang
    Zhao Kai-Sheng
    Guan An-Quan
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 2006, 25 (03) : 221 - 224
  • [27] Correction method for pixel response nonuniformity of CCD
    Cheng, Wan-Sheng
    Zhao, Jie
    Cai, He-Gao
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2008, 16 (02): : 314 - 318
  • [28] A nonuniformity correction method based on Bayesian framework
    Qian Wei-Xian
    Ren Jian-Le
    Chen Qian
    Gu Guo-Hua
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 2013, 32 (06) : 491 - +
  • [29] Method of nonuniformity correction for irfpa with nonlinear response
    College of Optoelectronic Information, University of Electronics Science and Technology of China, Chengdu 610054, China
    Hongwai Yu Haomibo Xuebao, 2006, 3 (221-224):
  • [30] Correction for Intensity Inhomogeneity in MRI
    Hui, C.
    Narayana, P.
    MEDICAL PHYSICS, 2009, 36 (06)